Peter Huber Kältemaschinenbau CC-510w Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau CC-510w Kälte-Badumwälzthermostat — Thermostatic baths.
The CC-510w is a water-cooled refrigerated bath circulator with an 18 l bath, a variable-speed pressure/suction pump and the Pilot ONE controller, covering -50 to 200 °C at ±0.02 K stability. Its 3 kW heater and 2.4 kW cooling at 0 °C (0.4 kW at -40 °C) position it above the smaller CC-405/410/415/505 siblings for higher heat loads. It serves as a temperature-control source for external circuits and bath immersion applications requiring three-phase power and building cooling water.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection
| Model | Noise level, dB | Power supply | Pump connection | Temperature range, °C | Weight, kg | Filling volume, m³ |
|---|---|---|---|---|---|---|
| CC-510w Kälte-Badumwälzthermostat Current | — | 400V 3~N 50Hz | M16x1 AG male | -50–200 | 106 | 0.016 |
| CC-410wl Kälte-Badumwälzthermostat | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -45–200 | 71 | 0.017 |
| CC-415wl Kälte-Badumwälzthermostat | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -40–200 | 61 | 0.004 |
| CC-508w Kälte-Badumwälzthermostat | 57 | 208-230V 1~/2~ 60Hz | M16x1 AG male | -55–200 | — | 0.004 |
| CC-515w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 105 | 0.016 |
| CC-520w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 161 | 0.015 |
| CC-525w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 163 | 0.015 |
| CC-905 Kälte-Badumwälzthermostat | 64 | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 171 | 0.019 |
| CC-905w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 175 | 0.019 |
| CC-906w Kälte-Badumwälzthermostat | — | 460V 3~ 60Hz | M16x1 AG male | -90–200 | 234 | 0.025 |
| CC-K12 Kälte-Badthermostat | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 29 | 0.007 |
| CC-K15 Kälte-Badthermostat | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 28 | — |
| CC-K20 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.012 |
| CC-K25 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.017 |
| CC-K6 Kälte-Badumwälzthermostat | 57 | 208-240V 1~/2~ 50/60Hz | M16x1 AG male | -25–200 | 25 | — |
| Variostat Kältethermostat für variable Badgefäße | 54 | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -30–150 | 25 | — |
Full specifications
| Bath volume | 18 l |
| Filling volume | 16 l (min) |
| Heating capacity | 3 kW |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | -50...200 °C |
| Temperature stability mfr | ±0, 02 K |
| Dimensions (W x D x H) mfr | 455 x 515 x 1014 mm |
| Delivery capacity pressure pump max. mfr | 25 l/min; 0, 7 bar |
| Delivery capacity suction pump max. mfr | 18, 5 l/min; 0, 4 bar |
| Bath opening WxD/BD mfr | 270 x 150 / 200 mm |
| Weight | 106 kg |
| Power supply | 400V 3~N 50Hz |
| Current consumption max. mfr | 10, 5 A (max) |
| Fuse mfr | 3x16A |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water differential pressure | 3 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Cooling capacity mfr | 100:2, 4; 20:2, 4; 0:2, 4; -20:1; -40:0, 4 °C / (multiple temperature points) |
| Safety class | III / FL |
Description
Construction and series positioning
- Water-cooled refrigeration machine (R-452A, 1 kg charge) with a 400 V 3~N 50 Hz three-phase supply, 10.5 A max current and 3x16 A fuse; the CC-510w is the higher-capacity member of the CC series alongside the CC-405w, CC-410wl, CC-415wl and CC-505wl.
- Bath volume 18 l (minimum filling 16 l; 11 l with displacement rack), bath opening 270 x 150 mm with 200 mm depth, external dimensions 455 x 515 x 1014 mm, 106 kg.
- Pump is a variable-speed pressure and suction type: pressure side up to 25 l/min at 0.7 bar, suction side up to 18.5 l/min at 0.4 bar; pump connection M16x1 male, cooling-water connection G1/2 male.
- Controller: Pilot ONE; interfaces Ethernet, USB host and device, RS232; external Pt100 sensor connection; safety class III/FL; IP20.
Behavior under load and media
- Temperature range -50 to 200 °C with ±0.02 K stability. Cooling capacity falls with temperature: 2.4 kW at 100, 20 and 0 °C, 1 kW at -20 °C and 0.4 kW at -40 °C, so the lowest setpoints carry a much reduced heat-load budget.
- Heating capacity 3 kW. The medium must not exceed 50 mm²/s kinematic viscosity, which limits the pumpable fluids at low temperatures.
- Cooling water must be supplied at 5–20 °C intake with a differential pressure of 3–6 bar; condensation on uninsulated pipework must be prevented.
Installation, control and operation
- Three-phase devices below 63 A are delivered with cable but without a plug; the site must provide a suitable plug and a 3x16 A protected supply. Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not permitted.
- Ambient operating range 5–40 °C, IP20, Class A Group 1 (EN55011) for electromagnetic disturbance, so a non-domestic industrial environment is assumed.
- The Pilot ONE controller provides Ethernet, USB and RS232 connectivity and accepts an external Pt100 sensor for control at the external circuit.
Integration & Operating Notes
The water-cooled refrigeration circuit requires a differential pressure between 3 and 6 bar and an intake temperature between 5 °C and 20 °C. Verify these conditions in the building water system before commissioning; insulate pipework where condensation can form.
The maximum permissible kinematic viscosity of the bath medium is 50 mm²/s. Selecting a fluid that is more viscous, particularly at low temperatures, will prevent proper circulation through the external circuit and bath.
Cooling power falls from 2.4 kW at 0 °C to 1 kW at -20 °C and 0.4 kW at -40 °C. The heat load of the application at the lowest operating temperature must be checked against this reduced capacity to avoid losing temperature control.
The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient temperature. It is classified Class A, Group 1 for electromagnetic disturbance to EN55011, so it is intended for non-domestic environments and may require EMC measures.
The refrigerant R-452A has a GWP of 2141 and a charge of 1 kg, giving a CO2 equivalent of 2.1 t. This is relevant for leak checks and F-gas handling requirements at the installation site.
Selection & Integration Q&A
Choose the CC-510w when the heat load at low temperature requires its 2.4 kW cooling at 0 °C and 0.4 kW at -40 °C, and when the 18 l bath and 3 kW heater suit the application. The smaller siblings offer less capacity; select them only if their lower cooling and bath volume meet the duty.
A 400 V 3~N 50 Hz supply fused at 3x16 A with a suitable plug (cable supplied, plug not included), and cooling water at 5–20 °C with 3–6 bar differential pressure on G1/2 male connections. Voltage tolerance is ±5% only with ±2% frequency tolerance.
Operate between 5 °C and 40 °C ambient at IP20. The unit is Class A, Group 1 (EN55011) for electromagnetic disturbance, so it is intended for non-domestic environments and may require EMC measures. Cooling water pipes must be insulated where condensation can occur.
The three-phase unit is supplied with cable but without a plug, so a site plug and 3x16 A protection must be provided. External circuit connections use M16x1 male pump fittings and G1/2 male cooling-water fittings; an external Pt100 sensor is supported but not stated as included.