Contents

  1. Overview p. 1
    Product name and cover of CC-510w circulator
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 2 table, 11 rows

Technical data according to DIN 12876

ParameterValue
Height of bath opening815 mm
Overall dimensions WxDxH **455x515x1014 mm
Net weight106 kg
Power supply requirement (3 phase)400V 3~N 50Hz
max. current (3 Phase)10,5 A
Fuse (3 phase)3x16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
max. ambient temperature40 °C
min. ambient temperature5 °C
from Serial-No.:S401658

** Please respect space requirements. See operating conditions at www.huber-online.com

Engineering Facts

12 facts · every claim linked to a page
Installation

The unit requires a three-phase 400 V supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable and protection device needed before installation.

Source
“Power supply requirement (3 phase) 400V 3~N 50Hz”
Installation

The cooling water circuit needs a differential pressure between 3 and 6 bar, which must be verified in the building water system.

Insufficient cooling water pressure will prevent the water-cooled refrigeration machine from operating correctly.

Source
“min. cooling water differential pressure 3 bar”
Installation

Cooling water intake temperature must stay between 5 °C and 20 °C, and pipes must be insulated if condensation can occur.

Defines the permissible cooling water supply conditions and condensation protection required at the site.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops significantly at low temperatures, from 2.4 kW at 0 °C to 0.4 kW at -40 °C, which affects process design.

Engineers must check that available cooling capacity matches the heat load at the lowest operating temperature.

Source
“at -40°C 0,4 kW”
Performance

The pressure pump delivers up to 25 l/min at 0.7 bar, and the suction pump up to 18.5 l/min at 0.4 bar.

Pump performance determines whether the external circuit can achieve the required flow and pressure drop.

Source
“max. delivery 25 l/min”
Operation

The maximum permissible kinematic viscosity of the medium is 50 mm²/s, limiting the fluids that can be pumped.

Prevents selection of media too viscous for the pump and ensures proper circulation.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient temperature.

Defines the environmental limits and protection level for the installation location.

Source
“max. ambient temperature 40 °C”
Safety

The refrigerant R-452A has a GWP of 2141 and a charge of 1 kg, giving a CO2 equivalent of 2.1 t relevant for leak and F-gas obligations.

Affects environmental compliance, leak checks and refrigerant handling requirements.

Source
“Global Warming Potential (GWP) 2141”
Compliance

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Ensures the supply quality at the installation site is within the unit's permissible operating envelope.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices below 63 A are supplied with cable but without a plug, so a suitable plug must be provided on site.

Clarifies the scope of delivery and what electrical accessories the installer must supply.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The bath holds 18 l, with a minimum filling of 16 l and 11 l when a displacement rack is used.

Determines the required medium volume and safe operating fill level.

Source
“Bath volume 18 l”
Compliance

The equipment is classified as Class A, Group 1 for electromagnetic disturbance, which has implications for the installation environment.

Class A equipment may require EMC measures and is intended for non-domestic environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”