Peter Huber Kältemaschinenbau Unichiller 017Tw-H
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 017Tw-H

Peter Huber Kältemaschinenbau Unichiller 017Tw-H — Recirculating chillers with Noise level 64 dB.

The Unichiller 017Tw-H is a water-cooled recirculating chiller with a Pilot ONE controller, rated for -10 to 100 °C and 208 V 2~ 60 Hz. It is the 'Tw' (water-cooled) modification of the 017 platform, distinguished from the air-cooled 015 siblings by its G1/2 cooling-water circuit. It serves externally closed process loops requiring heat rejection to a facility cooling-water supply.

Noise level
64 dB

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBControllerSafety classTemperature range, °CInterfacePower supply
Unichiller 017Tw-H Current 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 015 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller 015-H OLÉ 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015w-H OLÉ 54OLÉI / NFL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 017T 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 017T-H 64Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020T —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020T-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 020Tw 60Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020Tw-H 60Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 022-H 65Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022w-H —Pilot ONEIII / FL-10–100——
Unichiller 022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025T 74Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025T-H 74Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025Tw-H —Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P007 58Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Unichiller P007-H —Pilot ONEIII / FL-20–100——
Unichiller P007-H OLÉ 58OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller P007 OLÉ 58OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller P010 57Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P010-H 57Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010-H OLÉ 57OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P010 OLÉ 57OLÉI / NFL-20–40USB (Device), RS232 Inte110-120V 1~ 50/60Hz
Unichiller P010w —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012 59Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P012-H 59Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012-H OLÉ 59OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012 OLÉ 59OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012w-H OLÉ —OLÉIII / FL-20–100USB (Device), RS232 Inte208V 2~ 60Hz
Unichiller P012w OLÉ —OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P015 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller P015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P015-H OLÉ 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P015 OLÉ 63OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P015w 54Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller P015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P015w-H OLÉ 54OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50Hz
Unichiller P022 65Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022-H 65Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P022w-H 60Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P025 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz

Full specifications

Core Performance & Functional Specifications
Noise level 64 dB(A)
Expansion volume 4 l
Pump connection G3/4 AG male
Pump speed adjustable no
Pump type B
External sensor connection Pt100
Temperature range -10...100 °C
Electrical, Power & Drive Specifications
Power supply 208V 2~ 60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 25 л/час (at 15°C, flow -10°C)
at 15°C, flow 0°C: 40 l/h · at 15°C, flow 15°C: 53 l/h
Cooling water differential pressure 1 bar
Cooling Water Pressure 6 bar
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Working range and behavior

Control, installation and operation

Integration & Operating Notes

Process Limits
Externally closed circuits only; no open baths

The manual restricts this chiller to externally closed applications, so the hydraulic circuit must be designed as a sealed loop with the process. Open-tank or atmospheric bath arrangements are outside the intended use and invalidate the thermal-fluid and safety-class assumptions.

Unichiller Operation Manual, Function Description, p. 34
Cooling & Media
Cooling water 5–20 °C, 1–6 bar differential

The facility loop must deliver intake water between 5 and 20 °C with at least 1 bar differential pressure and no more than 6 bar at the G1/2 connection. Below 1 bar the unit cannot reject rated capacity; above 6 bar the cooling-water circuit is overstressed. Insulate pipes where water can fall below dew point.

Unichiller 017Tw-H Technical Datasheet, p. 1–2
Cooling & Media
Shut off cooling water for overnight stops

Even short downtimes such as overnight require closing the cooling-water supply. Leaving it open invites condensation on the water-cooled assembly and corrosion or water damage in the enclosure, and the manual treats this as a standing operating rule rather than an optional measure.

Unichiller Operation Manual, Installation, p. 28
Mounting
Keep 50–200 cm clearance per side for service

Free-standing installation requires a wall clearance of 50 to 200 cm on either side to permit maintenance access. The unit is indoor-only, needs at least 300 lx illuminance and must be sited below 2000 m altitude; check refrigerant charge against room volume to stay within the limit per m³ of ambient air.

Unichiller Operation Manual, Installation, p. 25
Mounting
Protect cables below the unit from condensate

Because the unit is water-cooled, condensation can drip from the cooling-water circuit. Electrical lines routed directly below the chiller must be protected against liquid ingress, and the IP20 enclosure itself must be kept clear of water and dust. Plan cable trays accordingly.

Unichiller Operation Manual, Installation, p. 28
Electrical
Voltage and frequency tolerances are coupled

The ±5% voltage tolerance applies only together with a ±2% frequency tolerance, and combinations such as -5% voltage with +2% frequency are not permitted. Verify the site supply against this coupled window before connection; three-phase units under 63 A arrive with a cable but no plug, so a suitable plug must be provided on site.

Unichiller 017Tw-H Technical Datasheet, p. 2
Control & I/O
Choose power-failure response before commissioning

The Pilot ONE can be set to OFF/Standby, ON or Power Failure Auto after a mains loss. For exothermic processes, Process Safety mode instead releases maximum cooling when the overtemperature trigger is reached rather than switching all poles off. Decide the mode during commissioning, since it defines restart behavior of the process loop.

Unichiller Operation Manual, Normal Operation, p. 18–19
Maintenance
Refrigerant work restricted to approved contractors

Any work on the refrigerant circuit must be carried out by approved refrigeration/air-conditioning contractors. The refrigerant is R-449A (A1, H280, GWP 1397), so handling and ventilation precautions apply and service planning must account for certified personnel and refrigerant reporting obligations.

Unichiller Operation Manual, Maintenance, p. 17

Selection & Integration Q&A

When should I choose the 017Tw-H instead of the air-cooled Unichiller 015 siblings?

Pick the 017Tw-H when a facility cooling-water loop is available and room heat rejection is undesirable; it needs 5–20 °C water at 1–6 bar differential. Choose the 015 air-cooled siblings where no water loop exists, accepting their 63 dB(A) fan noise versus 64 dB(A) here.

What must the facility cooling-water supply and process piping provide for this model?

Cooling water at 5–20 °C, minimum 1 bar differential and maximum 6 bar at G1/2 male connections; flow demand is 25–53 l/h depending on flow temperature. Process side uses G3/4 AG male pump connections with matching couplings and counterheld tightening.

What installation and duty-cycle limits does the manual impose on the 017Tw-H?

Indoor use only, 5–40 °C ambient, at least 300 lx, below 2000 m, 50–200 cm side clearance, no ATEX zone. Leveling feet or parking brakes must be set before start-up, and cooling water must be shut off even overnight.

What maintenance and accessories are needed beyond the base delivery?

Refrigerant-circuit service only by approved contractors (R-449A, GWP 1397). Three-phase units under 63 A ship with cable but no plug, so a site plug is required; external Pt100 sensor and hose couplings for G3/4 and G1/2 are not included and must be sourced separately.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF
PDF
Unichiller 017Tw-H Technical Datasheet
Datasheet ·63,7 КБ ·2 pp.
View PDF