Product introduction and general description of the chiller unit
Specpp. 1-2
Technical data, performance figures and operating parameters per DIN 12876
p. 2table, 8 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
Overall dimensions WxDxH **
400x440x1230 mm
sound pressure level +/- 4 dB(A)
64 dB(A)
Power supply requirement
208V 2~ 60Hz
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/21
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876 — Unichiller 017Tw-H (Order-No.: 3013.0070.01)
Parameter
Value
Operating temperature range
-10...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power
1,6 kW
Cooling power at 15°C
1,7 kW
Cooling power at 0°C
0,9 kW
Cooling power at -10°C
0,4 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Circulation pump
B
max. delivery
30 l/min
max. delivery pressure
3,8 bar
Delivery at 0,12 bar
29 l/min
Delivery at 0,5 bar
25 l/min
Delivery at 1,0 bar
21 l/min
Delivery at 1,5 bar
17 l/min
Delivery at 2,0 bar
13,5 l/min
Delivery at 2,5 bar
10 l/min
Delivery at 3,0 bar
6,5 l/min
Pump connection
G3/4 male
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 15°C
53 l/h
Consumption at water 15°C, flow 0°C
40 l/h
Consumption at water 15°C, flow -10°C
25 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
2,5 l
Volume of expansion
4 l
Engineering Facts
12 facts · every claim linked to a page
Installation
The chiller is water-cooled and requires a cooling water supply with a minimum differential pressure of 1 bar and a maximum pressure of 6 bar.
Determines whether the site's cooling water loop can supply adequate pressure and whether a booster pump or pressure reducer is needed.
Source
“min. cooling water differential pressure 1 bar”
Installation
Cooling water inlet temperature must be between 5 °C and 20 °C, and pipes must be insulated if the water temperature can fall below the dew point.
Defines the acceptable cooling water supply conditions and the need for pipe insulation to prevent condensation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling capacity drops as cooling water temperature rises, and the required cooling water flow rate may increase.
Affects sizing of the cooling water supply and expected performance at the actual site conditions.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality and material compatibility checks to avoid corrosion or damage to the cooling circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Compliance
Voltage tolerance is ±5% only when frequency tolerance is simultaneously ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.
Ensures the electrical supply at the installation site stays within the permitted voltage and frequency window.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Three-phase devices drawing less than 63 A are supplied with a cable but without a plug, so a suitable plug must be provided on site.
Affects electrical installation planning and the need to procure and fit a plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The unit is rated for ambient temperatures from 5 °C to 40 °C.
Defines the allowable room temperature range where the chiller can be installed and operated.
Source
“min. ambient temperature 5 °C”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280, so refrigerant handling and ventilation precautions apply.
Informs safety measures for refrigerant handling and room ventilation during installation and service.
Source
“R-449A (A1, H280)”
Selection
The unit has a global warming potential of 1397, which matters for refrigerant regulations and environmental reporting.
Relevant for compliance with refrigerant regulations and for environmental impact assessment during selection.
Source
“Global Warming Potential (GWP) 1397”
Installation
The pump connection is G3/4 male and the cooling water connection is G1/2 male, so matching hose couplings are required.
Determines the fittings and hoses needed for process and cooling water connections.
Source
“Pump connection G3/4 male”
Operation
The minimum filling capacity is 2.5 l and the expansion volume is 4 l, which sets the minimum heat transfer fluid charge.
Needed to plan the heat transfer fluid volume and expansion tank capacity for the application.
Source
“min. filling capacity 2,5 l”
Installation
The unit has degree of protection IP20, so it must be protected from water and dust ingress.
Restricts installation location to dry, protected areas and affects enclosure or room requirements.