Product introduction and key features of KISS 202C
Specp. 2
Technical data according to DIN 12876 standard
p. 1table, 37 rows
Technical data according to DIN 12876
Parameter
KISS 202C
Operating temperature range
45...200 °C
with water cooling
20...200 °C
with refrigerator
-30...200 °C
Temperature stability at 70°C
0,05 K
temperature set point / display
digital
Internal temperature sensor
Pt100
Interface digital
USB (Device), RS232
Interface
Alarm message
optic, acoustic
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Heating power at 200V
1,5 kW
Pressure / Suction pump
max. delivery
14 l/min
max. delivery pressure
0,25 bar
max. delivery (suction)
10,5 l/min
max. delivery pressure (suction)
0,17 bar
Pump connection
M16x1 male
Bath volume
3,5 l
Filling capacity
2 l
Bath opening diameter
25 mm
Bath depth
150 mm
Height of bath opening
190 mm
Overall dimensions WxDxH **
178x260x355 mm
Net weight
8 kg
Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
N/A
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.
S393769
** Overall dimensions WxDxH
Engineering Facts
14 facts · every claim linked to a page
Selection
The operating temperature range depends on the cooling method chosen, so the required range dictates whether water cooling or a refrigerator is needed.
Determines whether the base unit suffices or a water-cooling loop or refrigerator option must be specified.
Source
“Operating temperature range 45...200 °C with water cooling 20...200 °C with refrigerator -30...200 °C”
Selection
The pump delivers up to 14 l/min with 0.25 bar pressure, which constrains the external circuit resistance the unit can drive.
Needed to check that the external loop pressure drop is compatible with the pump curve.