Peter Huber Kältemaschinenbau CC-K12 Kälte-Badthermostat
Peter Huber Kältemaschinenbau CC-K12 Kälte-Badthermostat — Thermostatic baths.
The CC-K12 is a refrigerated/heating bath circulator with a 10.5 L bath and a variable-speed pressure/suction pump, covering -20 to 200 °C with ±0.02 K stability. It is the compact 7 L-minimum-fill modification in the CC series, distinguished by its Pilot ONE controller and 1.8–2.1 kW heater. It serves as a temperature-control source for external circuits or direct bath immersion in laboratory processes.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection
| Model | Noise level, dB | Power supply | Pump connection | Temperature range, °C | Weight, kg | Filling volume, m³ |
|---|---|---|---|---|---|---|
| CC-K12 Kälte-Badthermostat Current | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 29 | 0.007 |
| CC-410wl Kälte-Badumwälzthermostat | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -45–200 | 71 | 0.017 |
| CC-415wl Kälte-Badumwälzthermostat | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -40–200 | 61 | 0.004 |
| CC-508w Kälte-Badumwälzthermostat | 57 | 208-230V 1~/2~ 60Hz | M16x1 AG male | -55–200 | — | 0.004 |
| CC-510w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -50–200 | 106 | 0.016 |
| CC-515w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 105 | 0.016 |
| CC-520w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 161 | 0.015 |
| CC-525w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 163 | 0.015 |
| CC-905 Kälte-Badumwälzthermostat | 64 | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 171 | 0.019 |
| CC-905w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 175 | 0.019 |
| CC-906w Kälte-Badumwälzthermostat | — | 460V 3~ 60Hz | M16x1 AG male | -90–200 | 234 | 0.025 |
| CC-K15 Kälte-Badthermostat | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 28 | — |
| CC-K20 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.012 |
| CC-K25 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.017 |
| CC-K6 Kälte-Badumwälzthermostat | 57 | 208-240V 1~/2~ 50/60Hz | M16x1 AG male | -25–200 | 25 | — |
| Variostat Kältethermostat für variable Badgefäße | 54 | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -30–150 | 25 | — |
Full specifications
| Dimensions (w × D × H) | 350 x 560 x 430 mm |
| Filling volume | 7 l (minimum) |
| Maximum delivery flow rate | 27 l/min (pressure pump; 0, 7 bar) suction pump; 0, 4 bar: 22 l/min |
| Pump connection | (Option, #19607) (Option) |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | -20...200 °C |
| Temperature stability mfr | ±0, 02 K |
| Heating capacity mfr | 1, 8 - 2, 1 kW |
| Bath opening WxD/BD mfr | 290 x 152 / 150 mm (WxD/depth) |
| Bath volume mfr | 10, 5 l |
| Weight | 29 kg |
| Power supply | 220-240V 1~/2~ 50/60Hz |
| Current consumption max. mfr | 10 + 1, 5 A (max.) |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Cooling capacity mfr | 20:0, 25; 0:0, 2; -10:0, 12; -20:0, 05 °C / kW |
| Safety class | III / FL |
Description
Construction and series positioning
- The CC-K12 is a refrigerated/heating circulator with an integrated 10.5 L bath (7 L minimum filling volume) and a pressure/suction pump; it is the smallest bath-volume model in the CC series, below the CC-405w/410wl/415wl/505wl siblings.
- Temperature range spans -20 to 200 °C with ±0.02 K stability, using the Pilot ONE controller and a Pt100 external sensor connection.
- Heating capacity is 1.8–2.1 kW; cooling capacity is 0.25 kW at 20 °C, 0.2 kW at 0 °C, 0.12 kW at -10 °C and 0.05 kW at -20 °C.
- The unit uses R-290 (A3, H220) refrigerant, which imposes flammable-refrigerant site requirements.
Behavior under load and media
- Cooling capacity falls sharply toward -20 °C, so low-temperature duty requires careful load sizing; above 20 °C ambient, cooling capacity may drop further.
- The variable-speed pressure/suction pump delivers up to 27 l/min at 0.7 bar pressure and 22 l/min at 0.4 bar suction, allowing adaptation to external circuit resistance.
- Polyglycol must not be used as heat transfer fluid when the Huber pump adapter is fitted.
Installation, control and operation
- Power supply is 220–240 V, 1~ or 2~, 50/60 Hz; maximum current consumption is 10 + 1.5 A; the circuit must be fused between 10 A and 16 A.
- Voltage tolerance ±5% with simultaneous frequency tolerance ±2%; some combinations are not allowed.
- Protection rating is IP20, requiring a dry indoor location; ambient temperature must remain 5–40 °C.
- Interfaces include Ethernet, USB (host and device) and RS232; the unit has adjustable overtemperature protection per DIN 12876.
- EMC classification is Class A, Group 1 (EN 55011), with implications for residential environments.
Integration & Operating Notes
The site circuit must provide 220–240 V single- or two-phase at 50/60 Hz and be protected by a fuse rated between 10 A and 16 A; maximum current consumption is 10 + 1.5 A, so undersized protection will trip and oversized protection will not protect the unit.
Voltage may deviate ±5% only with a simultaneous frequency deviation of ±2%, and some combinations are not allowed; supply quality outside these combined limits can cause malfunction or damage, so the site supply should be verified before commissioning.
The unit is rated IP20 and must be installed in a dry indoor location protected from water ingress; ambient temperature must stay between 5 °C and 40 °C, and cooling capacity may drop above 20 °C room temperature.
Cooling capacity falls from 0.25 kW at 20 °C to 0.05 kW at -20 °C, so processes requiring sustained cooling near the lower limit must be sized against the reduced capacity to avoid inability to reach or hold setpoint.
When the Huber pump adapter is used, polyglycol must not be used as heat transfer fluid; selecting an incompatible fluid can damage the pump or compromise safety, so fluid choice must be confirmed against the adapter instructions.
Three-phase devices with current consumption below 63 A are supplied with cable but without a plug, so the customer must provide a suitable plug and connection for commissioning.
The equipment is classified as Class A, Group 1 for electromagnetic disturbance to EN 55011, which has implications for residential EMC environments and may require additional measures if installed near sensitive equipment.
The unit has adjustable overtemperature protection according to DIN 12876; it must be set for the application before unattended operation to prevent overheating if the control loop fails.
The unit uses R-290 (A3, H220) flammable natural refrigerant, so the installation location must be assessed for flammable-refrigerant requirements, including ventilation and compliance approvals, before siting.
Selection & Integration Q&A
Choose the CC-K12 when a 10.5 L bath and 7 L minimum fill suffice and the -20 to 200 °C range with 0.05–0.25 kW cooling meets the load; the larger CC-405w/410wl/415wl/505wl siblings are for higher bath volumes or duty beyond this compact unit's capacity.
Provide 220–240 V, 1~ or 2~, 50/60 Hz with a 10–16 A fuse; maximum current is 10 + 1.5 A. Voltage ±5% and frequency ±2% apply together, and some combinations are not allowed. Three-phase units below 63 A ship without a plug.
Install indoors in a dry location: IP20, ambient 5–40 °C, with cooling capacity possibly reduced above 20 °C room temperature. The R-290 (A3) refrigerant requires a flammable-refrigerant site assessment, and Class A EMC may restrict residential use.
The base unit does not include a plug for three-phase versions below 63 A; the customer must provide it. The Huber pump adapter is an option (#19607), and polyglycol must not be used with it. Overtemperature protection must be set per DIN 12876 before unattended operation.