Peter Huber Kältemaschinenbau Ministat 230w Kompakter Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau Ministat 230w Kompakter Kälte-Badumwälzthermostat — Thermostatic baths.
The Ministat 230w is a water-cooled refrigerated/heating bath circulator with a 3.5 l bath and a variable-speed pressure/suction pump, covering -40 to 200 °C with ±0.02 K stability. Its water-cooled condenser and G1/2 cooling-water connection distinguish it from air-cooled siblings in the Ministat series. It serves as a precise temperature source for external circuits and bath immersion in laboratory and industrial processes.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Ministat Series — Model Comparison & Selection
| Model | Max delivery flow rate, liter/minute | Power supply | Temperature range, °C | Weight, kg | Ambient temperature limit, °C | Controller |
|---|---|---|---|---|---|---|
| Ministat 230w Kompakter Kälte-Badumwälzthermostat Current | 16 | 208-240V 1~/2~ 50/60Hz | -40–200 | 36 | 5–40 | Pilot ONE |
| Ministat 125w Kompakter Kälte-Badumwälzthermostat | 22 | 110-120V 1~ 60Hz | -25–150 | 26 | 5–40 | Pilot ONE |
| Ministat 240w Kompakter Kälte-Badumwälzthermostat | 22 | 220-240V 1~/2~ 50/60Hz | -45–200 | 41 | 5–40 | Pilot ONE |
Full specifications
| Filling volume | 3 l (min.) |
| Maximum delivery flow rate | 22 l/min (pressure pump; 0, 7 bar) suction pump; 0, 4 bar: 16 l/min |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| Refrigerator | water-cooled, natural refrigerant |
| External sensor connection | Pt100 |
| Temperature range | -40...200 °C |
| Temperature stability mfr | ±0, 02 K |
| Dimensions (W x D x H) mfr | 255 x 450 x 476 mm |
| Heating capacity mfr | 1, 6 - 2, 1 kW |
| Bath opening WxD/BD mfr | 170 x 85 / 135 mm |
| Weight | 36 kg |
| Current consumption | 11 A (max.) |
| Fuse rating | 16 A |
| Power supply | 208-240V 1~/2~ 50/60Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Refrigerant | R-290 (A3, H220) (1st stage) |
| Maximum viscosity | 50 mm²/s |
| Refrigerant quantity 1st stage mfr | 0, 06 kg (1st stage) |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water differential pressure | 2 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Cooling capacity mfr | 100 0, 42; 20 0, 42; 0 0, 38; -10 0, 33; -20 0, |
| Cooling water consumption at 15°C, outlet 0 °C mfr | 20, 4 l/h (at 15°C, outlet 0 °C) |
| Safety class | III / FL |
Description
Construction and series positioning
- The Ministat 230w is the water-cooled variant of the Ministat 230, sharing the 3.5 l bath and Pilot ONE controller but rejecting heat to facility water rather than room air.
- Cooling circuit uses natural refrigerant R-290 (A3, H220) with a 0.06 kg charge; the condenser is water-cooled via G1/2 male connections.
- Bath opening measures 170 x 85 mm (W x D) with 135 mm depth; overall dimensions are 255 x 450 x 476 mm and weight is 36 kg.
- Safety classification III / FL per DIN 12876 with adjustable overtemperature protection.
Thermal behavior and media
- Working range spans -40 to 200 °C with ±0.02 K stability; heating capacity is 1.6–2.1 kW.
- Cooling capacity falls from 0.42 kW at 20 °C to 0.05 kW at -40 °C, so heat loads at the low end must be kept small.
- The variable-speed pressure/suction pump delivers up to 22 l/min at 0.7 bar pressure and 16 l/min at 0.4 bar suction, setting the allowable external circuit resistance.
- Bath fluid viscosity must not exceed 50 mm²/s; minimum filling is 3 l, and a displacement rack reduces usable volume to 1.7 l.
Installation, control and operation
- The Pilot ONE controller provides Ethernet, USB (host and device) and RS232 interfaces, plus a Pt100 external sensor connection.
- Power supply is 208–240 V, 1~/2~, 50/60 Hz with a 16 A fuse and 11 A maximum current draw; a country-specific plug must be specified at order.
- Cooling water must be supplied at 5–20 °C with differential pressure between 2 and 6 bar; circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.
- The unit is EN55011 Class A Group 1 and IP20, intended for industrial EMC environments and indoor installation at 5–40 °C ambient.
Integration & Operating Notes
Facility water supply must provide at least 2 bar differential pressure to actuate the water-cooled condenser, but must never exceed 6 bar static pressure or the cooling circuit can be damaged. Verify pressure at the G1/2 male connections before commissioning.
Water below 5 °C or above 20 °C degrades cooling performance and may cause condensation on uninsulated pipes. If the water can fall below the dew point, insulate all cooling-water piping to prevent corrosion and dripping.
The supply may deviate ±5% in voltage only when frequency deviation is simultaneously within ±2%; the combination of -5% voltage with +2% frequency is explicitly not allowed. Check the site supply against this rule before connecting the 208–240 V unit.
Single/two-phase devices are delivered with a power cable and a country-specific plug that must be specified at the time of ordering. Confirm the plug type for the destination country to avoid a non-usable power connection on delivery.
Heat load from samples or external circuits must remain below the available cooling capacity at the setpoint. At -40 °C only 0.05 kW is available, so large or warm loads will prevent the bath from reaching or holding the low-end setpoint.
Highly viscous fluids overload the pump and reduce circulation. Select bath fluid with a kinematic viscosity at the lowest operating temperature below 50 mm²/s, especially when running near -40 °C where many fluids thicken significantly.
The installation room must remain between 5 °C and 40 °C. Operation outside this range degrades cooling performance and may trip protection. The IP20 enclosure also requires a dry, dust-controlled indoor location.
The unit is classified EN55011 Class A Group 1, intended for industrial electromagnetic environments. In residential or sensitive EMC areas, additional filtering or physical separation from sensitive equipment may be required.
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Aggressive or incompatible water will corrode or degrade these materials and cause leaks; verify water quality before connecting the facility supply.
Selection & Integration Q&A
Choose the 230w when facility cooling water is available and heat rejection to room air is undesirable; it shares the 3.5 l bath and -40…200 °C range with the air-cooled 230. The 125w offers a smaller bath, while the 240w provides greater cooling capacity for higher heat loads.
The variable-speed pressure/suction pump delivers up to 22 l/min at 0.7 bar and suctions 16 l/min at 0.4 bar. Size external tubing and heat exchangers so pressure drop stays within these limits, and keep fluid viscosity below 50 mm²/s to avoid overloading the pump.
Install indoors at 5–40 °C ambient with IP20 protection. Cooling water must be 5–20 °C at 2–6 bar differential pressure. The unit is EN55011 Class A Group 1 for industrial EMC environments; residential use may require additional filtering.
The base unit includes a power cable with country-specific plug only if specified at order. Bath fluid, external Pt100 sensor, and displacement rack are not included. Cooling-water hoses and fittings for the G1/2 male connections must be supplied separately; no periodic maintenance schedule is given in the datasheet.