BINDER ED 56
BINDER ·Made in Germany · In Active Production

BINDER ED 56

BINDER ED 56 — Drying ovens with Temperature 5…300 °C.

The BINDER ED 56 is a 57 L natural-convection drying and heating oven with a +5 to 300 °C working range, 1.05 kW single-phase heating and 8 air changes per hour. Within the ED series it is the smallest standard chamber, distinguished by 30 kg per rack and 70 kg total load capacity. It serves precise tempering, drying and heat treatment of non-flammable solids, powders and glassware in laboratory and production processes.

Temperature
5…300 °C

BINDER ED Series — Model Comparison & Selection

Model Temperature, °CAir change rate, 1/hourDepth, mmHeating up time to 150 c, minHeight, cmInterior volume, L
ED 56 Current 5–30085655062.557
ED 115 5–30066054570.5114
ED 23 5–300135202849.520
ED 260 5–30037605094255
ED 400 5–3001076571102.5400
ED 720 5–300287085159743

Full specifications

Core Performance & Functional Specifications
Temperature 5...300 °C
Air change rate 8 x/h
Heating time to 150°C 50 min
Interior volume 57 L
Load per rack 30 kg (ED056-230V / ED056UL-120V)
Nominal power 1.05 kW
Permitted load 70 kg (ED056-230V / ED056UL-120V)
Temperature fluctuation 0.4 K
Temperature uniformity 2.3 K
Rear wall clearance 160 mm (ED056-230V / ED056UL-120V)
Side wall clearance 100 mm (ED056-230V / ED056UL-120V)
Article Number mfr 9010
Recovery time after door was opened for 30 s at 150°C mfr 19 min
Dimensions, Weight & Mechanical Design
Depth 565 mm (ED056-230V / ED056UL-120V)
Height 625 mm (ED056-230V / ED056UL-120V)
Interior width 360 mm (ED056-230V / ED056UL-120V)
Width 560 mm (ED056-230V / ED056UL-120V)
Net weight of the unit (empty) mfr 39 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Unit fuse 6.3 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Behaviour under load and in operation

Installation, connection and control

Integration & Operating Notes

Mounting
Maintain 160 mm rear and 100 mm side clearance

Insufficient clearance prevents heat dissipation and can cause overheating or invalid temperature uniformity. Install on a non-flammable surface in a ventilated room, never in an unventilated recess, and keep the specified distances to walls and adjacent equipment.

Factory Specification Sheet; Quick Start Guide, Requirements for the location of installation
Electrical
Supply must stay within ±10 % of nominal voltage

The 230 V variant requires a 230 V 1~ 50/60 Hz supply with a 6.3 A time-lag (T) 5×20 mm / 250 V external chamber fuse. Verify supply quality before connection; fluctuations beyond ±10 % may prevent the chamber from meeting its performance and safety limits.

Quick Start Guide, Electrical connection data and installation
Control & I/O
Set safety controller 2–5 °C above the temperature set-point

The safety controller set-point is an absolute maximum limit temperature. After every change of the temperature set-point, re-check and adjust the safety controller set-point to 2–5 °C above it; the permissible range is 10 °C to 300 °C with 1 °C accuracy.

Quick Start Guide, Set-point entry for temperature; Setting the safety controller set-point
Maintenance
Schedule annual maintenance and electrical safety testing

Maintenance must be performed at least once a year by licensed electricians or experts authorized by BINDER. Include repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 before initial startup and after maintenance or repair.

Operating Manual BD-ED-FD (E2), Section 10.1 and 10.2
Maintenance
Replace door gasket only when the chamber is cold

Replacing the door gasket while the chamber is warm can damage the gasket. Allow the chamber to cool to ambient temperature before replacement; use only original BINDER spare parts.

Operating Manual BD-ED-FD (E2), Section 10.2
Process Limits
Charging material must be non-flammable and below its flash point

The chamber is intended for harmless materials only. Charging material must not be explosive or flammable, must not form an explosive mixture with air, and the operating temperature must lie below its flash point or sublimation point.

Quick Start Guide, Intended use
Cooling & Media
Hot-air sterilization possible at 190 °C for at least 30 minutes

The ED series can be hot-air sterilized at 190 °C / 374 °F for at least 30 minutes. Remove all inflammable goods from the interior before sterilization. Racks can be removed and autoclaved.

Quick Start Guide, Cleaning and decontamination

Selection & Integration Q&A

When should I choose the ED 56 over the larger ED 115 or ED 260?

Choose the ED 56 for small loads up to 70 kg total and 30 kg per rack in a 57 L chamber. Step up to ED 115 or ED 260 when you need more usable volume or higher heating power (1.25 kW or 2.25 kW); all share the +5 to 300 °C range and natural convection.

What supply and fusing does the ED 56 require, and can it be connected to a standard wall outlet?

The 230 V variant (ED056-230V) uses a CEE 7/7 plug on a 230 V 1~ 50/60 Hz supply with a 6.3 A time-lag (T) 5×20 mm / 250 V external chamber fuse. The UL variant (ED056UL-120V) uses a NEMA 5-15 plug at 120 V 1~ 60 Hz. Verify ±10 % voltage tolerance.

What ambient conditions and clearances are mandatory for installing the ED 56?

Install indoors on a non-flammable surface with 160 mm rear and 100 mm side clearance, ambient temperature +18 to +40 °C, humidity ≤70 % r.h. non-condensing, altitude ≤2000 m. Do not install in unventilated recesses or explosive areas.

What maintenance and accessories are required beyond the base delivery?

Schedule annual maintenance by BINDER-authorized personnel and electrical safety testing per EN 50678/VDE 0701. Recommended consumables: neutral cleaning agent Art. 1002-0016 and disinfectant spray Art. 1002-0022. Optional RS 422 interface for APT-COM 4 and analog temperature output are not included.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model ED 56 en
Datasheet ·363,4 КБ ·7 pp.
View PDF
PDF
QuickStartGuide BD BF ED FD FED E31 03 2023 dt en fr
Manual ·1,7 МБ ·35 pp.
View PDF
PDF
BD ED FD (E2) 09 2026 en
Manual ·12,9 МБ ·71 pp.
View PDF