Peter Huber Kältemaschinenbau KISS 202C
Peter Huber Kältemaschinenbau KISS 202C — Thermostatic baths.
The KISS 202C is a 3.5 l stainless-steel bath immersion circulator with a non-variable pressure/suction pump (14 l/min, 0.25 bar) and 1.5–2.1 kW heating. Its 45–200 °C base range extends to 20–200 °C with water cooling or −30–200 °C with a refrigerator, positioning it as the high-temperature member of the KISS series for precise (±0.05 K) external loop tempering.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 200C Series — Model Comparison & Selection
| Model | Controller | Interface | Pump connection | Temperature range, °C | Weight, kg | Max delivery flow rate, liter/minute |
|---|---|---|---|---|---|---|
| KISS 202C Current | KISS | USB (Device), RS232 Inte | M16x1 AG male | 45–200 | 8 | 14 |
| CC-202C | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 45–200 | 9 | 22 |
| CC-205B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 45–200 | 9 | 27 |
| CC-208B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 25–200 | 10 | 27 |
| CC-212B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 11 | 22 |
| CC-215B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 25–200 | 12 | — |
| CC-220B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 14 | 27 |
| CC-225B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 16 | 27 |
| KISS 205B Badumwälzthermostat mit Edelstahlbad | KISS | USB (Device), RS232 Inte | M16x1 AG male | 45–200 | 9 | 14 |
| KISS 208B | KISS | USB (Device), RS232 Inte | M16x1 AG male (Option, # | 25–200 | 10 | 14 |
| KISS 212B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 11 | 14 |
| KISS 215B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 12 | 14 |
| KISS 220B | KISS | USB (Device), RS232 Inte | M16x1 AG male (Option, # | 25–200 | 14 | 14 |
| KISS 225B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 16 | 14 |
Full specifications
| Filling volume | 2 l |
| Maximum delivery flow rate | 14 l/min (pressure pump; 0, 25 bar) |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | no |
| Pump type | Pressure and suction pump, speed not variable |
| Temperature range | 45...200 °C with cooling unit: -30...200 °C · with water cooling: 20...200 °C |
| Temperature stability mfr | ±0, 05 K |
| Alarm message mfr | optic, acoustic |
| Dimensions (W x D x H) mfr | 178 x 260 x 355 mm |
| Heating capacity mfr | 1, 5 - 2, 1 kW |
| Delivery capacity suction pump max. mfr | 10, 5 l/min (suction pump; 0, 17 bar) |
| Bath opening WxD/BD mfr | ø = 25 / 150 |
| Bath volume mfr | 3, 5 l |
| Weight | 8 kg |
| Current consumption | 10 A |
| Power supply | 200-240V 1~/2~ 50/60Hz |
| Controller | KISS |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Immersion circulator with 3.5 l bath volume, 2 l filling capacity and a 25 mm bath opening; 178 × 260 × 355 mm, 8 kg.
- Stainless-steel bath construction (series designation "Mit Edelstahlbaedern bis 200 °C") supports the full 200 °C upper limit.
- Within the KISS/CC series, the 202C is the 200 °C-capable modification; sibling CC-202C/205B/208B/212B differ in size class, so select the 202C when the 3.5 l bath and 200 °C ceiling match the load, and a larger sibling when sample volume or immersion depth exceeds the 25 mm opening / 2 l filling.
- Controller: KISS with OLED display; interfaces USB (Device) and RS232.
Thermal behavior and media
- Base operating range 45–200 °C; 20–200 °C requires water cooling; −30–200 °C requires a refrigerator. The cooling method is a selection decision, not an accessory detail.
- Temperature stability ±0.05 K; heating capacity 1.5–2.1 kW.
- Pump: pressure and suction pump, speed not variable; max delivery 14 l/min at 0.25 bar, max suction 10.5 l/min at 0.17 bar; pump connection M16×1 male.
- The plastic pump is not resistant to acetone or polyglycols (manufacturer-dependent); for freeze protection, mix water with glysantine or ethylene glycol.
Installation, control and operation
- Supply 200–240 V 1~/2~ 50/60 Hz, max current 10 A, min fuse 10 A; external branch circuit protection to UL 489 required; country-specific plug must be specified at order.
- Voltage tolerance ±10% only if frequency tolerance does not run in the opposite direction.
- IP20 enclosure; ambient 5–40 °C; EN55011 Class A Group 1 (industrial EMC).
- Adjustable overtemperature protection per DIN 12876, safety class III/FL; optical and acoustic alarm; autostart after power failure must be assessed for unattended processes.
Integration & Operating Notes
The unit draws up to 10 A on a 200–240 V 1~/2~ 50/60 Hz supply and requires a minimum 10 A fuse plus external branch circuit protection according to UL 489; the panel must include a listed breaker, and the country-specific plug must be stated at order time.
Without auxiliary cooling the unit operates only from 45 °C upward; reaching 20 °C requires a water-cooling loop and reaching −30 °C requires a refrigerator, so the required process low point dictates which cooling configuration must be specified.
The non-variable plastic pump is not resistant to acetone or polyglycols (depending on manufacturer), so these heat transfer fluids must be excluded; for freeze protection water should be mixed with glysantine or ethylene glycol rather than used pure.
The autostart function restarts the unit automatically after a power failure; for sensitive or unattended processes this behavior may be unacceptable and must be evaluated during commissioning alongside the adjustable overtemperature cut-out per DIN 12876 (safety class III/FL).
With IP20 protection the unit must be installed away from water spray in a dry location, and ambient temperature must remain between 5 °C and 40 °C; the EN55011 Class A Group 1 classification means it is intended for industrial, not residential, electromagnetic environments.
Selection & Integration Q&A
Pick the 202C when a 3.5 l bath, 2 l filling and 25 mm opening suit the samples and 200 °C is the ceiling; move to a larger sibling when immersion depth or sample volume exceeds the 25 mm opening.
The loop must be drivable by 14 l/min at 0.25 bar through an M16×1 male connection; the supply is 200–240 V 1~/2~ 50/60 Hz with a 10 A fuse and UL 489 branch protection, and the plug type must be ordered.
Install in a dry, spray-free location at 5–40 °C ambient; IP20 and EN55011 Class A Group 1 restrict it to industrial settings, and the ±10% voltage tolerance applies only if frequency does not deviate oppositely.
The country-specific power plug must be specified at order; water cooling or a refrigerator is required to go below 45 °C; and heat transfer fluid (water with glysantine or ethylene glycol for freeze protection) is a separate consumable.