CC-520w Circulating Bath Datasheet

Contents

  1. Overview p. 1
    Introduces the CC-520w circulating bath and its main features
  2. Spec pp. 1-2
    Performance, temperature range, and dimensional data per DIN 12876
  3. p. 2 table, 12 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Width bath opening WxD / bath depth270 x 150 / 200 mm
Height of bath opening903 mm
Overall dimensions WxDxH **539x629x1102 mm
Net weight161 kg
Power supply requirement (3 phase)400V 3~N 50Hz
max. current (3 Phase)13,5 A
Fuse (3 phase)3x16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
max. ambient temperature40 °C
min. ambient temperature5 °C
from Serial-No.:S373466

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-55...200 °C
Temperature stability at -10°C0,02 K
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power3 kW
Cooling power at 100°C5 kW
Cooling power at 20°C5 kW
Cooling power at 0°C5 kW
Cooling power at -20°C3 kW
Cooling power at -40°C1,5 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity1,65 kg
CO2 equivalent3,5 t
Pressure pumpyes
max. delivery25 l/min
max. delivery pressure0,7 bar
Suction pumpyes
max. delivery (suction)18,5 l/min
max. delivery pressure (suction)0,4 bar
Pump connectionM16x1 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 15°C288 l/h
min. cooling water differential pressure3 bar
max. cooling water pressure6 bar
Bath volume17 l
min. filling capacity15 l
Bath capacity with displacement rack10 l

Order-No.: CC-520w; 2022.0001.01

Engineering Facts

12 facts · every claim linked to a page
Installation

The unit requires a three-phase 400 V supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply and protection device needed before installation.

Source
“Power supply requirement (3 phase) 400V 3~N 50Hz”
Installation

The cooling water circuit needs a differential pressure of at least 3 bar and must not exceed 6 bar.

Ensures the available cooling water supply is adequate and safe for the water-cooled condenser.

Source
“min. cooling water differential pressure 3 bar”
Installation

Cooling water inlet temperature must stay between 5 °C and 20 °C, and pipes must be insulated if condensation can occur.

Prevents condensation damage and loss of cooling performance at the installation site.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling capacity drops as cooling water temperature rises, and the required cooling water flow rate may increase.

Affects sizing of the cooling water supply and expected process performance.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water quality should be used.

Guides water treatment and material compatibility checks to avoid corrosion or fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Defines acceptable site power quality for reliable operation and warranty compliance.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices below 63 A are delivered with a cable but without a plug, so the customer must provide the plug.

Affects electrical installation planning and the bill of materials for commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The pump connection is M16x1 male and the maximum permissible kinematic viscosity is 50 mm²/s.

Ensures the external circuit and heat transfer fluid are compatible with the pump.

Source
“max. permissible kin. viscosity 50 mm²/s”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.

Confirms the safety concept required for flammable or sensitive bath fluids.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation

The refrigerant is R-452A with GWP 2141 and a CO2 equivalent of 3.5 t, which matters for leak monitoring and environmental reporting.

Relevant for refrigerant regulations, leak checks and environmental compliance.

Source
“CO2 equivalent 3,5 t”
Installation

The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient temperature.

Defines the permissible room conditions and protection against ingress.

Source
“max. ambient temperature 40 °C”
Selection

The bath holds 17 l, with a minimum filling of 15 l and 10 l when a displacement rack is used.

Helps select the right bath volume and displacement accessory for the application.

Source
“Bath capacity with displacement rack 10 l”