Introduces the CC-202C immersion circulator device
Specp. 2
Lists technical data according to DIN 12876 standard
p. 2table, 3 rows
p. 1table, 36 rows
Technical data according to DIN 12876
Parameter
Value
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
402842
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
45...200 °C
with water cooling
20...200 °C
with refrigerator
-30...200 °C
Temperature stability at 70°C
0,02 K
temperature set point / display
5,7" colour Touchscreen
Absolute accuracy
setup for calibration
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Heating power at 200V
1,5 kW
Pressure pump
yes
max. delivery
27 l/min
max. delivery pressure
0,7 bar
Suction pump
yes
max. delivery (suction)
22 l/min
max. delivery pressure (suction)
0,4 bar
Pump connection
M16x1 male
Bath volume
3,5 l
Filling capacity
2 l
Bath opening diameter
25 mm
Bath depth
150 mm
Height of bath opening
190 mm
Overall dimensions WxDxH **
178x260x355 mm
Net weight
9 kg
Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
alternative max. current
10 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
** Overall dimensions WxDxH
Engineering Facts
14 facts · every claim linked to a page
Selection
The operating temperature range depends on the cooling method: 45–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator.
Determines whether the unit can reach the required process temperature with the available cooling utility.
Source
“Operating temperature range 45...200 °C with water cooling 20...200 °C with refrigerator -30...200 °C”
Installation
The unit requires a 200–240 V single- or two-phase supply with a minimum 10 A fuse and maximum 16 A fuse.
Electrical supply and branch protection must be sized accordingly before installation.
Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A alternative max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation
External branch circuit protection according to UL 489 is required.
The installer must provide a UL 489-listed branch breaker; this affects panel and wiring design.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Voltage and frequency tolerances are ±10% only if they do not run in opposite directions; for example, -10% voltage with +3% frequency is not allowed.
Site power quality must be checked; combined opposite tolerances can make the supply non-compliant.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation
Single- and two-phase devices are delivered with a power cable and country-specific plug, which must be specified at ordering.
The correct plug type must be selected on the order to avoid rework at commissioning.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation
The unit needs a water cooling coil connection of 3/8 inch for tap water cooling.
Cooling water piping and fittings must be prepared for a 3/8" connection if water cooling is used.
Source
“Cooling coil for (tap) water (3/8").”
Installation
The pump connection is M16x1 male, and sleeve nuts with M16x1 thread are standard equipment.
External circuit hoses and adapters must match the M16x1 male thread.
Source
“Pump connection M16x1 male”
Operation
The permissible ambient temperature range is 5 °C to 40 °C.
The installation location must stay within this range for reliable operation and performance.
Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance
The pressure pump delivers up to 27 l/min at 0.7 bar, while the suction pump delivers up to 22 l/min at 0.4 bar.
External circuit pressure drop and required flow must be checked against these pump limits.