Selection
The usable temperature range depends on the cooling method, so the chiller/cooling option must be chosen to reach the required low temperature.
Determines whether the bath can reach the process temperature without an external refrigerator or water cooling.
Source
“Operating temperature range
25...100 °C
with water cooling
20...100 °C
with refrigerator
15...100 °C”
Performance
Heating power is voltage-dependent, so the actual available heating capacity must be checked against the site supply voltage.
Affects heat-up time and whether the unit meets the process requirement at the installed voltage.
Source
“Heating power at 230V
2 kW”
Installation
The unit requires a 200-240 V single- or two-phase supply with a 10 A minimum and 16 A maximum fuse.
Needed to size the branch circuit and protective device correctly.
Source
“Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
min. Fuse
10A
max. Fuse
16A”
Installation
An external branch circuit protection according to UL 489 is mandatory.
Must be provided in the installation for compliance and safe operation.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Voltage and frequency tolerances are limited and must not deviate in opposite directions.
Prevents specifying the unit on a supply where combined voltage/frequency deviation is out of tolerance.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Safety
The device is safety class III / FL with adjustable overtemperature protection per DIN 12876.
Relevant for risk assessment and for setting the overtemperature cut-out during commissioning.
Source
“With adjustable
overtemperature protection according to DIN 12876.”
Operation
The autostart function restarts the unit automatically after a power failure.
Important for unattended processes and for deciding whether this behavior is acceptable.
Source
“Autostart function for power failure”
Selection
The plastic pump is not resistant to acetone and polyglycols, so these media require an alternative pump material.
Prevents selecting the standard unit for incompatible bath fluids.
Source
“Special Case: Acetone and Polyglycol: The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure.
Guides the choice of heat transfer fluid for low-temperature operation.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
The Pt100 process sensor option is factory-fitted only and is for display, not control.
Must be ordered with the unit and cannot be used for closed-loop control of process temperature.
Source
“Option: Pt100 sensor connection #10688 to display (not control) e.g. of the process temperature (only available factory
fitted, additional charge).”
Installation
The unit is IP20 and must be installed in an environment between 5 °C and 40 °C.
Defines the permissible installation location and ambient conditions.
Source
“Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Installation
Overall dimensions and space requirements must be respected at the installation site.
Ensures adequate clearance and access for installation and operation.
Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”
Selection
The bath holds 4.4 l with a minimum filling capacity of 2.5 l.
Determines the required fluid volume and whether the bath suits the application.
Source
“Bath volume
4,4 l
min. filling capacity
2,5 l”