BINDER M 56
BINDER ·Made in Germany · In Active Production

BINDER M 56

BINDER M 56 — Drying ovens with Temperature 23…300 °C.

The BINDER M 56 is a 59 L forced-convection drying and heating oven with a +23 to 300 °C working range, 2 kW single-phase heating, and 175 air changes per hour. As the smallest chamber in the M series, it delivers 0.2 K fluctuation and 1.4 K uniformity, with 9 min heat-up to 150 °C and 4 min recovery after a 30 s door opening. It is intended for reproducible laboratory drying, heating, and thermal testing where fast air exchange and tight spatial uniformity are required.

Temperature
23…300 °C

BINDER M Series — Model Comparison & Selection

Model Temperature, °CAir change rate, 1/hourDepth, mmHeating up time to 150 c, minHeight, cmInterior volume, L
M 56 Current 23–300175596962.259
M 115 23–3001056361173.2113
M 260 14–300657861396.2256
M 720 14–3004081614159735

Full specifications

Core Performance & Functional Specifications
Temperature 23...300 °C
Air change rate 175 x/h
Heating time to 150°C 9 min
Interior volume 59 L
Load per rack 30 kg
Nominal power 2 kW
Permitted load 70 kg
Temperature fluctuation 0.2 K
Temperature uniformity 1.4 K
Rear wall clearance 160 mm
Side wall clearance 100 mm
Article Number mfr 9010
Recovery time after door was opened for 30 s at 150°C mfr 4 min
Dimensions, Weight & Mechanical Design
Depth 596 mm
Height 622 mm
Width 560 mm
Net weight of the unit (empty) mfr 43 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Unit fuse 10.0 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Thermal behavior

Control, installation, and operation

Integration & Operating Notes

Mounting
Maintain 160 mm rear and 100 mm side clearance

The chamber must not be installed in unventilated recesses; prescribed minimum distances ensure heat dispersal and prevent overheating. Verify these clearances before finalizing bench or cabinet layout, and keep the rear panel accessible for the SUB-D alarm/analog sockets.

Operating Manual M (E3.1), chap. 1.8 and chap. 3.4 Installation
Electrical
Size supply for 230 V 1~ 50 Hz, 2 kW, 10 A fuse

The M 56 draws 2 kW at 230 V single-phase with a 10.0 A unit fuse. Confirm the branch circuit is dedicated and protected accordingly; the CEE 7/7 plug is supplied. Voltage must remain within tolerance at the indicated power frequency to avoid controller faults and heating underperformance.

Data Sheet Model M 56 en.pdf; Operating Manual M (E3.1), chap. 3.4 Electrical Connection
Process Limits
Charging material must stay below flash point

Do not introduce substances that are combustible or explosive at working temperature; the chamber temperature must lie below the flash point or sublimation point of the load. The chamber has no explosion-protection measures and must not be operated in potentially explosive areas or with combustible dust/air-solvent mixtures nearby.

Operating Manual M (E3.1), chap. 1.8 Safety
Control & I/O
Wire zero-voltage alarm contact for external monitoring

The rear SUB-D 'ALARM / CONTACT' socket provides a zero-voltage relay alarm contact that can be integrated into a site alarm or interlock circuit. The optional 'ANALOG OUTPUT' SUB-D socket provides analog temperature output for recording; plan cabling and signal conditioning before commissioning.

Operating Manual M (E3.1), chap. 2.2 and chap. 2.5 Rear Interfaces
Maintenance
Annual maintenance by BINDER-authorized personnel required

Maintenance must be performed at least once a year by licensed electricians or experts authorized by BINDER; warranty is void if non-authorized personnel perform the work. Annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is also required, plus testing before initial startup and after any maintenance or repair.

Operating Manual M (E3.1), chap. 20.1 and chap. 20.2 Maintenance
Maintenance
Reset class 1 overtemperature trip by 10 s power disconnect

If the class 1 overtemperature protective device has turned off the chamber, disconnect it from the power supply for at least 10 seconds and let it cool down. If the device responds again, contact BINDER service; do not repeatedly reset without investigating the cause.

Operating Manual M (E3.1), chap. 20.3 Troubleshooting

Selection & Integration Q&A

When should I choose the M 56 over the larger M 115, M 260, or M 720?

Pick the M 56 for 59 L loads on a 230 V single-phase, 2 kW supply with 10 A protection. Choose M 115 for more volume on the same 230 V supply; M 260/M 720 require 400 V three-phase and offer larger interiors with rear main power switches.

What supply and protection does the M 56 require, and can it share a circuit?

It requires 230 V 1~ 50 Hz, 2 kW, with a 10.0 A unit fuse and CEE 7/7 plug. Size a dedicated branch circuit; sharing with high-inrush loads may cause voltage dips that disturb the MB2 controller and heating performance.

What are the installation clearances and environmental limits for the M 56?

Maintain 160 mm rear and 100 mm side clearance, install splash-proof per IP 20, and avoid unventilated recesses and explosive atmospheres. Do not operate in damp areas or puddles; ambient conditions per manual chap. 3.4 apply.

What maintenance and accessories are required beyond the base chamber?

Annual maintenance by BINDER-authorized personnel and annual electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 are required. Optional accessories include APT-COM 4 software, Pt100 object sensor, analog output, HEPA fresh-air filter, and inert-gas/gas-tight version; these are not included in base delivery.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model M 56 en
Datasheet ·504,4 КБ ·7 pp.
View PDF
PDF
M (E3.1) 06 2026 en
Manual ·11,0 МБ ·137 pp.
View PDF