BINDER UF V 500
BINDER UF V 500 — Ultra-low temperature freezers with Temperature -90…-40 °C.
The UF V 500 is a 501 L ultra-low temperature freezer using a two-stage hydrocarbon cascade (R290/R170) to reach -90 °C, positioned between the UF V 350 and UF V 700 in the series. Its 230 V / 50 Hz single-phase supply draws 7.0 A at 1.6 kW nominal, with 1.5 K fluctuation and 2.5 K uniformity at -80 °C. It serves as a high-capacity storage stage for temperature-sensitive samples requiring a DIN 12880 class 3.1 safety controller.
Direct drop-in replacements
BINDER UF V Series — Model Comparison & Selection
Full specifications
| Temperature | -90...-40 °C |
| Interior volume | 501 L (UFV500-230V, 50 Hz) |
| Nominal power | 1.6 kW |
| Temperature fluctuation (-80 °C) | 1.5 K |
| Temperature uniformity (-80 °C) | 2.5 K |
| Article Number mfr | 9020 |
| Depth | 938 mm (UFV500-230V, 50 Hz) |
| Height | 1966 mm (UFV500-230V, 50 Hz) |
| Interior width | 826 mm (UFV500-230V, 50 Hz) |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Cool-down time | 360 min |
Description
Construction and series positioning
- The UF V 500 is the mid-capacity model in the UF V (E3) series, with 501 L interior volume and 826 mm interior width; the UF V 350 and UF V 700 share the same -90…-40 °C range but differ in chamber size.
- It uses a two-stage cascade refrigeration system with R290 (0.150 kg) in stage 1 and R170 (0.150 kg) in stage 2; maximum operating pressure is 28 bar.
- The standard 230 V model (UFV500-230V) is single-phase (1 N~) at 50 Hz; UL variants (115 V/60 Hz and 208–230 V/60 Hz 2~) and a water-cooled H2O version are separate models, not options on this unit.
- The RD4 controller provides set-point entry, password-protected access levels, and a temperature safety device per DIN 12880 class 3.1.
Behavior under load and in operation
- Factory temperature fluctuation at -80 °C is 1.5 K and uniformity 2.5 K; cooling-down time to nominal is 360 min.
- The chamber operates from -90 °C to -40 °C; the safety controller (class 3.1) monitors overtemperature independently of the main control loop.
- Refrigerant stage pressures up to 28 bar require service procedures to respect this limit; hydrocarbon refrigerants demand ventilation and no ignition sources.
Installation, connection and control
- The chamber requires rear-wall spacers for clearance; support feet are specific to the UF V 350 and not used here.
- Electrical connection is 230 V / 50 Hz with ±10 % voltage tolerance; nominal current is 7.0 A.
- The RD4 controller supports Ethernet (MAC/IP/subnet/gateway/DNS), USB data transfer, and an optional DL1 data recorder with settable storage rate.
- Alarm functions include a zero-voltage relay output for external monitoring, door-open alarm delay, and tolerance-range alarm delay.
- Optional CO2 emergency cooling and door access system are available for the UF V 500; these options affect GS and cUL marks as noted in the manual.
Integration & Operating Notes
The UF V 500-230V draws 7.0 A nominal at 1.6 kW; voltage must stay within ±10% of 230 V at 50 Hz. Undersized circuits or unstable supply can trip protection or degrade cooling performance; an advanced voltage booster is available as an option for marginal grids.
The chamber must be installed with spacers to maintain rear-wall distance for ventilation, and IP 20 rating means it cannot be exposed to water or damp areas. Support feet are only for the UF V 350, so leveling for the UF V 500 relies on its own base arrangement.
Maximum operating pressure of the refrigerating system is 28 bar. Service or connection work on the cascade must respect this limit; the system contains flammable R290 and R170, so ventilation and ignition-source control are mandatory during any intervention.
A volt-free relay output is provided for alarm forwarding. Wire it to the site monitoring system so that safety-controller overtemperature, tolerance-range, door-open, and power-failure alarms trigger a remote response; the relay must be integrated with the facility's alarm hierarchy.
Door handle, hinges, and locking counterpart must be cleaned and lubricated with medical vaseline at least once a year. Annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is also required, plus testing after any maintenance or repair.
chambers fitted with it do not bear the GS mark and do not bear the cUL test mark. If certification marks are part of the compliance requirement, this option must be evaluated against the project's acceptance criteria.
Selection & Integration Q&A
Pick the UF V 500 for 501 L interior volume when the UF V 350 is too small and the UF V 700 exceeds space or power budget. All share -90…-40 °C; the UF V 500 uses a 230 V / 50 Hz single-phase supply at 7.0 A, while the UF V 700 may demand more capacity. CO2 emergency cooling and door access are available on the 500 and 700, not the 350.
A 230 V / 50 Hz single-phase circuit rated for at least 7.0 A with ±10% voltage tolerance. For remote monitoring, use the Ethernet interface (configurable IP/DNS) and the zero-voltage relay alarm output; an optional analog output for temperature and a DL1 data recorder are available but not included in base delivery.
Install with rear-wall spacers for ventilation; IP 20 rating prohibits damp areas or water exposure. Ambient must avoid ignition sources due to flammable R290/R170 refrigerants. Support feet are not used on this model (UF V 350 only). Ensure the floor can carry the unit and that the 28 bar refrigerant limit is respected during any service.
Annual door-lock lubrication with medical vaseline and annual electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702. Optional items not in base delivery include CO2 emergency cooling, door access system, battery-buffered alarm, advanced voltage booster, analog output, DL1 data recorder, and storage racks/cryo boxes. A USB stick is needed for data transfer.