WIC 09 2026

Technical Reference Summary

Covered Models & Document Scope

Covers the BINDER Environmental Walk-in Chambers WIC (E1) for stability testing and storage: WIC 1 (WIC1230V, Art. No. 9800-0011), WIC 2 (WIC2230V, Art. No. 9800-0012) and WIC 3 (WIC3230V, Art. No. 9800-0013). The assembly consists of the chamber with fittings and the conditioner; optional fittings include stainless steel heavy-duty shelves, a trolley ramp, and front-mounted conditioner. Document issue 09/2026, Art. no. 7001-0431.

Section-by-Section Technical Breakdown

pp. 15-16
Device description, sizes and construction

The WIC chambers are walk-in environmental chambers for temperature and climate testing/stability testing of non-hazardous goods and storage under constant climatic conditions. Three sizes are available with different volumes and floor areas. The chamber is built from galvanized, polyester-coated steel panels with 100 mm PU foam insulation, a chrome-nickel steel floor overlay, and a separate conditioner mounted at the rear (front mounting optional).

  • WIC 1: 12.5 m3 volume, 5.5 m2 floor area (plus ramp)
  • WIC 2: 18.5 m3 volume, 8.2 m2 floor area (plus ramp)
  • WIC 3: 24.5 m3 volume, 10.8 m2 floor area (plus ramp)
  • Temperature range 5 °C to 50 °C (41 °F to 122 °F); relative humidity 20 % to 90 % r.h.
  • Chamber panels: galvanized sheet steel, polyester coated, 100 mm / 4 inch PU foam, U value 0.21 W/sqmK
  • Floor overlay: chrome-nickel steel 1.4301, 0.7 mm on 15 mm multiplex board, slip resistance R12 (DIN 51130)
  • Modular design allows installation in 2 hours with 2 persons; WIC 2 and WIC 3 allow door on long or short side
  • Conditioner uses refrigerant with GWP 547; operated with tap water; SmartPad control
pp. 7-8
Type plate, electrical ratings and safety device class

The type plate of the WIC overall assembly is located next to the door below the controller cover. It lists nominal temperature, nominal power and current, voltage/frequency, IP protection, temperature safety device class, refrigerant charge and maximum operating pressure. The example WIC 1 rating is 2.70 kW / 11.1 A at 230 V / 50 Hz or 60 Hz, 1 N~.

  • Nominal temperature 50 °C / 122 °F
  • Nominal power 2.70 kW; nominal current 11.1 A (WIC 1 example)
  • Nominal voltage 230 V / 50 Hz or 230 V / 60 Hz, 1 N~, tolerance ±10 %
  • IP protection 20 according to EN 60529
  • Temperature safety device according to DIN 12880:2007, Class 1 (type plate) / Class 3.1 (example label)
  • Refrigerant R450a, filling weight 0.737 kg; max operating pressure 17 bar
  • Contains fluorinated greenhouse gases covered by the Kyoto Protocol
pp. 9-11
Installation site requirements and permissible floor load

The installation site must meet the ambient conditions in chapter 3.3 and must be sufficiently ventilated to disperse heat. The chamber must not be installed in hazardous locations, damp areas or puddles, and must be splash-proof. Floor loading limits are specified for surface load and wheel load.

  • Surface load: 50,000 N/m2 / 1,000 lbs/sq. ft.
  • Wheel load (rubber wheel): 1000 N/4 cm2 / 224 lbs on a 0.62 sq. inch
  • Wheel load (steel wheel): not passable
  • Do NOT install in unventilated recesses; observe prescribed minimum distances (chap. 3.3)
  • Do NOT install or operate in potentially explosive areas
  • Set up splash-proof; do not install in damp areas or puddles
pp. 17, 24-27
Water supply and drain connections

The conditioner is operated with tap water and requires freshwater and wastewater connections. The freshwater inlet and unit drain inlet are 1/2" female NPT. Suitable water quality, freshwater connection, wastewater connection and draining of the conditioner are described in chapter 4.3.

  • Freshwater inlet: 1/2" female NPT
  • Unit drain inlet: 1/2" female NPT
  • Suitable water quality per chap. 4.3.1
  • Freshwater connection per chap. 4.3.2; wastewater connection per chap. 4.3.3
  • Draining the conditioner per chap. 4.3.4
pp. 16-18, 28-29
Electrical connection and junction box

Electrical connection is made via the junction box, which contains overcurrent protection (FI) and circuit breakers. The junction box is connected to the customer's power supply according to chapter 4.5.1. The main power switch / Emergency Off (EMO) switch and Emergency Stop switches are part of the safety chain.

  • Junction box contains overcurrent protection (FI) and circuit breakers
  • Connection of junction box to customer power supply per chap. 4.5.1
  • Main power switch / Emergency Off (EMO) switch (8)
  • Emergency Stop switch (10) and additional Emergency Stop switch on the conditioner (16)
  • Central grounding connection (5) provided
pp. 30-33
Commissioning, SmartPad setpoints and terminal lock

Commissioning covers water connections, start-up, quick start, the SmartPad display, entering temperature and humidity setpoints, behavior during and after power failures, and locking the SmartPad terminal for set-point entry. The SmartPad is the controller interface for the chamber.

  • Water connections must be established before start (chap. 5.1)
  • Quick start and SmartPad display described in chap. 5.2.1
  • Temperature and humidity setpoints entered via SmartPad (chap. 5.3)
  • Performance during and after power failures per chap. 5.4
  • SmartPad terminal can be locked for set-point entry (chap. 5.5)
pp. 36
Humidification and dehumidification system

Chapter 6 describes the humidification and dehumidification system of the WIC chambers. The system supports the relative humidity range of 20 % to 90 % r.h. and is supplied with tap water via the freshwater connection.

  • Humidity range 20 % to 90 % r.h.
  • System supplied with tap water
  • Detailed description in chapter 6 (p. 36)
pp. 32, 37
Operation after power failure

Chapter 7 describes chamber behavior and required actions after a power failure. Chapter 5.4 additionally covers performance during and after power failures.

  • Behavior after power failure described in chap. 7
  • Performance during and after power failures per chap. 5.4
pp. 14, 37-38
Unit malfunctions, low tub water level, thermal protection and emergency stop

Chapter 8 covers unit malfunctions including low tub water level, the adjustable thermal protection for the interior of the chamber, a malfunctions table, and emergency stop / emergency off. The adjustable thermal protection protects the chamber interior against overtemperature.

  • Low tub water level (chap. 8.1.1)
  • Adjustable thermal protection for the interior of the chamber (chap. 8.1.2)
  • Malfunctions table (chap. 8.1.3)
  • Emergency stop and emergency off (chap. 8.1.4)
  • Door frame heater controlled via separate controller with maximum setpoint 55 °C and temperature limiter
pp. 12, 39-41
Cleaning and decontamination

Chapter 9 covers cleaning and decontamination / chemical disinfection of the chamber. Water must not penetrate the chamber during cleaning, and cleaning while the chamber is turned on is prohibited.

  • Cleaning per chap. 9.1
  • Decontamination / chemical disinfection per chap. 9.2
  • Do NOT spray with water; chamber must not become wet
  • Cleaning activity while the chamber is turned on is a foreseeable misuse
pp. 42-43
Maintenance, service, troubleshooting and electrical testing

Maintenance must be performed at least once a year by licensed electricians or experts authorized by BINDER. Troubleshooting beyond simple checks, repairs, and electrical testing must follow the Service Manual. An annual repeat electrical inspection and tests before initial startup and after maintenance/repair are required, recommended under EN 50678:2020 / EN 50699:2020.

  • Regular maintenance at least once a year
  • Disconnect chamber before maintenance; turn off main power switch / Emergency off (8)
  • Maintenance by non-authorized personnel voids warranty
  • Annual repeat electrical inspection required; test before initial startup and before resuming operation after maintenance or repair
  • Recommended electrical testing under EN 50678:2020 / EN 50699:2020
  • After maintenance, chamber must be tested prior to resuming operation
  • Troubleshooting and repair per Service Manual by BINDER Service or qualified partners
pp. 46-49
Technical description and technical data

Chapter 12 covers factory calibration and adjustment, overcurrent protection, definition of usable volume, technical data, equipment and options, and accessories and spare parts. The technical data table provides the detailed performance and connection values for the WIC series.

  • Factory calibration and adjustment (chap. 12.1)
  • Overcurrent protection (chap. 12.2)
  • Definition of usable volume (chap. 12.3)
  • Technical Data (chap. 12.4)
  • Equipment and options extract (chap. 12.5)
  • Accessories and spare parts extract (chap. 12.6)
pp. 19, 50-52
Dimensions

Chapter 13 provides the dimensional drawings and dimensions of the WIC chambers, including door and ramp arrangements for the different sizes.

  • Dimensional drawings per chap. 13
  • Door on short side or long side for WIC 2 and WIC 3
  • Ramp included in floor area planning

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Temperature range 5 °C to 50 °C (41 °F to 122 °F); relative humidity range 20 % to 90 % r.h. p.2 ↗ Chamber selection must match the required stability test conditions within these limits; ICH Q1A (R2) long-term and accelerated conditions are covered. WIC 1, WIC 2, WIC 3
WIC 1: 12.5 m3 / 5.5 m2; WIC 2: 18.5 m3 / 8.2 m2; WIC 3: 24.5 m3 / 10.8 m2 (volume / floor area plus ramp). p.2 ↗ Size selection is based on required usable volume and floor area; ramp must be included in site planning. WIC 1, WIC 2, WIC 3
Nominal power 2.70 kW and nominal current 11.1 A at 230 V / 50 Hz or 60 Hz, 1 N~, voltage tolerance ±10 %. p.1 ↗ Customer power supply must be sized for at least 11.1 A at 230 V single-phase with ±10 % tolerance; verify local frequency 50 or 60 Hz. WIC 1 (example type plate)
Permissible surface load 50,000 N/m2 (1,000 lbs/sq. ft.); rubber wheel load 1000 N/4 cm2 (224 lbs on 0.62 sq. inch); steel wheels not passable. p.1 ↗ Floor construction and access route must support these loads; steel-wheeled transport equipment must not enter the chamber. WIC series
Maximum stay with door closed: 1 person 20 minutes, 2 people 10 minutes; more than two people with door closed is NOT allowed. p.1 ↗ Operating procedures must limit occupancy and duration; door can be opened from inside at any time even when locked. WIC series
Chamber panels: 100 mm / 4 inch PU foam insulation with U value 0.21 W/sqmK; floor overlay 0.7 mm chrome-nickel steel 1.4301 on 15 mm multiplex, slip resistance R12 (DIN 51130). p.2 ↗ Thermal performance and floor durability are defined; floor is passable for rubber wheels only. WIC series
Refrigerant R450a, filling weight 0.737 kg; maximum operating pressure 17 bar; GWP 547. p.1 ↗ Service and leak checks must account for R450a and the 17 bar maximum operating pressure; refrigerant is fluorinated greenhouse gas covered by Kyoto Protocol. WIC 1 (example type plate)
Temperature safety device according to DIN 12880:2007, Class 1 (type plate) / Class 3.1 (example label). p.1 ↗ Overtemperature protection class must be verified against process risk assessment; adjustable thermal protection for chamber interior is described in chap. 8.1.2. WIC series
Door frame heater controlled via separate controller with maximum setpoint temperature 55 °C, equipped with temperature limiter (temperature fuse via overtemperature protection); two additional temperature fuses in the conditioner. p.1 ↗ Door frame heating is independently limited; overtemperature protection devices must not be bypassed. WIC series
Freshwater inlet and unit drain inlet are 1/2" female NPT. p.2 ↗ Plumbing connections must be prepared with matching 1/2" NPT fittings; water quality per chap. 4.3.1. WIC series
Maintenance at least once a year; annual repeat electrical inspection required, recommended under EN 50678:2020 / EN 50699:2020. p.10 ↗ Maintenance schedule and electrical safety testing must be planned annually; test also required before initial startup and before resuming operation after maintenance or repair. WIC series
Chamber installation time approximately 2 hours with 2 persons due to modular design. p.2 ↗ Site planning can allocate a short installation window; door can be placed on long or short side for WIC 2 and WIC 3. WIC series
Lifting capacity of hoisting equipment must be no less than double the weight of the load being lifted; lift conditioner with 4 persons or mechanical assistance using supplied lifting brackets. p.3 ↗ Rigging plan must use at least 2x safety factor and the provided top lifting brackets; brackets may be removed after final placement. Conditioner
IP protection 20 according to EN 60529. p.1 ↗ Chamber is not protected against water jets; must not become wet during operation, cleaning or maintenance. WIC series
Interior LED lighting has IP66 protection; panels, ventilation ducts, fittings, internal parts and shelves are earthed; door frame heater heating cable protection class 2, double insulation. p.1 ↗ Earthing and double insulation are part of the electrical safety concept; verify grounding during installation and maintenance. WIC series
Optional stainless steel heavy-duty shelves with shelf height 2,007 mm / 8'7"; optional stainless steel trolley ramp. p.2 ↗ Interior fitting selection must account for shelf height and ramp access; options are listed in chap. 2.2 and 12.5. WIC series

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Low tub water level p.37 ↗ Water level in the tub is too low (humidification system water supply). Refer to chapter 8.1.1 for corrective action; check water supply connections per chap. 4.3 and 5.1. Chapter 8.1.1
Adjustable thermal protection for the interior of the chamber p.37 ↗ Interior temperature exceeds the adjustable thermal protection setpoint. Refer to chapter 8.1.2 for corrective action; check thermal protection setting and chamber overtemperature condition. Chapter 8.1.2
Malfunctions p.38 ↗ Various unit malfunctions listed in the malfunctions table. Refer to the malfunctions table in chapter 8.1.3; if errors cannot be identified with simple troubleshooting, contact BINDER Service or qualified service partners per the Service Manual. Chapter 8.1.3; chap. 10.1
Emergency stop and emergency off p.38 ↗ Emergency stop switch (10) or Emergency Stop switch on the conditioner (16) actuated, or main power switch / Emergency Off (EMO) switch (8) turned off. Refer to chapter 8.1.4 for emergency stop and emergency off procedures; disconnect chamber before maintenance work. Chapter 8.1.4