Peter Huber Kältemaschinenbau CC-520w Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau CC-520w Kälte-Badumwälzthermostat — Thermostatic baths.
The CC-520w is a water-cooled circulating bath thermostat with a 17 l bath, a variable-speed pressure/suction pump, and a 3 kW heater, covering -55 to 200 °C with ±0.02 K stability. It is the 5-series modification with the largest cooling capacity in the Cc series, distinguished from the CC-505wl by its higher cooling output and 400 V three-phase supply. It serves as a temperature-control source for external circuits and bath applications requiring high heat rejection.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection
| Model | Noise level, dB | Power supply | Pump connection | Temperature range, °C | Weight, kg | Filling volume, m³ |
|---|---|---|---|---|---|---|
| CC-520w Kälte-Badumwälzthermostat Current | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 161 | 0.015 |
| CC-410wl Kälte-Badumwälzthermostat | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -45–200 | 71 | 0.017 |
| CC-415wl Kälte-Badumwälzthermostat | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -40–200 | 61 | 0.004 |
| CC-508w Kälte-Badumwälzthermostat | 57 | 208-230V 1~/2~ 60Hz | M16x1 AG male | -55–200 | — | 0.004 |
| CC-510w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -50–200 | 106 | 0.016 |
| CC-515w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 105 | 0.016 |
| CC-525w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 163 | 0.015 |
| CC-905 Kälte-Badumwälzthermostat | 64 | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 171 | 0.019 |
| CC-905w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 175 | 0.019 |
| CC-906w Kälte-Badumwälzthermostat | — | 460V 3~ 60Hz | M16x1 AG male | -90–200 | 234 | 0.025 |
| CC-K12 Kälte-Badthermostat | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 29 | 0.007 |
| CC-K15 Kälte-Badthermostat | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 28 | — |
| CC-K20 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.012 |
| CC-K25 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.017 |
| CC-K6 Kälte-Badumwälzthermostat | 57 | 208-240V 1~/2~ 50/60Hz | M16x1 AG male | -25–200 | 25 | — |
| Variostat Kältethermostat für variable Badgefäße | 54 | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -30–150 | 25 | — |
Full specifications
| Bath volume | 17 l |
| Filling volume | 15 l (min.) |
| Heating capacity | 3 kW |
| Maximum delivery flow rate | 25 l/min (pressure pump max.; 0, 7 bar) |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | -55...200 °C |
| Temperature stability mfr | ±0, 02 K |
| Dimensions (W x D x H) mfr | 539 x 629 x 1102 mm |
| Delivery capacity suction pump max. mfr | 18, 5 l/min; 0, 4 bar |
| Bath opening WxD/BD mfr | 270 x 150 / 200 |
| Weight | 161 kg |
| Power supply | 400V 3~N 50Hz |
| Current consumption max. mfr | 13, 5 A (max.) |
| Fuse mfr | 3x16A |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 288 l/h (at 15°C, flow 15°C) |
| Cooling water differential pressure | 3 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Cooling capacity mfr | 100:5; 20:5; 0:5; -20:3; -40:1, 5 °C / kW |
| Safety class | III / FL |
Description
Construction and series positioning
The CC-520w is a floor-standing (539 x 629 x 1102 mm, 161 kg) circulating bath thermostat with a 17 l bath and a 270 x 150 mm opening (200 mm depth). It uses a water-cooled condenser (R-452A refrigerant) rather than an air-cooled one, which is why it requires a cooling-water circuit. Within the Cc series, the CC-520w is the highest-capacity water-cooled model; the CC-505wl is a smaller sibling, and the CC-405/410/415 models are lower-capacity variants. The CC-520w is the only sibling listed with a 400 V three-phase supply and 13.5 A maximum current.
Thermal behavior and media
- Temperature range: -55 to 200 °C; stability ±0.02 K.
- Cooling capacity (kW at °C): 5 at 100 °C, 5 at 20 °C, 5 at 0 °C, 3 at -20 °C, 1.5 at -40 °C.
- Heating capacity: 3 kW.
- Pump: variable-speed pressure and suction pump; max delivery 25 l/min at 0.7 bar (pressure side), max suction 18.5 l/min at 0.4 bar.
- Bath filling: 15 l minimum; 10 l with displacement rack.
- Max permissible kinematic viscosity: 50 mm²/s.
Installation, connection and control
- Power: 400 V 3~N 50 Hz, 13.5 A max, 3x16 A fuse; delivered with cable but without plug (three-phase devices below 63 A).
- Cooling water: G1/2 male connections; differential pressure min. 3 bar, max. 6 bar; consumption 288 l/h at 15 °C; inlet temperature 5–20 °C; circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.
- Pump connection: M16x1 male.
- Controller: Pilot ONE; interfaces Ethernet, USB (host and device), RS232; external Pt100 sensor connection.
- Safety: adjustable overtemperature protection per DIN 12876, safety class III / FL; IP20; ambient 5–40 °C.
- Voltage tolerance ±5% with frequency tolerance ±2% (certain combinations not allowed).
Integration & Operating Notes
The unit draws up to 13.5 A per phase and requires a 3x16 A fuse. The electrical installation must be planned for this load, and the customer must provide a suitable plug because the unit is delivered with cable but without plug. Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%.
The water-cooled condenser requires a differential pressure of at least 3 bar and no more than 6 bar. Inlet temperature must be between 5 °C and 20 °C; higher temperatures reduce cooling capacity and may increase required flow. Pipes must be insulated if condensation can occur. The circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so water quality must be compatible.
The pump is rated for a maximum kinematic viscosity of 50 mm²/s. Using a more viscous fluid can overload the pump and reduce circulation. The pump connection is M16x1 male, and the bath requires a minimum filling of 15 l (10 l with displacement rack).
The refrigerant is R-452A with a GWP of 2141 and a CO2 equivalent of 3.5 t. Regular leak monitoring and environmental reporting may be required depending on local regulations. The cooling water circuit should be checked for fouling and corrosion to maintain heat transfer.
Selection & Integration Q&A
Choose the CC-520w when you need the highest cooling capacity in the Cc series (5 kW at 20 °C) and a 400 V three-phase supply. The CC-505wl and 4-series models are lower-capacity variants; select them if your heat load is smaller or you only have a single-phase supply.
Cooling water: G1/2 male connections, differential pressure 3–6 bar, inlet temperature 5–20 °C, flow 288 l/h at 15 °C. Pump circuit: M16x1 male connection, fluid viscosity ≤50 mm²/s. Ensure water quality is compatible with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.
The unit is delivered with a cable but without a plug; the customer must provide a suitable three-phase plug. Refrigerant R-452A (GWP 2141) requires leak monitoring. Bath fluid and external Pt100 sensor are not included. Regular checks of cooling water quality and circuit fouling are recommended.