Selection
The operating temperature range depends on the cooling method: 28–100 °C without cooling, 15–100 °C with water cooling or refrigerator.
Determines whether the bath can reach sub-ambient or near-ambient setpoints required by the viscometer method.
Source
“Operating temperature range
28...100 °C
with water cooling
15...100 °C
with refrigerator
15...100 °C”
Installation
The unit requires a 200–240 V single- or two-phase supply with a minimum 10 A fuse and draws up to 10 A.
Electrical supply and branch protection must be sized accordingly before installation.
Source
“Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
min. Fuse
10A”
Installation
External branch circuit protection according to UL 489 is required.
Dictates the type of upstream breaker that must be provided in the installation.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Voltage and frequency tolerances are ±10% only if they do not run in opposite directions.
Site power quality must be checked; combined opposite deviations are not permitted.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation
Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.
The correct plug type must be stated on the order to avoid rework at commissioning.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation
Ambient temperature must stay between 5 °C and 40 °C and the enclosure is IP20.
Defines the permissible room conditions and limits protection against water and dust ingress.
Source
“Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Operation
The bath holds 30 l with a minimum filling of 25,5 l and has three bath openings of 90 x 90 mm, 310 mm deep.
Determines the required bath fluid volume and the space available for inserting viscometers.
Source
“Bath volume
30 l
min. filling capacity
25,5 l”
Performance
The pressure pump delivers up to 27 l/min at a maximum pressure of 0,7 bar through an M16x1 male connection.
Needed to match external circuit pressure drop and hose/connector thread.
Source
“max. delivery
27 l/min
max. delivery pressure
0,7 bar
Pump connection
M16x1 male”
Performance
Temperature stability at 70 °C is 0,01 K, supported by a Pt100 internal sensor and an external Pt100 connection.
Confirms the bath meets the tight stability needed for capillary viscometry and density measurement.
Source
“Temperature stability at 70°C
0,01 K”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.
Overtemperature cut-out must be set and verified during commissioning for safe unattended operation.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Compliance
The unit is classified as pressure equipment under Art. 4.3 PED.
Affects the regulatory classification and documentation required for the installation.
Source
“Pressure equipment category
Art. 4.3 PED”
Compliance
The unit is EMC class A, group 1 according to EN55011.
Class A equipment is intended for non-residential environments and may cause interference in domestic settings.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance
An integral cooling coil for cooling water is provided for operation below or near ambient temperature.
Requires a cooling water supply to be plumbed in when sub-ambient operation is needed.
Source
“An integral cooling coil for cooling water is provided for temperature applications below or near ambient temperatures.”
Selection
A drain valve and a holder for Ubbelohde viscometers are optional accessories, not included.
These must be ordered separately if draining or viscometer mounting is required.
Source
“Optional accessories:
drain valve, holder Ubbelohde viscosimeter #9586”