BINDER UF V 500

BINDER UF V 500 — Ultra-low temperature freezers with Temperature -90…-40 °C.

The UF V 500 is a 501 L ultra-low temperature freezer using a two-stage hydrocarbon cascade (R290/R170) to reach -90 °C, positioned between the UF V 350 and UF V 700 in the series. Its 230 V / 50 Hz single-phase supply draws 7.0 A at 1.6 kW nominal, with 1.5 K fluctuation and 2.5 K uniformity at -80 °C. It serves as a high-capacity storage stage for temperature-sensitive samples requiring a DIN 12880 class 3.1 safety controller.

Temperature
-90…-40 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER UF V Series — Model Comparison & Selection

Model Temperature, °CCooling down time, minInterior volume, LInterior width, mmTemperature fluctuation at 80 c, KTemperature uniformity at 80 c, K
UF V 500 Current -90–-403605018261.52.5
UF V 350 -90–-4021040270643.5
UF V 700 -90–-4045073611101.52.5

Full specifications

Core Performance & Functional Specifications
Temperature -90...-40 °C
Interior volume 501 L (UFV500-230V, 50 Hz)
Nominal power 1.6 kW
Temperature fluctuation (-80 °C) 1.5 K
Temperature uniformity (-80 °C) 2.5 K
Article Number mfr 9020
Dimensions, Weight & Mechanical Design
Depth 938 mm (UFV500-230V, 50 Hz)
Height 1966 mm (UFV500-230V, 50 Hz)
Interior width 826 mm (UFV500-230V, 50 Hz)
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Operating Conditions, Environment & Protection
Cool-down time 360 min

Description

Construction and series positioning

Behavior under load and in operation

Installation, connection and control

Integration & Operating Notes

Electrical
Supply must deliver 7.0 A at 230 V with ±10% tolerance

The UF V 500-230V draws 7.0 A nominal at 1.6 kW; voltage must stay within ±10% of 230 V at 50 Hz. Undersized circuits or unstable supply can trip protection or degrade cooling performance; an advanced voltage booster is available as an option for marginal grids.

Type plate / p.10; Instruction Manual, Section 4.7 Advanced Voltage Booster
Mounting
Rear-wall spacers required; IP 20 prohibits damp locations

The chamber must be installed with spacers to maintain rear-wall distance for ventilation, and IP 20 rating means it cannot be exposed to water or damp areas. Support feet are only for the UF V 350, so leveling for the UF V 500 relies on its own base arrangement.

Instruction Manual, Section 4.3 Installation and Connections; Type plate / p.11
Process Limits
Refrigerant system pressure must not exceed 28 bar

Maximum operating pressure of the refrigerating system is 28 bar. Service or connection work on the cascade must respect this limit; the system contains flammable R290 and R170, so ventilation and ignition-source control are mandatory during any intervention.

Type plate / p.10-11
Control & I/O
Zero-voltage relay can signal alarms to external systems

A volt-free relay output is provided for alarm forwarding. Wire it to the site monitoring system so that safety-controller overtemperature, tolerance-range, door-open, and power-failure alarms trigger a remote response; the relay must be integrated with the facility's alarm hierarchy.

Instruction Manual, Section 14.5 Alarm Functions
Maintenance
Annual door-lock lubrication and electrical safety test required

Door handle, hinges, and locking counterpart must be cleaned and lubricated with medical vaseline at least once a year. Annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is also required, plus testing after any maintenance or repair.

Instruction Manual, Section 25.2.4 / p.98; Section 25.1 / p.95
Cooling & Media
CO2 emergency cooling option changes certification marks

chambers fitted with it do not bear the GS mark and do not bear the cUL test mark. If certification marks are part of the compliance requirement, this option must be evaluated against the project's acceptance criteria.

Instruction Manual, Section 21 / p.83; Section 1.6 / p.12

Selection & Integration Q&A

When should I choose the UF V 500 instead of the UF V 350 or UF V 700?

Pick the UF V 500 for 501 L interior volume when the UF V 350 is too small and the UF V 700 exceeds space or power budget. All share -90…-40 °C; the UF V 500 uses a 230 V / 50 Hz single-phase supply at 7.0 A, while the UF V 700 may demand more capacity. CO2 emergency cooling and door access are available on the 500 and 700, not the 350.

What supply and monitoring infrastructure does the UF V 500 require?

A 230 V / 50 Hz single-phase circuit rated for at least 7.0 A with ±10% voltage tolerance. For remote monitoring, use the Ethernet interface (configurable IP/DNS) and the zero-voltage relay alarm output; an optional analog output for temperature and a DL1 data recorder are available but not included in base delivery.

What are the mounting and environmental limits for the UF V 500?

Install with rear-wall spacers for ventilation; IP 20 rating prohibits damp areas or water exposure. Ambient must avoid ignition sources due to flammable R290/R170 refrigerants. Support feet are not used on this model (UF V 350 only). Ensure the floor can carry the unit and that the 28 bar refrigerant limit is respected during any service.

What maintenance and accessories are required beyond base delivery?

Annual door-lock lubrication with medical vaseline and annual electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702. Optional items not in base delivery include CO2 emergency cooling, door access system, battery-buffered alarm, advanced voltage booster, analog output, DL1 data recorder, and storage racks/cryo boxes. A USB stick is needed for data transfer.

Dimensional Drawings & Technical Documentation

Dimensional drawings

UF V 500 — drawing 1
performance curve

Click drawing to inspect vector dimensions

Technical documentation

PDF
Data Sheet Model UF V 500 en
Datasheet ·544,6 КБ ·9 pp.
View PDF
PDF
UF V (E3) 09 2026 en
Manual ·10,2 МБ ·127 pp.
View PDF