Data Sheet Model M 56
Technical Reference Summary
Covered Models & Document Scope
Model M 56 drying and heating chamber with forced convection and advanced program functions; standard version M056-230V, Art.-No. 9010-0371, 230 V 1~ 50/60 Hz, CEE 7/7 plug, 59 L interior volume.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Interior volume | 59 L p.1 ↗ | M 56, 50/60 Hz |
| Rated voltage | 230 V p.2 ↗ | M 56, 50/60 Hz |
| Power frequency | 50/60 Hz p.2 ↗ | M 56 |
| Phase (nominal voltage) | 1~ p.2 ↗ | M 56 |
| Nominal power | 2 kW p.2 ↗ | M 56 |
| Unit fuse | 10,0 A p.2 ↗ | M 56 |
| Temperature range | +23 °C above ambient temperature to 300 °C p.2 ↗ | M 56 |
| Temperature uniformity at 150°C | 1.4 ± K p.2 ↗ | M 56, at 150 °C |
| Temperature fluctuation at 150°C | 0.2 ± K p.2 ↗ | M 56, at 150 °C |
| Heating up time to 150°C | 9 min p.2 ↗ | M 56 |
| Recovery time after door was opened for 30 s at 150°C | 4 min p.2 ↗ | M 56, at 150 °C |
| Air change rate at 100°C | 175 x/h p.2 ↗ | M 56, at 100 °C |
| Net weight of the unit (empty) | 43 kg p.2 ↗ | M 56 |
| Load per rack | 30 kg p.2 ↗ | M 56 |
| Permitted load | 70 kg p.2 ↗ | M 56 |
| Wall clearance back | 160 mm p.2 ↗ | M 56 |
| Wall clearance sidewise | 100 mm p.2 ↗ | M 56 |
Operating Limits
- Temperature range: +23 °C above ambient temperature to 300 °C p.2 ↗
- Permitted load: 70 kg p.2 ↗
- Load per rack: 30 kg p.2 ↗
- Wall clearance back: 160 mm p.2 ↗
- Wall clearance sidewise: 100 mm p.2 ↗
Type Code Decoder
| Position | Token | Meaning |
|---|---|---|
| 1-4 p.1 ↗ | M056 | Model series and size designation (M series, size 56) |
| field 2 p.1 ↗ | 230V | Rated voltage variant 230 V |
Ordering Data
| Ordering Code | Part No. | Variant / Code | Applies To | Note |
|---|---|---|---|---|
| M056-230V p.1 ↗ | 9010-0371 | Standard, 230 V 1~ 50/60 Hz, CEE 7/7 plug | Standard model version |