BINDER M 56
BINDER M 56 — Drying ovens with Temperature 23…300 °C.
The BINDER M 56 is a 59 L forced-convection drying and heating oven with a +23 to 300 °C working range, 2 kW single-phase heating, and 175 air changes per hour. As the smallest chamber in the M series, it delivers 0.2 K fluctuation and 1.4 K uniformity, with 9 min heat-up to 150 °C and 4 min recovery after a 30 s door opening. It is intended for reproducible laboratory drying, heating, and thermal testing where fast air exchange and tight spatial uniformity are required.
BINDER M Series — Model Comparison & Selection
Full specifications
| Temperature | 23...300 °C |
| Air change rate | 175 x/h |
| Heating time to 150°C | 9 min |
| Interior volume | 59 L |
| Load per rack | 30 kg |
| Nominal power | 2 kW |
| Permitted load | 70 kg |
| Temperature fluctuation | 0.2 K |
| Temperature uniformity | 1.4 K |
| Rear wall clearance | 160 mm |
| Side wall clearance | 100 mm |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 150°C mfr | 4 min |
| Depth | 596 mm |
| Height | 622 mm |
| Width | 560 mm |
| Net weight of the unit (empty) mfr | 43 kg |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Unit fuse | 10.0 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- The M 56 is the 59 L entry size of the BINDER M series; siblings M 115, M 260, and M 720 extend interior volume and, for M 260/M 720, use 400 V three-phase supply and a rear main power switch.
- All M-series chambers share forced convection, the MB2 microprocessor controller, and a selectable DIN 12880 class 2/3.1 overtemperature safety device; the M 56 is distinguished by its 230 V 1~ 50 Hz, 2 kW electrical configuration and 10.0 A unit fuse.
- Interior dimensions are 560 mm W × 622 mm H × 596 mm D (net), with a 70 kg total permitted load and 30 kg per rack.
Thermal behavior
- The chamber reaches 150 °C in 9 min from ambient and recovers within 4 min after a 30 s door opening at 150 °C, supported by a 175 x/h air-change rate.
- Temperature fluctuation is 0.2 K and spatial uniformity 1.4 K across the working range; the air flap position is adjustable via the controller to balance fresh-air exchange against heat-up rate.
- The inner surfaces, exhaust duct, door gaskets, and optional access ports reach near set-point temperature during operation and must not be touched.
Control, installation, and operation
- The MB2 controller supports fixed-value operation, timer/stopwatch programs, time programs, and week programs with ramp/step set-point entry, tolerance ranges, and section repetition.
- Rear interfaces include a SUB-D 'ALARM / CONTACT' zero-voltage relay alarm contact and an optional SUB-D 'ANALOG OUTPUT' for temperature recording; Ethernet and USB are available for data transfer and APT-COM 4 networking.
- Installation requires 160 mm rear and 100 mm side wall clearance, a splash-proof location with IP 20 protection, and exclusion of unventilated recesses and explosive atmospheres.
Integration & Operating Notes
The chamber must not be installed in unventilated recesses; prescribed minimum distances ensure heat dispersal and prevent overheating. Verify these clearances before finalizing bench or cabinet layout, and keep the rear panel accessible for the SUB-D alarm/analog sockets.
The M 56 draws 2 kW at 230 V single-phase with a 10.0 A unit fuse. Confirm the branch circuit is dedicated and protected accordingly; the CEE 7/7 plug is supplied. Voltage must remain within tolerance at the indicated power frequency to avoid controller faults and heating underperformance.
Do not introduce substances that are combustible or explosive at working temperature; the chamber temperature must lie below the flash point or sublimation point of the load. The chamber has no explosion-protection measures and must not be operated in potentially explosive areas or with combustible dust/air-solvent mixtures nearby.
The rear SUB-D 'ALARM / CONTACT' socket provides a zero-voltage relay alarm contact that can be integrated into a site alarm or interlock circuit. The optional 'ANALOG OUTPUT' SUB-D socket provides analog temperature output for recording; plan cabling and signal conditioning before commissioning.
Maintenance must be performed at least once a year by licensed electricians or experts authorized by BINDER; warranty is void if non-authorized personnel perform the work. Annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is also required, plus testing before initial startup and after any maintenance or repair.
If the class 1 overtemperature protective device has turned off the chamber, disconnect it from the power supply for at least 10 seconds and let it cool down. If the device responds again, contact BINDER service; do not repeatedly reset without investigating the cause.
Selection & Integration Q&A
Pick the M 56 for 59 L loads on a 230 V single-phase, 2 kW supply with 10 A protection. Choose M 115 for more volume on the same 230 V supply; M 260/M 720 require 400 V three-phase and offer larger interiors with rear main power switches.
It requires 230 V 1~ 50 Hz, 2 kW, with a 10.0 A unit fuse and CEE 7/7 plug. Size a dedicated branch circuit; sharing with high-inrush loads may cause voltage dips that disturb the MB2 controller and heating performance.
Maintain 160 mm rear and 100 mm side clearance, install splash-proof per IP 20, and avoid unventilated recesses and explosive atmospheres. Do not operate in damp areas or puddles; ambient conditions per manual chap. 3.4 apply.
Annual maintenance by BINDER-authorized personnel and annual electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 are required. Optional accessories include APT-COM 4 software, Pt100 object sensor, analog output, HEPA fresh-air filter, and inert-gas/gas-tight version; these are not included in base delivery.