Product introduction and general description of the Ministat 230w circulator
Specpp. 1-2
Technical data, performance figures and operating limits per DIN 12876
p. 2table, 17 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
Consumption at water 15°C, flow 0°C
20,4 l/h
min. cooling water differential pressure
2 bar
max. cooling water pressure
6 bar
Bath volume
3,5 l
min. filling capacity
3 l
Bath capacity with displacement rack
1,7 l
Width bath opening WxD / bath depth
170 x 85 / 135 mm
Overall dimensions WxDxH **
255x450x476 mm
Net weight
36 kg
Power supply requirement
208-240V 1~/2~ 50/60Hz
max. current
11 A
Fuse
16 A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
max. ambient temperature
40 °C
min. ambient temperature
5 °C
from Serial-No.:
S393233
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876 — Ministat 230w (Order-No.: 2015.0007.01)
Characteristic
Value
Operating temperature range
-40...200 °C
Temperature stability at -10°C
0,02 K
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Cooling power at 100°C
0,42 kW
Cooling power at 20°C
0,42 kW
Cooling power at 0°C
0,38 kW
Cooling power at -10°C
0,33 kW
Cooling power at -20°C
0,25 kW
Cooling power at -30°C
0,14 kW
Cooling power at -40°C
0,05 kW
Refrigeration machine
water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)
Global Warming Potential (GWP)
0,02
Refrigerant quantity
0,06 kg
CO2 equivalent
0,0 t
UN-number
UN 3358
Pressure pump
yes
max. delivery
22 l/min
max. delivery pressure
0,7 bar
Suction pump
yes
max. delivery (suction)
16 l/min
max. delivery pressure (suction)
0,4 bar
Pump connection
M16x1 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G1/2 male
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires cooling water with a differential pressure between 2 bar and 6 bar, so the facility water supply must be sized accordingly.
Insufficient or excessive cooling water pressure will prevent proper operation or damage the cooling circuit.
Source
“min. cooling water differential pressure
2 bar
max. cooling water pressure
6 bar”
Installation
The cooling water intake temperature must be between 5°C and 20°C, and pipes must be insulated if the water can fall below the dew point.
Out-of-range cooling water temperature reduces cooling power and can cause condensation and corrosion.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water quality may be used.
Aggressive or incompatible cooling water will corrode or degrade these materials and cause leaks.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation
Voltage tolerance is ±5% only when frequency tolerance is simultaneously ±2%, and the combination of -5% voltage with +2% frequency is not allowed.
The supply must be checked against this combined tolerance rule before connecting the unit.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The unit is classified EN55011 Class A Group 1, so it is intended for industrial electromagnetic environments and may cause interference in residential areas.
Installation in a residential or sensitive EMC environment may require additional filtering or separation.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.
The correct plug type must be stated on the order to avoid a non-usable power connection.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Selection
Cooling power falls from 0.42 kW at 20°C to 0.05 kW at -40°C, so the cooling capacity at the lowest operating temperature is very limited.
The heat load at low temperature must be kept below this value or the set point cannot be reached.
Source
“at -40°C
0,05 kW”
Selection
The pump delivers up to 22 l/min at 0.7 bar pressure and suction up to 16 l/min at 0.4 bar, which limits the external circuit resistance that can be handled.
External loop pressure drop must be checked against the pump curve to ensure adequate flow.