Induztr
Engineering Digest Document Viewer Related Models

Unichiller 100-H Technical Datasheet

Contents

  1. Overview p. 1
    Introduces the Unichiller 100-H chiller and its main features
  2. Spec pp. 1-2
    Performance, temperature, cooling capacity and dimensional data per DIN 12876
  3. p. 2 table, 9 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
sound pressure level +/- 4 dB(A)58 dB(A)
Power supply (3 Phase)460V 3~ 60Hz
max. current (3 Phase)13,5 A
Fuse (3 phase)3x16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S438249

Technical data according to DIN 12876 — Unichiller 100-H (Order-No.: 3059.0006.01)

ParameterValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power2 kW
Cooling power at ambient temperature +20°C
at 100°C10 kW
at 20°C10 kW
at 0°C9,7 kW
at -10°C6,7 kW
at -20°C4,6 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity1,1 kg
CO2 equivalent1,5 t
Circulation pump:E1
max. delivery56 l/min
max. delivery pressure3,8 bar
Delivery at 0,3 bar53 l/min
Delivery at 0,5 bar51 l/min
Delivery at 1,0 bar46 l/min
Delivery at 1,5 bar40 l/min
Delivery at 2,0 bar34 l/min
Delivery at 2,5 bar27 l/min
Delivery at 3,0 bar18 l/min
Pump connectionG1 1/4 male
min. filling capacity18 l
Volume of expansion48 l
Overall dimensions WxDxH **740x1160x1165 mm
Net weight284 kg

** см. сноску в оригинале (на странице не раскрыта).

Engineering Facts

14 facts · every claim linked to a page
Installation

The chiller requires a three-phase 460 V 60 Hz supply with a 3x16 A fuse, so the site electrical infrastructure must be provisioned accordingly.

Determines the electrical supply, breaker/fuse rating and cabling needed at the installation site.

Source
“Power supply (3 Phase) 460V 3~ 60Hz”
Installation

An external branch circuit protection according to UL 489 is required, so a suitable UL 489 listed breaker must be installed upstream.

Affects the electrical panel design and compliance of the installation.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so the customer must provide and wire the plug or connection.

Impacts site wiring preparation and the need to source a suitable plug/terminal connection.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The unit must be installed in an environment between 5 °C and 40 °C ambient, so the room climate must be kept within this range.

Room conditions outside this range can prevent proper operation and affect cooling performance.

Source
“min. ambient temperature 5 °C”
Performance

Cooling capacity is rated at 20 °C room temperature and may drop if ambient temperature rises, so capacity margins must account for the actual room temperature.

Prevents undersizing the chiller when the installation room runs warmer than 20 °C.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Installation

The unit has IP20 protection, so it must be installed in a dry indoor location protected from water and dust ingress.

Restricts where the chiller can be placed and what environmental protection is needed.

Source
“Degree of Protection IP20”
Installation

The pump connection is G1 1/4 male, so hoses and fittings must match this thread size.

Determines the hose couplings and adapters required for the external circuit.

Source
“Pump connection G1 1/4 male”
Selection

The unit is intended for externally closed applications, so the process loop must be a closed circuit.

Confirms the chiller is not suitable for open-loop or once-through cooling setups.

Source
“For externally closed applications.”
Selection

The minimum filling capacity is 18 l and expansion volume is 48 l, so the external circuit volume must be considered when sizing the system.

Helps verify the external loop volume and heat transfer fluid quantity needed.

Source
“min. filling capacity 18 l”
Safety

The refrigerant is R-449A with GHS classification H280, so refrigerant handling and ventilation precautions apply.

Affects safety planning, room ventilation and refrigerant handling procedures.

Source
“R-449A (A1, H280)”
Compliance

The equipment is classified Class A, Group 1 for electromagnetic disturbance, so it may require EMC measures in residential or sensitive environments.

Class A equipment is not intended for residential use without additional EMC precautions.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively, but not simultaneously in the unfavorable combination, so supply quality must be checked.

Ensures the site power quality stays within the allowed envelope to avoid damage or malfunction.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation

The chiller has a sound pressure level of 58 dB(A) ±4 dB(A), so acoustic planning is needed for the installation room.

Relevant for noise-sensitive environments and room placement.

Source
“sound pressure level +/- 4 dB(A) 58 dB(A)”
Maintenance

The refrigeration machine is air-cooled, so adequate airflow and clearance around the unit are required for heat rejection.

Air-cooled units need unobstructed ventilation and space for maintenance access.

Source
“air-cooled, CFC- and HCFC-free”