Data Sheet Model UF V 500
Technical Reference Summary
Covered Models & Document Scope
BINDER UF V 500 ultra-low temperature freezer family: UFV500-230V (standard, 230 V 1~ 50 Hz), UFV500-230V-W (water cooling, 230 V 1~ 50 Hz), UFV500UL-120V (115 V 1~ 60 Hz), UFV500UL-240V (208...230 V 2~ 60 Hz); 501 L interior volume
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Temperature range | -90...-40 °C p.2 ↗ | UFV500-230V, 50 Hz |
| Temperature uniformity at -80°C | 2.5 ± K p.2 ↗ | UFV500-230V, 50 Hz |
| Temperature fluctuation at -80°C | 1.5 ± K p.2 ↗ | UFV500-230V, 50 Hz |
| Cooling down time from 22°C to -80°C | 360 min p.2 ↗ | UFV500-230V, 50 Hz |
| Rated voltage | 230 V p.2 ↗ | UFV500-230V, 50 Hz |
| Power frequency | 50 Hz p.2 ↗ | UFV500-230V |
| Nominal power | 1,6 kW p.2 ↗ | UFV500-230V, 50 Hz |
| Unit fuse | 10,0 A p.2 ↗ | UFV500-230V, 50 Hz |
| Phase (nominal voltage) | 1~ p.2 ↗ | UFV500-230V, 50 Hz |
| Interior volume | 501 L p.2 ↗ | UFV500-230V, 50 Hz |
| Net weight of the unit (empty) | 247 kg p.2 ↗ | UFV500-230V, 50 Hz |
| Permitted load | 200 kg p.2 ↗ | UFV500-230V, 50 Hz |
| Wall clearance back | 100 mm p.2 ↗ | UFV500-230V, 50 Hz |
| Wall clearance sidewise | 250 mm p.2 ↗ | UFV500-230V, 50 Hz |
| Width net | 826 mm p.2 ↗ | UFV500-230V, 50 Hz |
| Height net | 1,966 mm p.2 ↗ | UFV500-230V, 50 Hz |
| Depth net | 938 mm p.2 ↗ | UFV500-230V, 50 Hz |
| Interior width | 606 mm p.2 ↗ | UFV500-230V, 50 Hz |
| Interior height | 1,300 mm p.2 ↗ | UFV500-230V, 50 Hz |
| Interior depth | 636 mm p.3 ↗ | UFV500-230V, 50 Hz |
| Inner doors | 2 p.3 ↗ | UFV500-230V, 50 Hz |
| Unit doors | 1 p.3 ↗ | UFV500-230V, 50 Hz |
| Sound-pressure level | 47 dB(A) p.3 ↗ | UFV500-230V, 50 Hz |
| Number of shelves (std./max.) | 3/11 p.3 ↗ | UFV500-230V, 50 Hz |
| Cryoboxes, 50 mm | 352 p.3 ↗ | UFV500-230V, 50 Hz |
Operating Limits
- Temperature range: -90...-40 °C p.2 ↗
- Permitted load: 200 kg p.2 ↗
- Wall clearance back: 100 mm p.2 ↗
- Wall clearance sidewise: 250 mm p.2 ↗
Type Code Decoder
| Position | Token | Meaning |
|---|---|---|
| 1-6 p.1 ↗ | UFV500 | Model family UF V 500 ultra-low temperature freezer, 501 L |
| suffix p.1 ↗ | UL | UL/NEMA variant for 60 Hz markets (absent = European 230 V 50 Hz version) |
| suffix p.1 ↗ | -W | Water cooling option model (absent = standard air-cooled) |
| voltage field p.1 ↗ | 230V | 230 V 1~ 50 Hz, CEE 7/7 plug, 10,0 A fuse |
| voltage field p.1 ↗ | 120V | 115 V 1~ 60 Hz, NEMA 5-15 plug, 15,0 A fuse |
| voltage field p.1 ↗ | 240V | 208...230 V 2~ 60 Hz, NEMA 6-15 plug, 10,0 A fuse |
Ordering Data
| Ordering Code | Part No. | Variant / Code | Applies To | Note |
|---|---|---|---|---|
| UFV500-230V p.1 ↗ | 9020-0347 | 230 V 1~ 50 Hz, standard | CEE 7/7 plug, 10,0 A fuse | |
| UFV500-230V-W p.1 ↗ | 9020-0349 | 230 V 1~ 50 Hz, water cooling | CEE 7/7 plug, 10,0 A fuse | |
| UFV500UL-120V p.1 ↗ | 9020-0351 | 115 V 1~ 60 Hz, standard | NEMA 5-15 plug, 15,0 A fuse | |
| UFV500UL-240V p.1 ↗ | 9020-0353 | 208...230 V 2~ 60 Hz, standard | NEMA 6-15 plug, 10,0 A fuse |