BINDER MKT 240
BINDER MKT 240 — Climate chambers with Temperature -70…180 °C.
The BINDER MKT 240 is a 228 L alternating climate chamber with a -70 to 180 °C working range, using a two-stage cascade refrigeration system (R452a/R23) for rapid temperature changes. It is the mid-size MKT variant, positioned between the MKT 115 and MKT 720, and is distinguished by its 400 V three-phase supply and 6.5 kW nominal power. It is intended for dynamic temperature cycling of components and materials in laboratory and industrial test environments.
BINDER MKT Series — Model Comparison & Selection
Full specifications
| Temperature | -70...180 °C |
| Interior volume | 228 L |
| Load per rack | 30 kg (MKT240-400V, 50 Hz) |
| Nominal power | 6.5 kW |
| Article Number mfr | 9020 |
| Depth | 925 mm (MKT240-400V, 50 Hz) |
| Height | 1935 mm (MKT240-400V, 50 Hz) |
| Interior height | 700 mm (MKT240-400V, 50 Hz) |
| Interior width | 735 mm (MKT240-400V, 50 Hz) |
| Width | 1115 mm (MKT240-400V, 50 Hz) |
| Net weight of the unit (empty) mfr | 380 kg |
| Power frequency | 50 Hz |
| Rated voltage | 400 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 3 N/PE |
| Cooling down time from 180°C to -70°C mfr | 95 min |
| Average heating-up rate according to IEC 60068-3-5 between -40°C and 180°C mfr | 5.0 K/min |
Description
Construction and series positioning
- The MKT 240 is a floor-standing chamber with 228 L interior volume (735 mm W × 700 mm H × 925 mm D net) and a net weight of 380 kg. It uses a two-stage cascade refrigeration circuit with R452a (2.20 kg) and R23 (0.38 kg) to reach -70 °C.
- Within the MKT series, the 240 sits between the MKT 115 and MKT 720. All share the -70 to 180 °C range, but the 240 is the mid-size unit with a 400 V 3~/N/PE 50 Hz supply and 6.5 kW nominal power. The -C suffix variant (MKT240-400V-C) adds a voltage and frequency changer for 480 V 3~ 60 Hz operation.
- The chamber is equipped with an Ethernet serial interface and a rear power switch (from size 115 upward). The MB2 controller provides program control, chart recorder display, and alarm functions.
Behavior under load and in different modes
- Average heating rate is 5.0 K/min between -40 °C and 180 °C (per IEC 60068-3-5); average cooling rate is 4.2 K/min between 180 °C and -40 °C. Cooling down from 180 °C to -70 °C takes 95 minutes.
- Temperature uniformity is 0.1...1.0 K and fluctuation 0.1...0.4 K, both dependent on set value. The chamber supports fixed value mode, time programs, and week programs with set-point ramps and steps.
- The over temperature protective device class 1 trips if the nominal temperature is exceeded by 20 °C due to chamber failure; a DIN 12880 class 2.0 temperature safety device is fitted as standard. An optional over/under temperature safety device class 2 is available.
Installation, connection, and operation
- The chamber requires a 400 V 3~/N/PE 50 Hz supply with 16.0 A unit fuse and 11.3 A nominal current. It must not be installed in unventilated recesses; minimum distances per chapter 3.4 of the manual must be observed for heat dispersal.
- The chamber must not be operated in potentially explosive areas or with combustible dust/air-solvent mixtures. It must not become wet during operation, cleaning, or maintenance.
- The MB2 controller supports set-point entry via the Setpoints menu or normal display, program management, alarm reset, and user management with password protection. Network configuration and APT-COM™ 4 software (accessory) enable remote monitoring and documentation.
Integration & Operating Notes
The chamber draws 11.3 A nominal current at 400 V 3~/N/PE 50 Hz and must be protected by a 16.0 A unit fuse. Verify that the site supply can deliver 6.5 kW nominal power and that the over-current protection is sized accordingly; a voltage deviation beyond ±10% may affect performance and safety.
The chamber must not be installed in unventilated recesses. Sufficient ventilation and the minimum distances specified in chapter 3.4 of the operating manual are required to disperse heat from the refrigeration system and avoid overheating damage.
The chamber has no explosion protection and must not be operated in potentially explosive areas or near combustible dust/air-solvent mixtures. Water ingress creates a deadly electric shock hazard; keep electrical parts dry during operation, cleaning, and maintenance.
The over temperature protective device class 1 responds if the nominal temperature is exceeded by 20 °C due to chamber failure. A class 1 trip requires the chamber to be taken out of operation and BINDER Service informed; do not reset and continue without investigating the cause.
A maintenance message appears after 8760 operating hours or two years. After confirmation, the message window reappears every two weeks until reset by BINDER Service. Schedule service intervention at or before this interval to reset the counter and avoid recurring messages.
An annual repeat electrical inspection is required, as well as a test prior to initial startup and prior to resuming operation after maintenance or repair. Testing under EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended. Document the results.
Simple troubleshooting by operating personnel does not require technical intervention into the chamber or disassembly of chamber parts. Detailed troubleshooting and repair must be performed by BINDER Service or BINDER-qualified service partners per the Service Manual.
Selection & Integration Q&A
Choose the MKT 240 when you need a 228 L interior volume and a -70 to 180 °C range with 5.0 K/min heating and 4.2 K/min cooling. The MKT 115 offers smaller capacity, while the MKT 720 provides larger capacity; all share the same temperature range. Select based on sample size and required throughput.
Install the chamber with the minimum distances specified in chapter 3.4 of the manual to allow heat dispersal; do not install in unventilated recesses. Avoid potentially explosive areas, combustible dust/air-solvent mixtures, and damp locations. The chamber must not become wet during operation or cleaning.
A maintenance message appears after 8760 operating hours or two years, requiring BINDER Service to reset. Annual electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702 is mandatory. Base delivery excludes options such as RS485 interface, analog temperature outputs, Pt 100 object sensor, compressed air connection/dryer, and APT-COM™ 4 software.