Peter Huber Kältemaschinenbau Unichiller 100-H
Peter Huber Kältemaschinenbau Unichiller 100-H — Recirculating chillers with Noise level 58 dB.
The Unichiller 100-H is an air-cooled recirculating chiller for externally closed circuits, rated -20 to 100 °C with 2 kW heating and a fixed-speed E1 pump delivering up to 56 l/min at 3.8 bar. Its 48 l expansion volume and 18 l minimum fill distinguish it within the Unichiller series, and it requires a 460 V 3~ 60 Hz supply.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Pump connection | Pump type | Safety class | Temperature range, °C | Power supply |
|---|---|---|---|---|---|---|
| Unichiller 100-H Current | 58 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 030T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 460V 3~ 60Hz |
| Unichiller 030Tw | 61 | G3/4 AG male | B | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 040T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040T-H | 69 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 040Tw | 70 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040Tw-H | 70 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 050-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 050w-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 055T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 055T-H | 74 | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 055Tw | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 060Tw-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 070T-H | 66 | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 070Tw-H | — | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 075w | — | G1 1/4 AG male | E1 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 075w-H | — | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 100T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100Tw | 65 | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller GL 035w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 050w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075 | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 100w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050 | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w-H | 62 | G1 1/4 AG male | E2 | III / FL | -20–100 | — |
| Unichiller P075 | 61 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100 | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P100-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100w | 58 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P100w-H | 58 | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
Full specifications
| Noise level | 58 dB(A) |
| Expansion volume | 48 l |
| Heating capacity | 2 kW |
| Maximum delivery flow rate | 56 l/min (also 3, 8 bar pump pressure) |
| Minimum fill volume | 18 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | E1 |
| External sensor connection | Pt100 |
| Temperature range | -20...100 °C |
| Weight | 284 kg |
| Power supply | 460V 3~ 60Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Air-cooled refrigeration circuit using R-449A (A1, H280), CFC- and HCFC-free; no water cooling connection is required.
- Fixed-speed pump type E1 with G1 1/4 AG male connections; pump speed is not adjustable.
- 48 l expansion volume and 18 l minimum filling capacity set this model apart from smaller Unichiller siblings such as the 030T/030Tw variants, which are rated at higher noise levels (61–69 dB(A)) and do not share this hydraulic volume.
- Controller: Pilot ONE; interfaces Ethernet, USB (Host and Device), RS232; external Pt100 sensor input.
Behavior and process limits
- Working range -20 to 100 °C; heating capacity 2 kW.
- Cooling capacity is rated at 20 °C room temperature and may drop as ambient temperature rises; the permissible ambient range is 5 to 40 °C.
- Intended only for externally closed applications; not for open-loop or once-through cooling.
- Safety class III / FL per DIN 12876, with identification depending on whether the thermal fluid is combustible.
- Process Safety mode releases maximum cooling capacity when the overtemperature trigger value is reached, instead of switching off all poles.
Installation, connection and control
- Requires 460 V 3~ 60 Hz supply with 3x16 A fusing; external branch circuit protection to UL 489 must be provided upstream.
- Three-phase devices below 63 A are delivered with a cable but without a plug; the customer supplies and wires the plug or connection.
- Voltage tolerance ±5% and frequency tolerance ±2%, not simultaneously in the unfavorable combination.
- IP20; indoor use only, illuminance at least 300 lx, altitude up to 2000 m; not for ATEX zones.
- Wall clearance of 50 to 200 cm per side is required for maintenance on free-standing models; leveling feet or parking brakes must be set before start-up.
- Power-failure response configurable via Pilot ONE as OFF/Standby, ON, or Power Failure Auto.
- Sound pressure level 58 dB(A) ±4 dB(A); Class A, Group 1 per EN 55011.
Integration & Operating Notes
The datasheet requires external branch circuit protection according to UL 489. The site must include a listed breaker sized for the 3x16 A three-phase supply; the unit is delivered with a cable but without a plug, so the customer must also provide and wire the connection.
Free-standing models require a wall clearance of 50 to 200 cm per side for maintenance. Because the refrigeration machine is air-cooled, this clearance also supports heat rejection; the unit must be indoors at 5–40 °C ambient and at least 300 lx illuminance.
The chiller is designed for externally closed applications. The external loop volume must be considered against the 18 l minimum filling capacity and 48 l expansion volume; open-loop or once-through cooling is not permitted.
The Pilot ONE allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached rather than switching off all poles; the chosen behavior must match the process risk.
Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The refrigerant is R-449A (A1, H280); observe the charge and room size at installation to avoid exceeding the refrigerant limit per m³ of ambient air.
Voltage tolerance is ±5% with a simultaneous frequency tolerance of ±2%, but not in the unfavorable combination. Class A, Group 1 equipment per EN 55011 may require EMC measures in residential or sensitive environments.
Selection & Integration Q&A
Pick the 100-H when you need the larger 48 l expansion volume and 18 l minimum fill with a -20 to 100 °C range and 2 kW heating. The 030T/030Tw siblings are louder (61–69 dB(A)) and do not share this hydraulic volume; choose them only if their capacity matches your load.
A 460 V 3~ 60 Hz supply with 3x16 A fusing and external branch circuit protection to UL 489. The unit arrives with a cable but no plug, so you must supply and wire the plug or terminal connection; verify voltage within ±5% and frequency within ±2%.
Indoors only, 5–40 °C ambient, at least 300 lx, up to 2000 m altitude, IP20 dry location, not in an ATEX zone. Provide 50–200 cm wall clearance per side for maintenance and airflow, and set leveling feet or parking brakes before start-up.
The base unit ships without a plug and without thermal fluid; the external circuit must be filled to at least 18 l. Refrigerant circuit service must be performed by approved contractors. Shut off cooling water even for overnight downtime if a water-cooled variant is used; this air-cooled model needs no water supply.