Peter Huber Kältemaschinenbau CC-202C
Peter Huber Kältemaschinenbau CC-202C — Thermostatic baths.
The CC-202C is a 3.5-litre immersion circulator with a 25 mm × 150 mm bath opening, using a variable-speed pressure/suction pump (M16x1 male) and Pilot ONE controller. Its temperature span is configuration-dependent: 45–200 °C uncooled, 20–200 °C with water cooling, and -30–200 °C with a refrigerator. It serves as a compact temperature-control source for external circuits and small bath volumes.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 200C Series — Model Comparison & Selection
| Model | Controller | Interface | Pump connection | Temperature range, °C | Weight, kg | Max delivery flow rate, liter/minute |
|---|---|---|---|---|---|---|
| CC-202C Current | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 45–200 | 9 | 22 |
| CC-205B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 45–200 | 9 | 27 |
| CC-208B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 25–200 | 10 | 27 |
| CC-212B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 11 | 22 |
| CC-215B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 25–200 | 12 | — |
| CC-220B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 14 | 27 |
| CC-225B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 16 | 27 |
| KISS 202C | KISS | USB (Device), RS232 Inte | M16x1 AG male | 45–200 | 8 | 14 |
| KISS 205B Badumwälzthermostat mit Edelstahlbad | KISS | USB (Device), RS232 Inte | M16x1 AG male | 45–200 | 9 | 14 |
| KISS 208B | KISS | USB (Device), RS232 Inte | M16x1 AG male (Option, # | 25–200 | 10 | 14 |
| KISS 212B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 11 | 14 |
| KISS 215B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 12 | 14 |
| KISS 220B | KISS | USB (Device), RS232 Inte | M16x1 AG male (Option, # | 25–200 | 14 | 14 |
| KISS 225B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 16 | 14 |
Full specifications
| Bath volume | 3.5 l |
| Dimensions (w × D × H) | 178 x 260 x 355 mm |
| Filling volume | 2 l (min.) |
| Heating capacity | 1.5 - 2.1 kW |
| Maximum delivery flow rate | 27 l/min (pressure pump max.; 0, 7 bar) suction pump max.; 0, 4 bar: 22 l/min |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | 45...200 °C with cooling unit: -30...200 °C · with water cooling: 20...200 °C |
| Temperature stability | 0.02 K |
| Bath opening WxD/BD mfr | ø = 25 / 150 |
| Weight | 9 kg (net) |
| Current consumption | 10 A (max.) |
| Power supply | 200-240V 1~/2~ 50/60Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
The CC-202C is the smallest bath-volume unit in the CC-20xC family, with a 3.5 l bath and a 25 mm diameter, 150 mm deep opening. Unlike the sibling CC-205B/208B/212B/215B models, which are larger bath variants, this model is an immersion circulator built around a narrow, deep bath well. The pump is a variable-speed pressure and suction type with M16x1 male connections; sleeve nuts with M16x1 thread are included. The controller is the Pilot ONE, with Ethernet, USB host/device, and RS232 interfaces and a Pt100 external sensor input.
Thermal behavior and media
Temperature range is not a single figure: it depends on the cooling method. Without cooling, the unit operates from 45 to 200 °C; with water cooling, from 20 to 200 °C; with a refrigerator, from -30 to 200 °C. Temperature stability is 0.02 K. Heating capacity is 1.5–2.1 kW. The pump delivers up to 27 l/min at 0.7 bar in pressure mode and up to 22 l/min at 0.4 bar in suction mode, so external circuit pressure drop must be checked against these limits.
Installation, control and operation
The unit requires a 200–240 V, 1~ or 2~, 50/60 Hz supply with a maximum current of 10 A; branch protection must be 10 A minimum and 16 A maximum, with an external UL 489-listed breaker. Voltage and frequency tolerances are ±10% only if they do not run in opposite directions. A country-specific plug must be specified at ordering. The cooling-water coil connection is 3/8 inch. The unit is IP20 and Class A Group 1 to EN55011, so it must be protected from water and dust and may need additional EMC measures in sensitive environments. Ambient temperature must remain between 5 and 40 °C. Overtemperature protection is adjustable per DIN 12876, safety class III/FL.
Integration & Operating Notes
The supply is 200–240 V, 1~ or 2~, 50/60 Hz with a maximum current of 10 A. The external branch circuit must use a UL 489-listed breaker rated between 10 A minimum and 16 A maximum; undersizing trips on run-up, oversizing leaves the unit unprotected.
The degree of protection is IP20, so the unit must be installed where it is protected from water ingress and dust. Ambient temperature must stay between 5 and 40 °C. Class A Group 1 to EN55011 means additional EMC measures may be needed in residential or sensitive environments.
The unit carries safety class III/FL and adjustable overtemperature protection according to DIN 12876. The trip point must be set for the application before operation; it is the primary safeguard against thermal runaway in the bath and external circuit.
Selection & Integration Q&A
Choose the CC-202C when the sample or external circuit fits a 3.5 l bath with a 25 mm × 150 mm opening and the 1.5–2.1 kW heater is sufficient. The larger siblings suit bigger bath volumes and immersion depths; this model is the compact option.
The pump connection is M16x1 male, and sleeve nuts with M16x1 thread are standard. Hoses and adapters must match this thread, and the circuit pressure drop must stay within 0.7 bar at up to 27 l/min (pressure) or 0.4 bar at up to 22 l/min (suction).
Ambient temperature must remain 5–40 °C, and the IP20 enclosure requires protection from water and dust. The supply must be 200–240 V, 1~ or 2~, 50/60 Hz with a 10–16 A UL 489 branch breaker; voltage and frequency tolerances must not run in opposite directions.
The country-specific plug must be specified at ordering. A 3/8-inch water cooling coil connection is needed if water cooling is used, and a refrigerator is required for -30 °C operation. External hoses and adapters must match the M16x1 male pump thread; sleeve nuts are included.