Data Sheet Model MKT 240
Technical Reference Summary
Covered Models & Document Scope
BINDER MKT 240 dynamic climate chamber for rapid temperature changes with extended low temperature range, 228 L interior volume. Two model versions: MKT240-400V (standard, 400 V 3~/N/PE 50 Hz, Art.-No. 9020-0386) and MKT240-480V-C (with voltage and frequency converter, 480 V 3~ 60 Hz, Art.-No. 9020-0364). Document also lists accessories and services.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Interior volume | 228 L p.1 ↗ | MKT240-400V, 50 Hz |
| Temperature range | -70...180 °C p.2 ↗ | MKT240-400V, 50 Hz |
| Temperature uniformity dependent on set value | 0.1...1.0 ± K p.2 ↗ | MKT240-400V, 50 Hz |
| Temperature fluctuation dependent on set value | 0.1...0.4 ± K p.2 ↗ | MKT240-400V, 50 Hz |
| Cooling down time from 180°C to -70°C | 95 min p.2 ↗ | MKT240-400V, 50 Hz |
| Rated voltage | 400 V p.2 ↗ | MKT240-400V, 50 Hz |
| Power frequency | 50 Hz p.2 ↗ | MKT240-400V |
| Nominal power | 6,5 kW p.2 ↗ | MKT240-400V, 50 Hz |
| Unit fuse | 16,0 A p.2 ↗ | MKT240-400V, 50 Hz |
| Phase (nominal voltage) | 3~/N/PE p.2 ↗ | MKT240-400V, 50 Hz |
| Net weight of the unit (empty) | 380 kg p.2 ↗ | MKT240-400V, 50 Hz |
| Load per rack | 30 kg p.2 ↗ | MKT240-400V, 50 Hz |
| Permitted load | 70 kg p.2 ↗ | MKT240-400V, 50 Hz |
| Wall clearance back | 300 mm p.2 ↗ | MKT240-400V, 50 Hz |
| Wall clearance sidewise | 200 mm p.2 ↗ | MKT240-400V, 50 Hz |
| Width net | 1,115 mm p.2 ↗ | MKT240-400V, 50 Hz |
| Height net | 1,935 mm p.2 ↗ | MKT240-400V, 50 Hz |
| Depth net | 925 mm p.2 ↗ | MKT240-400V, 50 Hz |
| Interior width | 735 mm p.3 ↗ | MKT240-400V, 50 Hz |
| Interior height | 700 mm p.3 ↗ | MKT240-400V, 50 Hz |
| Interior depth | 443 mm p.3 ↗ | MKT240-400V, 50 Hz |
| Unit doors | 1 p.3 ↗ | MKT240-400V, 50 Hz |
| Sound-pressure level | 64 dB(A) p.3 ↗ | MKT240-400V, 50 Hz |
| Number of shelves (std./max.) | 1/6 p.3 ↗ | MKT240-400V, 50 Hz |
| Sound-pressure level | 67 dB(A) p.3 ↗ | MKT240-480V-C, 60 Hz |
Operating Limits
- Temperature range: -70...180 °C p.2 ↗
- Wall clearance back: 300 mm p.2 ↗
- Wall clearance sidewise: 200 mm p.2 ↗
- Permitted load: 70 kg p.2 ↗
- Load per rack: 30 kg p.2 ↗
Type Code Decoder
| Position | Token | Meaning |
|---|---|---|
| 1-6 p.1 ↗ | MKT240 | Model series and size: MKT dynamic climate chamber, 240 size (228 L) |
| suffix p.1 ↗ | -400V | Standard model, 400 V 3~/N/PE 50 Hz supply |
| suffix p.1 ↗ | -480V-C | Model with voltage and frequency converter, 480 V 3~ 60 Hz supply |
Ordering Data
| Ordering Code | Part No. | Variant / Code | Applies To | Note |
|---|---|---|---|---|
| MKT240-400V p.1 ↗ | 9020-0386 | Standard, 400 V 3~/N/PE 50 Hz | CEE 16 / 6 H plug 5-pin | |
| MKT240-480V-C p.1 ↗ | 9020-0364 | with voltage and frequency converter, 480 V 3~ 60 Hz | CEE 16 / 6 H plug 5-pin |