Peter Huber Kältemaschinenbau CC-130A Visco 3 Visco-Thermostat
Peter Huber Kältemaschinenbau CC-130A Visco 3 Visco-Thermostat — Thermostatic baths.
The CC-130A Visco 3 is a 30-litre viscometer thermostat bath built around a variable-speed pressure/suction pump and the Pilot ONE controller. Its 28–100 °C native range extends down to 15 °C only when water cooling or a refrigerator is added. It holds ±0.01 K at 70 °C for capillary viscometry and density work.
Direct drop-in replacements
Full specifications
| Bath volume | 30 l |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | 28...100 °C with cooling unit: 15...100 °C · with water cooling: 15...100 °C |
| Temperature stability mfr | ±0, 01 K |
| Dimensions (W x D x H) mfr | 500 x 240 x 490 mm |
| Heating capacity mfr | 1, 5 - 2, 1 kW |
| Delivery capacity pressure pump max. mfr | 27 l/min; 0, 7 bar l/min; bar |
| Bath opening WxD/BD mfr | 90 x 90 / 310 mm |
| Filling volume min. mfr | 25, 5 l |
| Weight | 14 kg |
| Current consumption | 10 A |
| Power supply | 200-240V 1~/2~ 50/60Hz V, Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Open bath of 30 l nominal volume (25.5 l minimum fill) with three 90 × 90 mm openings, 310 mm deep, sized for inserting Ubbelohde viscometers and similar glassware.
- Heating capacity 1.5–2.1 kW; pressure/suction pump, variable speed, delivering up to 27 l/min at 0.7 bar through an M16×1 male connection.
- Pilot ONE controller with Ethernet, USB host/device and RS232; internal Pt100 plus an external Pt100 connection for sample-side sensing.
- Safety class III / FL with adjustable overtemperature protection to DIN 12876; pressure equipment per Art. 4.3 PED; EMC class A, group 1 (EN 55011).
Thermal behaviour and media
- Native operating range is 28–100 °C. Reaching 15 °C requires either the integral cooling coil fed with cooling water or a refrigerator; without either, sub-ambient setpoints are not achievable.
- Stability is 0.01 K at 70 °C, which is the figure relevant to capillary viscometry methods run near that setpoint.
- The variable-speed pressure/suction pump allows external circuit flow to be matched to the connected viscometer or heat exchanger rather than fixed at one duty point.
Installation and operation
- Supply is 200–240 V, 1~ or 2~, 50/60 Hz, drawing up to 10 A with a 10 A minimum fuse; external branch protection to UL 489 is required. Voltage and frequency may each deviate ±10% only if the deviations do not run in opposite directions.
- Units ship with a power cable and country-specific plug, which must be specified at order.
- Ambient 5–40 °C, enclosure IP20: keep the unit out of wet or dusty locations.
- Overtemperature cut-out must be set and verified at commissioning before unattended runs.
Integration & Operating Notes
The unit draws up to 10 A on a 200–240 V single- or two-phase supply and requires external branch circuit protection according to UL 489. Undersized protection will trip during heat-up; the supply must also stay within ±10% on voltage and frequency without the two deviations running in opposite directions.
Without cooling the bath only reaches 28 °C; 15 °C operation requires the integral cooling coil plumbed to a cooling water supply or a refrigerator. Plan the cooling water connection and its temperature before specifying a method that runs below ambient.
A drain valve and a holder for Ubbelohde viscometers (#9586) are optional accessories and are not part of the base delivery. Without the drain valve, emptying the 30 l bath for fluid changes must be done another way; without the holder, viscometers must be supported independently.
Selection & Integration Q&A
Pick it when you need a 30 l open bath with three 90 × 90 mm openings for viscometers and 0.01 K stability at 70 °C. If your method runs below 28 °C, budget for water cooling or a refrigerator to reach 15 °C.
A 200–240 V, 1~ or 2~, 50/60 Hz supply with a 10 A minimum fuse and UL 489 branch protection, plus a country-specific plug specified at order. External loop connects via M16×1 male, up to 27 l/min at 0.7 bar.
Ambient must stay 5–40 °C and the IP20 enclosure must be kept dry. Supply voltage and frequency may deviate ±10% only if not in opposite directions. Overtemperature cut-out must be set before unattended runs.
The drain valve and the Ubbelohde viscometer holder (#9586) are optional accessories. The country-specific power plug must be specified at ordering. Cooling water supply or a refrigerator is needed for 15 °C operation.