Selection
The unit uses flammable natural refrigerant R-290 (A3, H220), so the installation location must be assessed for flammable-refrigerant requirements.
Refrigerant safety class A3 affects site risk assessment, ventilation and compliance approvals.
Source
“R-290 (A3, H220)”
Installation
The power supply must be 220-240 V single- or two-phase at 50/60 Hz, so the site must provide a matching circuit.
Mismatched supply voltage or phase configuration prevents correct commissioning.
Source
“220-240V 1~/2~ 50/60Hz”
Installation
The circuit must be protected with a fuse between 10 A and 16 A.
Correct upstream protection is required for safe electrical installation.
Source
“min. Fuse 10A max. Fuse 16A”
Installation
Voltage and frequency tolerances must be applied together: ±5% voltage with simultaneous ±2% frequency, and some combinations are not allowed.
Supply quality outside these combined limits can cause malfunction or damage.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
The unit is IP20, so it must be installed in a dry indoor location protected from water ingress.
IP20 is not water-protected and constrains the installation environment.
Source
“Degree of Protection IP20”
Operation
Ambient temperature must stay between 5 °C and 40 °C, and cooling capacity may drop above 20 °C room temperature.
Exceeding ambient limits reduces performance and may void safe operation.
Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance
Cooling capacity falls from 0.25 kW at 20 °C to 0.05 kW at -20 °C, so low-temperature duty needs careful sizing.
Available cooling capacity at the required working temperature determines process feasibility.
Source
“at -20°C 0,05 kW”
Operation
The pressure/suction pump delivers up to 27 l/min at 0.7 bar pressure and 22 l/min suction at 0.4 bar.
Pump flow and pressure determine whether the external circuit can be supplied adequately.
Source
“max. delivery 27 l/min”
Operation
When using the Huber pump adapter, polyglycol must not be used as heat transfer fluid.
Using the wrong fluid with the pump adapter can damage the pump or compromise safety.
Source
“Polyglycol is not permissible to be used as a heat transfer fluid”
Installation
Three-phase devices below 63 A are supplied with cable but without a plug, so the customer must provide the plug.
Planning must include a suitable plug and connection for three-phase units.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Compliance
The equipment is classified as Class A, Group 1 for electromagnetic disturbance, which has implications for residential EMC environments.
Class A equipment may not be suitable for residential use without additional EMC measures.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety
The unit has adjustable overtemperature protection according to DIN 12876, which must be set for the application.
Correct setting of overtemperature protection is essential for safe unattended operation.
Source
“With adjustable overtemperature protection according to DIN 12876.”