Introduces convenience and throughput benefits of small-volume testing
Overviewpp. 3-4
Describes first compendial small-volume apparatus and workflow
Specpp. 5-6
Sampling ports, media replacement, and hardware details
Specp. 7
Software for controlling dissolution testing
p. 8
Invitation for dissolution testing professionals to get in touch
Engineering Facts
12 facts · every claim linked to a page
Selection
The instrument is a compendial USP Apparatus 7 designed specifically for small-volume dissolution testing of drug eluting stents and other novel dosage forms.
Confirms the intended dosage forms and compendial apparatus type before selecting this system.
Source
“It’s designed for small-volume dissolution testing of drug eluting stents (DES), as well as other novel dosage forms.”
Selection
The system meets all USP Apparatus 7 compendial requirements and operates at significantly lower media volumes than typical for this product testing.
Verifies compendial compliance and low-volume capability, which are key selection criteria.
Source
“It meets all USP Apparatus 7 compendial requirements and operates at significantly lower media volumes than are common in this type of product testing.”
Selection
A choice between 5 mL and 10 mL dissolution cells is available, and with a 5 mL cell as little as 3 mL may be used.
Cell volume determines sample concentration and method feasibility for UV or LC analysis.
Source
“choice between a 5 or 10 mL dissolution cell. With a 5 mL cell, as little as 3 mL may be used”
Operation
Tests can be accelerated by running at up to 55 °C.
Defines the maximum operating temperature available for accelerated methods.
Source
“you can accelerate your workflow by running tests at up to 55 °C.”
Operation
The built-in fluidics module allows partial or complete media replacement at each time point per vessel, with up to five different media in a single method.
Media replacement capability and media count affect method design and fluidics planning.
Source
“media may be partially or completely replaced at each time point for each vessel using a built-in fluidics module. As many as five different media may be used during a single method.”
Performance
The apparatus can simultaneously test up to 12 samples plus a control or standard.
Throughput capacity determines batch sizing and QC/R&D workflow planning.
Source
“can simultaneously test up to 12 samples and a control or standard”
Installation
Heating is provided by a jacket around each dissolution cell, so no water bath is required.
Eliminates water bath utilities and associated maintenance from installation planning.
Source
“The heating jacket around the dissolution cell saves you time and maintenance—with no water bath required.”
Operation
Magnetically coupled agitators control holder reciprocation from 1 to 35 dips per minute.
DPM range must match the required method agitation for the device under test.
Source
“They enable between 1 and 35 dips per minute (DPM) to best match your testing requirements.”
Operation
The integrated autosampler includes 13 individual racks with 2 or 4 mL HPLC sampling vials for up to 36 time points.
Autosampler vial capacity and time-point count constrain method duration and sampling schedule.
Source
“It includes 13 individual racks with either 2 or 4 mL HPLC sampling vials for up to 36 time points.”
Installation
Sampling is performed from the sampling port at the bottom of each dissolution cell, and temperature is independently controlled via an integrated probe at the bottom of each cell.
Bottom sampling port and per-cell temperature probe placement affect cell setup and fluidics routing.
Source
“All sampling is done from the sampling port at the bottom of the dissolution cell.”
Compliance
The Windows-based control software supports 21 CFR Part 11 electronic records with a relational database for result storage, archival, and retrieval.
Part 11 compliance and data archival are mandatory for regulated QC environments.
Source
“The software can store the operating parameters and method data required for electronic records, 21 CFR Part 11 environment.”
Operation
One PC can independently control up to four 400-DS systems.
Multi-system control affects PC/network planning and lab scalability.
Source
“Independent control of up to four dissolution apparatus”