DNA/NGS Fragment Analysis

Kit NameNGS Fragment Kit (1-6000 bp) DNF-473HS NGS Fragment Kit (1-6000 bp) DNF-474Small Fragment Kit DNF-476HS Small Fragment Kit DNF-477
Sizing range100 bp – 6,000 bp100 bp – 6,000 bp50 bp – 1,500 bp50 bp – 1,500 bp
Sizing accuracy± 5%1,2± 5%2± 5%2,5,6± 5%8
Sizing precision2% CV1,22% CV22% CV5,72% CV8
Separation resolution10100 bp – 1,000 bp ≤ 5%; 1,000 bp – 6,000 bp ≤10%100 bp – 1,000 bp ≤ 5%; 1,000 bp – 6,000 bp ≤10%50 bp – 900 bp ≤ 5%; 900 bp – 1,500 bp ≤ 10%50 bp – 900 bp8 ≤ 5%; 900 bp – 1,500 bp8 ≤ 10%
Quantification accuracy± 25%1,3± 25%2,4± 25%5,6± 25%8
Quantification precision15% CV1,315% CV2,415% CV5,615% CV8
Quantification rangeDNA Fragment1: 0.1 ng/µL – 10 ng/µL; DNA Smear3: 5 ng/µL – 100 ng/µLDNA Fragment2: 5 pg/µL – 500 pg/µL; DNA Smear4: 50 pg/µL – 5 ng/µLDNA Fragment5: 0.1 ng/µL – 10 ng/µL; DNA Smear6: 5 ng/µL – 100 ng/µLDNA Fragment8: 5 pg/µL – 500 pg/µL; DNA Smear9: 100 pg/µL – 5 ng/µL
Physical Specifications
Analysis time1050 min50 min45 min40 min
Sample volume2 µl2 µl2 µl2 µl
Guaranteed shelf lifeMinimum 4 monthsMinimum 4 monthsMinimum 4 monthsMinimum 4 months
Kit size500 and 1,000 samples500 and 1,000 samples500 samples500 samples

1 Results using 300 bp and 1,000 bp DNA fragment standards in 1X TE buffer. 2 Results using DNA Ladder in 1X TE buffer. 3 Results using sheared gDNA with smear range 50 bp – 2,000 bp in 1X TE buffer. 4 Results using DNA samples in 1X TE buffer. 5 Results using 400 bp DNA fragment standard in 1X TE buffer. 6 Results using sheared gDNA with smear range from 10 bp – 1,400 bp in 1X TE buffer. 7 Specifications not applicable to the FA 12-Capillary Array Long, 55 cm and FA/ZAG 96-Capillary Array Long, 55 cm. 8 Results using 500 bp DNA fragment standards and DNA Ladder in 1X TE buffer. 9 Results using sheared gDNA with smear range from 10 bp – 1,400 bp in 1X TE buffer with 33-55 Array. 10 Short capillary array, 33 cm.

Large Fragment Analysis

HS Large Fragment 50 kb Kit DNF-464Large Fragment Kit DNF-492
Sizing range75 bp – 48,500 bp75 bp – 20,000 bp
Sizing accuracy± 15% at ≤ 15 kb1; ± 25% at ≥ 15 kb2± 15%4,5
Sizing precision± 10%
Quantification accuracy± 25%3± 25%4
Quantification precision20% CV320% CV4
Quantification rangeDNA Fragment3: 5 pg/μL – 600 pg/μL (Optimal concentration 500 – 600 pg/μL); DNA Smear3: 50 pg/μL – 5 ng/μL (Optimal concentration of 1 ng/μL)DNA Fragment4: 0.1 ng/µL – 10 ng/µL; DNA Smear4: 5 ng/µL – 100 ng/µL
Maximum DNA concentration600 pg/μL per fragment; 5 ng/μL per total sample
Physical Specifications
Analysis time655 min50 min
Sample volume2 µl2 µl
Guaranteed shelf lifeMinimum 4 monthsMinimum 4 months
Kit size500 samples500 samples

1 Results using DNA fragment standards at < 15 kb at 600 pg/µL and DNA smears at 1 ng/µL prepared in 1X TE buffer. 2 Results using DNA Fragment standards at > 15 kb at 600 pg/μL and DNA smears at 1 ng/μL prepared from 1X TE buffer. 3 Results using DNA fragment standards and DNA smears prepared in 1X TE buffer. 4 Results using DNA fragment standards in 1X TE buffer. 5 Results using DNA Ladder in 1X TE buffer. 6 Short capillary array, 33 cm.

Quantitative

Kit NameHS RNA kit (15 nt) DNF-472IVT mRNA Low ConcentrationIVT mRNA Mid Concentration
Sizing range200 nt – 6,000 nt200 nt – 6,000 nt2; 500 nt – 9,000 nt3200 nt – 6,000 nt5; 500 nt – 9,000 nt6
Sizing accuracy± 20%1± 15%4± 15%4
Sizing precision20% CV110% CV45% CV4
Limit of detection (S/N > 3) or sensitivityTotal RNA: 50 pg/μL; Ribo-depleted RNA: 250 pg/μL15 pg/μL4N/A
Input concentration rangeTotal RNA: 50 pg/µL – 5,000 pg/µL; Ribo-depleted RNA: 500 pg/µL – 5,000 pg/µL500 pg/μL – 2,500 pg/μL42,500 pg/μL – 10,000 pg/μL4
Quantitative rangeTotal RNA: 50 pg/µL – 5,000 pg/µL; Ribo-depleted RNA: 500 pg/µL – 5,000 pg/µLN/AN/A
Quantification accuracy± 30%1N/AN/A
Quantification precision20% CV1N/AN/A
Physical Specifications
Analysis time740 min45 min2; 90 min345 min5; 90 min6
Sample volume2 µl2 µl2 µl
Guaranteed shelf lifeMinimum 4 monthsMinimum 4 monthsMinimum 4 months
Kit size500 and 1,000 samples500 and 1,000 samples500 and 1,000 samples

1 Results using RNA Ladder as sample. 2 Analysis with the DNF-472A33 – HS IVT mRNA Low Concentration method. 3 Analysis with Lonza RNA Marker and the DNF-472AE33 – 4kV HS IVT mRNA Extended Low Concentration method. 4 Results using 1,800 nt, 4,300 nt, and 9,000 nt IVT mRNA with no modifications on the 5200 and 5300 Fragment Analyzer instruments with the FA Capillary Array Short, 33 cm. 5 Analysis with the DNF-472B33 – HS IVT mRNA Mid Concentration method. 6 Analysis with Lonza RNA Marker and the DNF-472BE33 – 4kV HS IVT mRNA Extended Mid Concentration method. 7 Short capillary array, 33 cm.

Kit Name

dsDNA 930 Reagent Kit (75-20000 bp) DNF-930dsDNA 935 Reagent Kit (1-1500 bp) DNF-935Plasmid DNA Kit DNF-940Plasmid Topology kit7 5191-6627
Sizing range75 bp – 20,000 bp100 bp – 1,500 bp2,000 bp – 10,000 bp2,000 bp – 10,000 bp
Sizing accuracy± 10%1± 5%1± 10%2N/A
Sizing precision15% CV2% CVN/AN/A
Separation resolution575 bp – 1,500 bp ≤ 10%; 1,500 bp – 20,000 bp ≤ 15%1 bp – 100 bp ≤ 15%; 100 bp – 1,500 bp ≤ 10%N/AN/A
DNA fragment concentration range0.5 ng/µL – 50 ng/µL10.5 ng/µL – 50 ng/µL10.1 ng/µL – 1 ng/µL2,30.1 ng/µL – 1 ng/µL6
Physical Specifications
Analysis time530 min20 min30 min45 min
Sample volume2 µl2 µlVariable4Variable
Guaranteed shelf lifeMinimum 4 monthsMinimum 4 monthsMinimum 4 monthsMinimum 4 months
Kit size1,000 samples1,000 samples500 samples275 samples

1 Results using DNA Ladder or DNA fragment standards initially prepared in 1X TE buffer. 2 Results obtained using plasmid DNA Ladder or supercoiled DNA sample prepared in 1X TE buffer. 3 Final concentration DNA in 1X TE buffer. 4 24 µL final sample volume required; optimal sample concentration range in 1X TE buffer. 5 Short capillary array, 33 cm. 6 Final in well concentration DNA in pDNA Dilution Buffer. 7 Instrument with updated Vent Valve (p/n M5301A) is required to run this assay.

Engineering Facts

12 facts · every claim linked to a page
Selection

Choose the kit by its sizing range, since each kit covers a different fragment size window.

The sizing range determines whether a given DNA or RNA sample can be analyzed at all, so it is the first selection criterion.

Source
“Sizing range”
Selection

The HS NGS Fragment Kit reaches down to 5 pg/µL for fragments, so it is required for very low-input NGS libraries.

Standard kits start at 0.1 ng/µL, so low-concentration libraries need the HS variant to stay in range.

Source
“DNA Fragment2: 5 pg/µL – 500 pg/µL;”
Operation

All listed kits require only a 2 µL sample volume, which constrains the minimum sample amount that must be reserved.

The 2 µL requirement drives sample prep and dilution planning, especially for precious or low-volume samples.

Source
“Sample volume 2 µl”
Maintenance

Every kit has a guaranteed shelf life of at least 4 months, which sets the maximum stock to keep on hand.

Ordering and inventory planning must respect the 4-month minimum shelf life to avoid expired reagents.

Source
“Guaranteed shelf life Minimum 4 months”
Selection

The Genomic DNA 50 kb kits cover up to 60,000 bp, so high-molecular-weight gDNA needs these kits rather than the NGS kits.

Samples above 6,000 bp fall outside the NGS kit range and require the dedicated gDNA 50 kb kit.

Source
“75 bp – 60,000 bp”
Selection

The HS Large Fragment 50 kb Kit has a maximum of 600 pg/µL per fragment, so overloaded samples must be diluted before analysis.

Exceeding the maximum concentration degrades quantification, so the loading must be checked against this limit.

Source
“600 pg/μL per fragment; 5 ng/μL per total sample”
Operation

The Plasmid DNA Kit requires a 24 µL final sample volume, unlike the 2 µL used by most other kits.

This larger volume changes the sample prep protocol and must be planned before running plasmid assays.

Source
“24 µL final sample volume required; optimal sample concentration range in 1X TE buffer”
Installation

The Plasmid Topology kit only runs on an instrument fitted with the updated Vent Valve (p/n M5301A).

Without the updated vent valve the assay cannot be run, so hardware must be verified before selecting this kit.

Source
“Instrument with updated Vent Valve (p/n M5301A) is required to run this assay.”
Selection

The FA 12-Capillary Array Ultrashort (22 cm) is only available for the 5200 Fragment Analyzer system.

Array compatibility depends on the instrument model, so the array must be matched to the installed system.

Source
“The FA 12-Capillary Array Ultrashort, 22 cm is only available for the 5200 Fragment Analyzer system.”
Performance

Some sizing specifications do not apply to the long 55 cm capillary arrays, so results differ on those arrays.

Choosing a long array invalidates part of the published spec, which affects method validation and expectations.

Source
“Specifications not applicable to the FA 12-Capillary Array Long, 55 cm and FA/ZAG 96-Capillary Array Long, 55 cm.”
Operation

Analysis times range from 20 to 90 minutes depending on kit and method, which sets the throughput per run.

Run time drives scheduling and throughput planning for the lab workflow.

Source
“Analysis time5 30 min”
Compliance

The products are for research use only and must not be used in diagnostic procedures.

This restricts the permitted use of the kits and arrays in regulated or clinical settings.

Source
“For Research Use Only. Not for use in diagnostic procedures.”