Peter Huber Kältemaschinenbau KISS 106A
Peter Huber Kältemaschinenbau KISS 106A — Thermostatic baths.
The KISS 106A is an immersion circulator with a 4.4 L polycarbonate bath opening of 130 x 110 mm and a pressure/suction pump, rated for 25...100 °C without auxiliary cooling. It is the mid-size 106 variant in the KISS series, distinguished from the 104/108/110 siblings by its 4.4 L bath volume. It serves as a stand-alone thermostatic bath for sample tempering in the 25–100 °C range.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Polycarbonatbaedern Bis 100C Series — Model Comparison & Selection
| Model | Controller | Interface | Max delivery flow rate, liter/minute | Pump connection | Weight, kg | Ambient temperature limit, °C |
|---|---|---|---|---|---|---|
| KISS 106A Current | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 5 | 5–40 |
| CC-104A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 6 | 5–40 |
| CC-106A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 5 | 5–40 |
| CC-108A | Pilot ONE | Ethernet, USB (Host u. D | 27 | M16x1 AG male (Option, # | 6 | 5–40 |
| CC-110A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 6 | 5–40 |
| CC-112A | Pilot ONE | Ethernet, USB (Host u. D | 27 | M16x1 AG male (Option, # | 8 | 5–40 |
| CC-118A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male (Option, # | 8 | 5–40 |
| KISS 104A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male | 5 | 5–40 |
| KISS 108A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 6 | 5–40 |
| KISS 110A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 6 | 5–40 |
| KISS 112A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 8 | 5–40 |
| KISS 118A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 8 | 5–40 |
Full specifications
| Alarm output | optic, acoustic |
| Dimensions (w × D × H) | 147 x 307 x 330 mm |
| Maximum delivery flow rate | 14 l/min (pressure pump; 0, 25 bar) |
| Pump connection | M16x1 AG male (Option, #19606) (Option, #19606) |
| Pump speed adjustable | no |
| Pump type | Pressure and suction pump, speed not variable |
| Temperature range | 25...100 °C with cooling unit: 15...100 °C · with water cooling: 20...100 °C |
| Temperature stability mfr | ±0, 05 K |
| Heating capacity mfr | 1, 5 - 2, 1 kW |
| Delivery capacity suction pump max. mfr | 10, 5 l/min (suction pump; 0, 17 bar) |
| Bath opening WxD/BD mfr | 130 x 110 / 150 mm (WxD/depth) |
| Bath volume mfr | 4, 4 l |
| Filling volume min. mfr | 2, 5 l (min) |
| Weight | 5 kg |
| Current consumption | 10 A (max) |
| Power supply | 200-240V 1~/2~ 50/60Hz |
| Controller | KISS |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Immersion circulator with integrated 4.4 L polycarbonate bath; bath opening 130 x 110 mm, depth 150 mm; minimum filling 2.5 L.
- Pressure and suction pump, speed not adjustable: max delivery 14 L/min at 0.25 bar (pressure side), max suction 10.5 L/min at 0.17 bar.
- Heating capacity 1.5–2.1 kW depending on supply voltage; temperature stability ±0.05 K.
- Within the KISS series the 106A sits between the 104A and 108A/110A siblings; the 106A is the 4.4 L bath size, so choose it when the sample load and immersion depth fit that opening, and step to a larger sibling when more bath volume or opening area is required.
Thermal behavior and media
- Base temperature range is 25...100 °C; extending downward requires auxiliary cooling: 20...100 °C with water cooling, 15...100 °C with a refrigerator. Without either, the unit cannot regulate below ambient.
- The plastic pump is not resistant to acetone or polyglycols; these media require an alternative pump material.
- For freeze protection below ambient, water should be mixed with glysantine or ethylene glycol rather than used pure.
- Safety class III / FL with adjustable overtemperature protection per DIN 12876; alarm output is optical and acoustic.
Control, installation and operation
- KISS controller with OLED display; interfaces USB (Device) and RS232.
- Supply 200-240 V 1~/2~ 50/60 Hz, max current 10 A; external branch circuit protection per UL 489 is mandatory; fuse min. 10 A, max. 16 A.
- Voltage tolerance ±10% provided the frequency tolerance does not run in the opposite direction.
- IP20, ambient 5...40 °C; autostart restarts the unit automatically after a power failure.
- Optional Pt100 process sensor connection (#10688) is factory-fitted only and is for display, not control; optional M16x1 male pump connection (#19606).
Integration & Operating Notes
The datasheet requires external branch circuit protection according to UL 489 and specifies a minimum fuse of 10 A and maximum of 16 A on a 200-240 V 1~/2~ 50/60 Hz supply. Size the circuit and protective device to these limits; the unit draws up to 10 A.
The base range starts at 25 °C. Reaching 20...100 °C requires water cooling and 15...100 °C requires a refrigerator; without auxiliary cooling the bath cannot regulate below ambient, so the cooling method must be selected with the required low temperature.
The plastic pump is not resistant to acetone or polyglycols (depending on manufacturer). For these media an alternative pump material is required; for freeze protection below ambient, mix water with glysantine or ethylene glycol rather than using pure water.
The optional Pt100 sensor connection #10688 is available only factory-fitted and is used to display (not control) e.g. the process temperature. It cannot be added in the field and cannot close a control loop.
The autostart function restarts the unit automatically after a power failure. For unattended processes, confirm this behavior is acceptable or provide external sequencing, since the bath will resume heating without a manual restart.
The unit is IP20 and must be installed in an environment between 5 °C and 40 °C. Respect the space requirements at the installation site for access and ventilation; the IP20 rating excludes wet or dusty locations.
Selection & Integration Q&A
Pick the 106A when the sample load and immersion depth fit its 4.4 L bath and 130 x 110 mm opening. Step to a smaller sibling for less bath volume or a larger one when more volume or opening area is needed; all share the 25...100 °C base range.
The base range is 25...100 °C; 20...100 °C requires water cooling and 15...100 °C requires a refrigerator. For freeze protection, mix water with glysantine or ethylene glycol rather than using pure water.
The Pt100 process sensor connection #10688 is factory-fitted only and display-only; the M16x1 male pump connection #19606 is an option. Water cooling or a refrigerator is a separate selection needed for sub-ambient operation.