VDL (E3.1) 05 2026
Technical Reference Summary
Covered Models & Document Scope
This operating manual (VDL (E3.1), issue 05/2026, Art. No. 7001-0384) covers the BINDER VDL series of vacuum drying ovens with microprocessor program controller MB2. Models and variants: VDL 23 (VDL023-230V, Art. No. 9630-0009; VDL023-120V, Art. No. 9630-0013), VDL 56 (VDL053-230V, Art. No. 9630-0010; VDL053-120V, Art. No. 9630-0014), VDL 115 (VDL115-230V, Art. No. 9630-0011; VDL115-120V, Art. No. 9630-0015). The document also covers accessories and options such as the VP5 EX vacuum pump, pump module, connection kits, APT-COM 4 software, analog outputs, and flexible Pt 100 object temperature sensor.
Section-by-Section Technical Breakdown
Describes the physical layout of the VDL chamber, the triangular instrument box housing the MB2 controller, and the connections on the rear of the chamber. It provides the reference for locating interfaces used during installation and operation.
- Triangular instrument box contains the MB2 controller
- Rear of chamber carries supply and process connections
- Area classification information is provided for chamber interior and surroundings
Specifies requirements for the installation site, including general installation requirements, ventilation and extraction for heat removal and process operation, equipotential bonding, ambient conditions, compressed air supply for flushing, fire extinguisher, and lightning protection.
- Technical ventilation required during chamber operation and when emptying the pump condensate catchpot
- Air supply for breaking vacuum when operating with inert gas
- Compressed air supply required for flushing the area for electrical equipment, preheating chamber and controller housing
- Equipotential bonding and lightning protection device required
Covers mechanical and electrical installation: vacuum expansion racks and rack holders, compressed air supply connection, pump module accessory, vacuum connection and vacuum pumps, inert gas supply, tilt protection holders, grounding concept, and electrical connection.
- Vacuum expansion racks and rack holders
- Compressed air supply connection for flushing
- Pump module accessory mounting and equipotential bonding
- Vacuum connection and VP5 EX vacuum pump installation
- Inert gas supply connection and pressure settings
- Equipotential bonding / grounding concept
- Electrical connection
Defines the scope and objective of explosion protection testing before initial commissioning, including verification of the explosion protection plan, implementation of measures, recurring test deadlines, maintenance plan, and tests of technical ventilation and protective devices.
- Functional test scope and explosion protection plan
- Testing before initial commissioning
- Inspection after changes requiring review
- Recurring tests for explosive safety
Describes the MB2 controller display views, icon overview, operating modes, menu structure including main menu, Settings and Service submenus, principle of controller entries, and behavior during and after power failures.
- Normal display, program display and chart-recorder display
- MB2 controller icons and operating modes
- Main menu, Settings submenu and Service submenu
- Performance during and after power failures
Covers requirements for safe commissioning, overview of the drying process, the VDL compressed-air flushing system for the electrical equipment area, preheating chamber and controller housing, standby mode, controller settings at start up, loading, evacuation, condensation avoidance, breaking vacuum, unloading and preparing a new drying process.
- Pre-flush with compressed air before starting up / restarting
- Continuous compressed air flow during chamber operation (pressurized encapsulation)
- Flushing after completion of chamber operation (recommended)
- Evacuation and avoidance of condensation in the interior
- Breaking the vacuum with ambient air or inert gas
- Removing the full condensate catchpot of the pump
Explains entering setpoints through the Setpoints menu and by direct setpoint entry via the Normal display.
- Set-point entry through the Setpoints menu
- Direct setpoint entry via Normal display
Describes special controller functions including the Functions on/off and Control on/off menus, use of the optional universal connection GAS/AIR 2 for ventilation, closing all valves, activating/deactivating temperature and pressure control, drying monitoring, and cyclical rinsing.
- Functions on/off and Control on/off menus
- Optional universal connection GAS/AIR 2 for ventilation
- Activating / deactivating temperature control
- Activating / deactivating pressure control
- Drying monitoring and cyclical rinsing
Covers user management, authorization levels, password protection, login/logout, user change, password assignment and change, deleting passwords, new password assignment for service or admin levels, and activation code.
- User management and authorization levels
- Log in, log out and user change
- Password assignment and password change
- Activation code
Covers selecting menu language, setting date and time, selecting temperature unit, display configuration and touchscreen calibration, event list, service contact page, current operating parameters, and technical chamber information.
- Menu language, date/time and temperature unit
- Display configuration and touchscreen calibration
- Event list and service contact page
- Current operating parameters and technical chamber information
Describes the safety temperature limiter (TL) class 2 and the overtemperature safety controller class 2, including safety controller mode, setting, alarm messages and measures, and function check.
- Safety temperature limiter (TL) class 2
- Overtemperature safety controller class 2
- Safety controller mode and setting
- Message and measures in the state of alarm
- Function check
Covers setting alarm delay times and tolerance ranges, and describes the alarm condition.
- Setting alarm delay times and tolerance ranges
- Alarm condition
Describes information messages, alarm messages, resetting an alarm, activating/deactivating the audible alarm (buzzer), and the test alarm of the safety temperature limiter (TL).
- Information messages and alarm messages
- Resetting an alarm
- Activating / deactivating the audible alarm (buzzer)
- Test alarm of the safety temperature limiter (TL)
Covers the timer program (stopwatch function) and time programs, including starting, cancelling, performance during program delay time, performance after program end, creating programs, program editor, section editor, value entry for program sections, set-point ramp and step modes, weekday, start time, setpoint entry, tolerance range, repeating sections, and saving programs.
- Timer program (stopwatch function)
- Time program creation and management
- Section editor and value entry for program sections
- Set-point ramp and set-point step modes
- Repeating one or several sections within a time program
Covers starting, cancelling and creating week programs, program editor and section management, and value entry for a program section including set-point ramp/step modes, weekday, start time, setpoint entry and special controller functions.
- Starting and cancelling a week program
- Creating a new week program
- Program editor and section editor
- Value entry for a program section
Covers Ethernet configuration and display of MAC address, and E-mail settings.
- Ethernet configuration
- Display of MAC address
Covers using the USB connection during chamber operation for data transfer.
- Using the USB connection during chamber operation
Describes chart recorder views, showing and hiding the legend, history display, and setting the parameters.
- Views and legend
- History display
- Setting the parameters
Covers checking the temperature in the inner chamber, checking the controller display, checking spatial temperature exactitude, and checking the function of the manometer for compressed air flushing.
- Checking the temperature in the inner chamber
- Checking the controller display
- Checking the spatial temperature exactitude
- Checking the function of the manometer for compressed air flushing
Describes available options and accessories: APT-COM 4 Multi Management software, analog outputs for temperature and pressure, and object temperature display with flexible Pt 100 temperature sensor including connection and display on the MB2 controller.
- APT-COM 4 Multi Management software
- Analog outputs for temperature and pressure (option)
- Object temperature display with flexible Pt 100 temperature sensor (option)
Covers safety instructions on cleaning and decontamination, cleaning procedures, and decontamination / chemical disinfection.
- Safety instructions on cleaning and decontamination
- Cleaning
- Decontamination / chemical disinfection
Covers general information and personnel qualifications, simple troubleshooting, maintenance and service including safety instructions and maintenance intervals, service reminder, BINDER service contact data, and sending the chamber back to BINDER GmbH.
- Simple troubleshooting table with fault description, possible cause and required measures
- Maintenance intervals
- Service Reminder
- Electrical testing and explosion protection testing requirements
Covers factory calibration and adjustment, over current protection, VDL technical data, equipment/options/accessories extract, accessories and spare parts extract, and dimensions for VDL 23, VDL 56 and VDL 115.
- Factory calibration and adjustment
- Over current protection
- VDL technical data
- Equipment, options, and accessories (extract)
- Accessories and spare parts (extract)
- Dimensions for VDL 23, VDL 56 and VDL 115
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| The VDL series is available in 230 V and 120 V variants: VDL 23 (9630-0009 / 9630-0013), VDL 56 (9630-0010 / 9630-0014), VDL 115 (9630-0011 / 9630-0015). p.1 ↗ | Select the correct model version and article number for the site supply voltage before ordering. | VDL 23, VDL 56, VDL 115 |
| Type plate example shows nominal temperature 110 °C, 1.60 kW / 7.0 A, 230 V / 60 Hz, IP protection 20, temperature safety device class 2.0 acc. to DIN 12880, temp. class T3. p.13 ↗ | Verify electrical supply and temperature safety class against the type plate of the specific chamber. | VDL series (type plate example) |
| An inlet pressure of 2 bar must be provided for compressed air flushing. p.17 ↗ | The compressed air supply line must be sized and regulated to deliver 2 bar at the chamber inlet. | VDL series |
| The BINDER vacuum pump VP 5 EX is designed for a gas inlet temperature of 40 °C / 104 °F max. p.18 ↗ | Do not exceed 40 °C gas inlet temperature at the pump; confirm pump/vacuum system is designed for the solvent temperature class. | VP5 EX accessory |
| The vacuum pump must be ATEX Directive 2014/34/EU compliant and suitable for suction from Zone 0 or Zone 1. p.18 ↗ | Only use an ATEX-compliant vacuum pump suitable for the zone classification of the application. | VDL series with vacuum pump |
| The chamber must not be installed in potentially explosive areas of Zone 1 or Zone 0. p.15 ↗ | Site the chamber only in a non-hazardous or appropriately classified area per the explosion protection document. | VDL series |
| Technical ventilation (extraction) must include the installation area of the vacuum drying oven and a vacuum pump, and must be provided before commissioning. p.16 ↗ | Design extraction to cover both chamber and pump installation area before commissioning. | VDL series |
| Equipotential bonding must be implemented according to the grounding concept. p.16 ↗ | Ground the chamber and pump module per the grounding plan before commissioning. | VDL series |
| An annual repeat electrical inspection is required, as well as a test prior to initial startup and prior to resuming operation after maintenance or repair; EN 50678/VDE 0701 and EN 50699/VDE 0702 are recommended. p.169 ↗ | Schedule annual electrical testing and post-maintenance testing per the Service Manual. | VDL series |
| Explosion protection testing is required before initial commissioning, before restarting after maintenance or repair, and as recurring tests according to the operator's explosion protection concept. p.169 ↗ | Do not put the chamber into operation if it did not pass the explosion protection test. | VDL series |
| The chamber must not heat up until the pressure threshold of 100 mbar is reached; no heater release occurs below this threshold. p.170 ↗ | Select a suitable pressure setpoint and wait until the pressure threshold is reached before expecting heating. | VDL series |
| The safety controller value must be set appropriately relative to the temperature setpoint; if the inner chamber temperature reaches the safety controller value, the safety controller responds with alarm message 'Safety controller'. p.170 ↗ | Check temperature setpoint and safety controller value settings if the safety controller alarm occurs. | VDL series |
| The safety temperature limiter (TL) class 2 turns off the heater and triggers alarm message 'Overtemperature' when the chamber is defective. p.170 ↗ | Let the chamber cool down, acknowledge the alarm and restart; if the error repeats, turn off the chamber and contact BINDER service. | VDL series |
| The chamber fuse must be checked if the chamber is permanently turned off; supply voltage must be checked for 115 V or 230 V. p.170 ↗ | Verify correct supply voltage and chamber fuse before contacting service. | VDL series |
| The controller must be calibrated and adjusted if the set-point temperature is not reached after the specified time. p.170 ↗ | Perform controller calibration and adjustment as part of troubleshooting for slow heating. | VDL series |
| The door gasket must be replaced if the chamber door is not properly closed or the door gasket is defective. p.170 ↗ | Check door closure and gasket condition when the chamber does not heat up. | VDL series |
| The Pt 100 sensor or heating element or semiconductor relay may be defective if the chamber does not heat up; contact BINDER service. p.170 ↗ | Do not attempt repair; contact BINDER service for defective sensor, heating element or semiconductor relay. | VDL series |
| The operator must provide ESD-protected personal protective equipment; gloves, shoes and other elements must be conductive. p.33 ↗ | Use conductive PPE to avoid electrostatic charging when handling the chamber and solvents. | VDL series |
| The correct auto-ignition temperature of the solvent must be determined and the maximum drying temperature (safety temperature) set according to the information panel 'Temperature setting'; for solvent mixtures, use the solvent with the lowest auto-ignition temperature. p.33 ↗ | Base the safety temperature setting on the solvent data sheet with the lowest auto-ignition temperature. | VDL series |
| The chamber must be tested prior to resuming operation after maintenance; an electrical test and an explosion protection test are required. p.169 ↗ | Plan electrical and explosion protection tests after any maintenance or repair before restarting. | VDL series |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Chamber permanently turned off. p.170 ↗ | No power supply. | Check connection to power supply. | Section 26.2 |
| Chamber permanently turned off. p.170 ↗ | Wrong voltage. | Check power supply for voltage of 115V or 230V. | Section 26.2 |
| Chamber permanently turned off. p.170 ↗ | Chamber fuse has responded. | Check chamber fuse. | Section 26.2 |
| Chamber permanently turned off. p.170 ↗ | Controller defective. | Contact BINDER Service. | Section 26.2 |
| Chamber doesn’t heat up. p.170 ↗ | Pressure threshold of 100 mbar not reached. No heater release. | Wait until pressure threshold is reached and heater released. Select suitable pressure setpoint. | Section 26.2 |
| Chamber doesn’t heat up. p.170 ↗ | Chamber door not properly closed. | Completely close chamber door. | Section 26.2 |
| Chamber doesn’t heat up. p.170 ↗ | Door gasket defective. | Replace door gasket, | Section 26.2 |
| Chamber doesn’t heat up. p.170 ↗ | Pt 100 sensor defective. | Contact BINDER service. | Section 26.2 |
| Chamber doesn’t heat up. p.170 ↗ | Heating element defective. Semiconductor relay defective | Contact BINDER service. | Section 26.2 |
| Chamber doesn’t heat up. Alarm message “Safety controller” on the controller display p.170 ↗ | Safety controller has responded: Inner chamber temperature has reached the safety controller value. Safety controller value set too low or temperature set-point too high, or error in the heating system. | Let the chamber cool down. Acknowledge the alarm on the controller. Check the settings of the temperature setpoint and safety controller value. If appropriate, select suitable safety controller value (chap. 14.2). | Section 26.2 |
| Chamber doesn’t heat up. Alarm message “Safety controller” on the controller display p.170 ↗ | Safety controller defective. | Contact BINDER service. | Section 26.2 |
| Chamber doesn’t heat up. Alarm message “Overtemperature” on the controller display p.170 ↗ | Safety temperature limiter (TL) has turned off the heater. Chamber defective. | Let the chamber cool down. Acknowledge the alarm on the controller and restart the chamber (reconnect power plug). If the error repeats, turn off the chamber and contact BINDER service. | Section 26.2 |
| Set-point temperature is not reached after specified time. p.170 ↗ | Chamber door not properly closed. | Completely close chamber door. | Section 26.2 |
| Set-point temperature is not reached after specified time. p.170 ↗ | Door gasket defective. | Replace door gasket, | Section 26.2 |
| Set-point temperature is not reached after specified time. p.170 ↗ | Controller not adjusted. | Calibrate and adjust controller. | Section 26.2 |