BINDER ED 56
BINDER ED 56 — Drying ovens with Temperature 5…300 °C.
The BINDER ED 56 is a 57 L natural-convection drying and heating oven with a +5 to 300 °C working range, 1.05 kW single-phase heating and 8 air changes per hour. Within the ED series it is the smallest standard chamber, distinguished by 30 kg per rack and 70 kg total load capacity. It serves precise tempering, drying and heat treatment of non-flammable solids, powders and glassware in laboratory and production processes.
BINDER ED Series — Model Comparison & Selection
Full specifications
| Temperature | 5...300 °C |
| Air change rate | 8 x/h |
| Heating time to 150°C | 50 min |
| Interior volume | 57 L |
| Load per rack | 30 kg (ED056-230V / ED056UL-120V) |
| Nominal power | 1.05 kW |
| Permitted load | 70 kg (ED056-230V / ED056UL-120V) |
| Temperature fluctuation | 0.4 K |
| Temperature uniformity | 2.3 K |
| Rear wall clearance | 160 mm (ED056-230V / ED056UL-120V) |
| Side wall clearance | 100 mm (ED056-230V / ED056UL-120V) |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 150°C mfr | 19 min |
| Depth | 565 mm (ED056-230V / ED056UL-120V) |
| Height | 625 mm (ED056-230V / ED056UL-120V) |
| Interior width | 360 mm (ED056-230V / ED056UL-120V) |
| Width | 560 mm (ED056-230V / ED056UL-120V) |
| Net weight of the unit (empty) mfr | 39 kg |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Unit fuse | 6.3 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- The ED 56 uses natural convection (no fan) and a stainless steel 1.4301 (AISI 304) inner chamber, pre-heating chamber and door interior.
- Interior dimensions are 360 mm wide × 625 mm high × 565 mm deep (57 L); external width 560 mm, net weight 39 kg.
- Within the ED series (siblings ED 23, ED 115, ED 260, ED 400) it is the smallest standard size, sharing the +5 to 300 °C range but with lower nominal power (1.05 kW) and a 6.3 A external chamber fuse.
- The controller displays temperature with 1 °C resolution (ED/FD/FED series), unlike the 0.1 °C resolution of BD/BF incubators.
Behaviour under load and in operation
- Temperature fluctuation is 0.4 K and temperature uniformity 2.3 K, measured under factory conditions.
- Heating from ambient to 150 °C takes 50 min; recovery after a 30 s door opening at 150 °C takes 19 min.
- The air-change rate is 8 x/h, supporting drying and moisture removal without forced airflow.
- Permitted load is 70 kg total, with a maximum of 30 kg per rack; the usable volume must not be loaded by more than half, and samples must keep at least 10 % of interior dimensions as wall clearance.
- Above 150 °C, metallic surfaces may discolour due to oxidation; this is harmless and does not impair function.
Installation, connection and control
- Two standard variants: ED056-230V (230 V 1~ 50/60 Hz, CEE 7/7 plug, Art. 9010-0333) and ED056UL-120V (120 V 1~ 60 Hz, NEMA 5-15 plug, Art. 9010-0334).
- Nominal power 1.05 kW; external chamber fuse 6.3 A time-lag (T) 5×20 mm / 250 V; unit fuse 6.3 A.
- The chamber uses an IEC connector plug with shockproof plug; only original BINDER connection cables should be used.
- The microprocessor temperature controller supports set-point entry with or without ramp, timer/continuous operation, and °C/°F change; an RS 422 interface is available as an option for APT-COM 4 software.
- Temperature safety device class 2 (DIN 12880) is standard; the safety controller set-point range is 10 °C to 300 °C (1 °C accuracy) and should be set 2–5 °C above the temperature set-point.
- Overtemperature protective device class 1 trips at 330 °C.
- Installation requires 160 mm rear clearance and 100 mm side clearance; ambient temperature +18 to +40 °C, humidity ≤70 % r.h. non-condensing, altitude ≤2000 m, supply fluctuation ±10 %.
- Housing protection IP 20; pollution degree 2, over-voltage category II (IEC 61010-1).
Integration & Operating Notes
Insufficient clearance prevents heat dissipation and can cause overheating or invalid temperature uniformity. Install on a non-flammable surface in a ventilated room, never in an unventilated recess, and keep the specified distances to walls and adjacent equipment.
The 230 V variant requires a 230 V 1~ 50/60 Hz supply with a 6.3 A time-lag (T) 5×20 mm / 250 V external chamber fuse. Verify supply quality before connection; fluctuations beyond ±10 % may prevent the chamber from meeting its performance and safety limits.
The safety controller set-point is an absolute maximum limit temperature. After every change of the temperature set-point, re-check and adjust the safety controller set-point to 2–5 °C above it; the permissible range is 10 °C to 300 °C with 1 °C accuracy.
Maintenance must be performed at least once a year by licensed electricians or experts authorized by BINDER. Include repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 before initial startup and after maintenance or repair.
Replacing the door gasket while the chamber is warm can damage the gasket. Allow the chamber to cool to ambient temperature before replacement; use only original BINDER spare parts.
The chamber is intended for harmless materials only. Charging material must not be explosive or flammable, must not form an explosive mixture with air, and the operating temperature must lie below its flash point or sublimation point.
The ED series can be hot-air sterilized at 190 °C / 374 °F for at least 30 minutes. Remove all inflammable goods from the interior before sterilization. Racks can be removed and autoclaved.
Selection & Integration Q&A
Choose the ED 56 for small loads up to 70 kg total and 30 kg per rack in a 57 L chamber. Step up to ED 115 or ED 260 when you need more usable volume or higher heating power (1.25 kW or 2.25 kW); all share the +5 to 300 °C range and natural convection.
The 230 V variant (ED056-230V) uses a CEE 7/7 plug on a 230 V 1~ 50/60 Hz supply with a 6.3 A time-lag (T) 5×20 mm / 250 V external chamber fuse. The UL variant (ED056UL-120V) uses a NEMA 5-15 plug at 120 V 1~ 60 Hz. Verify ±10 % voltage tolerance.
Install indoors on a non-flammable surface with 160 mm rear and 100 mm side clearance, ambient temperature +18 to +40 °C, humidity ≤70 % r.h. non-condensing, altitude ≤2000 m. Do not install in unventilated recesses or explosive areas.
Schedule annual maintenance by BINDER-authorized personnel and electrical safety testing per EN 50678/VDE 0701. Recommended consumables: neutral cleaning agent Art. 1002-0016 and disinfectant spray Art. 1002-0022. Optional RS 422 interface for APT-COM 4 and analog temperature output are not included.