Agilent Technologies Fragment Analyzer Capillary Arrays
Agilent Technologies Fragment Analyzer Capillary Arrays — Automated electrophoresis.
The Agilent Fragment Analyzer Capillary Arrays are 12-capillary consumable arrays for the 5200/5300/5400 Fragment Analyzer systems, offered in ultrashort, short and long formats. They enable automated electrophoresis with a 22 cm effective length, 2 µL sample consumption and 50 min analysis time, delivering sizing and quantification for DNA/RNA QC.
Agilent Technologies Fragment Analyzer Systems Series — Model Comparison & Selection
| Model | Measurement range | Plate capacity | Average sizing of 2 200 nt ivt mrna, nt | Lower marker size of hs rna kit, nt | Maximum ivt mrna sample size analyzed, nt | Sample volume per well, microliter |
|---|---|---|---|---|---|---|
| Fragment Analyzer Capillary Arrays Current | — | 12 | — | — | — | 2 |
| 5200 Fragment Analyzer | — | — | 2370 | 15 | 6000 | 2 |
| 5300 Fragment Analyzer | — | — | 2370 | 15 | 6000 | — |
| RNA QC, DNA QC, Fragment Analyzer | — | 12 | — | — | — | 2 |
| Ultra-high-throughput fragment analysis 5400 Fragment Analyzer | 0.5 | nine plates | 2370 | 15 | 6000 | — |
Full specifications
| Analysis time | 50 min |
| Analysis Time6 | 55 min |
| Analysis Time7 | 40 min |
| Detector type | Charged Coupled Device |
| Plate capacity | 12 samples |
| Sample Input Volume | 2 µL |
| Sample volume | 2 µl |
| Sample Volume Per Well | 2 µL |
| Separation resolution mfr | 3 bp |
| Dilutions per sample mfr | single single |
| Gel capacity mfr | Two gel gel |
| Input concentration dynamic range mfr | two orders of magnitude magnitude |
| RQN quality score scale mfr | 1...10 |
| Sizing range mfr | 200 nt |
| Sizing accuracy mfr | 20 |
| Sizing precision mfr | 20 |
| Quantification accuracy mfr | 30 |
| Quantification precision mfr | 20 |
| Quantification range (DNA Fragment) mfr | 0.1 ng/μL |
| Quantification range (DNA Smear) mfr | 5 ng/μL |
| Guaranteed shelf life mfr | 4 months |
| Kit size mfr | 500 samples |
| DNA fragment concentration range mfr | 0.5 ng/μL |
| Number of capillaries mfr | 12 capillaries |
| RQN scale mfr | 1...10 |
| Capillary arrays mfr | 12 arrays |
| n–1 resolution mfr | 60 nt |
| Samples separated per run mfr | 12 samples |
| Maximum queued samples mfr | 288 samples |
| Capillary counts mfr | 12 capillaries |
| sample separation time mfr | 30 minutes |
| samples per run mfr | 12 samples |
| capillary array options mfr | 12 arrays |
| resolution range mfr | 60 nt |
| Maximum DNA concentration mfr | 600 sample |
| Limit of detection (S/N > 3) or sensitivity mfr | 50 pg/μL |
| Input concentration range mfr | 50 pg/μL |
| Quantitative range mfr | 50 pg/μL |
| Maximum DNA concentration per fragment mfr | 600 pg/μL |
| Maximum DNA concentration per total sample mfr | 5 ng/μL |
| Final sample volume (Plasmid DNA Kit) mfr | 24 μL |
| Effective Length | 22 cm |
| Length | 47 cm |
| Effective length (FA 12-Capillary Array Ultrashort) mfr | 22 cm |
| Total length (FA 12-Capillary Array Ultrashort) mfr | 47 cm |
| Effective length (FA 12-Capillary Array Short) mfr | 33 cm |
| Total length (FA 12-Capillary Array Short) mfr | 55 cm |
| Effective length (FA 12-Capillary Array Long) mfr | 55 cm |
| Total length (FA 12-Capillary Array Long) mfr | 80 cm |
| Standard Warranty | One-year |
Description
Construction and series positioning
- The arrays are 12-capillary consumables used across the 5200, 5300 and 5400 Fragment Analyzer systems; the instruments differ only in throughput (5200: 12–96 samples one to three times per day; 5300: one to two 96-well plates up to three times per day; 5400: robotic automation, more than two 96-well plates per day).
- Array formats are ultrashort, short and long; the matched specification lists a 22 cm effective length, which corresponds to the separation length of the installed array.
- The array is a replaceable module, not a fixed instrument component; gel capacity is two gels per array.
Performance in operation
- With the 22 cm effective length, the system achieves 100 separation resolution, sizing accuracy of 5 and sizing precision of 2 % CV; quantification accuracy is 25 and precision 15 % CV.
- Sizing range starts at 100 bp; quantification ranges are 0.1 ng/µL for DNA fragments and 5 ng/µL for DNA smears, with a DNA fragment concentration range from 0.5 ng/µL and an input concentration dynamic range of two orders of magnitude.
- Each sample consumes 2 µL; analysis time is 50 min; throughput is 12 samples per day for the 12-capillary configuration.
- The RQN quality score scale runs from 1 to 10; dilutions per sample are single.
Installation and operation
- The arrays are part of the consumable set (FA 12/48/96-capillary arrays) alongside DNA/NGS reagent kits such as DNF-473, DNF-474, DNF-476, DNF-477, DNF-905, DNF-910, DNF-915 and DNF-920.
- Reagents are room-temperature stable with a guaranteed shelf life of 4 months; kit size is 500 samples.
- The detector is a charged coupled device (CCD); no further electrical or interface specifications are provided for the array itself.
Integration & Operating Notes
The 12-capillary arrays are intended for the 5200/5300/5400 Fragment Analyzer systems; installing an array of the wrong capillary count or format (ultrashort, short, long) will not fit the instrument's capillary drive and will invalidate separation length and sizing performance. Confirm the array format against the installed instrument configuration before ordering.
Quantification is specified from 0.1 ng/µL for DNA fragments and 5 ng/µL for DNA smears, with a DNA fragment concentration range starting at 0.5 ng/µL and a two-orders-of-magnitude input dynamic range. Samples outside this window require dilution; only single dilutions per sample are supported, so plan dilution before loading.
Each array supports two gels, and reagent kits are guaranteed for 4 months at room temperature. Track gel count and reagent expiry per array so that separation resolution (100) and sizing precision (2 % CV) are not degraded by exhausted gels or expired reagents.
The array is a passive consumable; the matched specifications list only a CCD detector and no voltage, current or interface ratings for the array itself. Electrical and control requirements are determined by the host Fragment Analyzer instrument, not by the array.
Selection & Integration Q&A
Pick the 12-capillary array when daily demand is 12–96 samples run one to three times per day, matching the 5200 system. For one to two 96-well plates up to three times per day choose the 5300; for more than two plates per day with robotic automation choose the 5400.
The array is supplied separately from the reagent kits; pair it with a DNA/NGS kit such as DNF-473, DNF-474, DNF-476, DNF-477, DNF-905, DNF-910, DNF-915 or DNF-920. Kits cover 500 samples and are room-temperature stable for 4 months.
Each array supports two gels; replace it after the gel capacity is exhausted. Reagent kits carry a guaranteed 4-month shelf life at room temperature, so schedule reagent replacement accordingly to preserve sizing precision of 2 % CV.