Unichiller 180Tw-H Technical Datasheet

Contents

  1. Overview p. 1
    Introduces the Unichiller 180Tw-H chiller unit
  2. Spec pp. 1-2
    Performance, temperature, and dimensional data per DIN 12876
  3. p. 2 table, 15 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity15 l
Volume of expansion22 l
Overall dimensions WxDxH **760x800x1615 mm
Net weight401 kg
sound pressure level +/- 4 dB(A)59 dB(A)
Power supply (3 Phase)400V 3~ 50Hz
max. current24 A
Fuse (3 phase)3x25A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S457845

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unichiller 180Tw-H (Order-No.: 3019.0044.01)

ParameterValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power2 kW
Cooling power at 15°C18 kW
Cooling power at 0°C18 kW
Cooling power at -10°C13 kW
Cooling power at -20°C6 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity2,6 kg
CO2 equivalent3,6 t
Circulation pumpD3
max. delivery210 l/min
max. delivery pressure4,7 bar
Delivery at 1,1 bar202 l/min
Delivery at 1,5 bar193 l/min
Delivery at 2,0 bar179 l/min
Delivery at 2,5 bar159 l/min
Delivery at 3,0 bar140 l/min
Delivery at 3,5 bar115 l/min
Delivery at 4,0 bar77 l/min
Delivery at 4,5 bar31 l/min
Pump connectionG1 1/4 male
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C666 l/h
Consumption at water 15°C, flow -10°C553 l/h
Consumption at water 15°C, flow -20°C427 l/h

Engineering Facts

14 facts · every claim linked to a page
Installation

The chiller requires cooling water supply pressure between 1 bar minimum differential and 6 bar maximum, so the building water system must be sized within this window.

Incorrect cooling water pressure will prevent safe commissioning or damage the unit.

Source
“min. cooling water differential pressure 1 bar max. cooling water pressure 6 bar”
Installation

The unit is water-cooled and must be fed with cooling water between 5°C and 20°C; below the condensation point the cooling water pipes inside the unit must be insulated.

Condensation on uninsulated pipes can cause water damage and electrical hazards.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling capacity is rated at 18 kW at 15°C and 0°C but drops to 13 kW at -10°C and 6 kW at -20°C, so low-temperature duty needs a larger unit or derating.

Sizing must account for the actual process temperature, not just the nominal 18 kW.

Source
“at 15°C 18 kW at 0°C 18 kW at -10°C 13 kW at -20°C 6 kW”
Installation

The unit needs a 400 V 3-phase 50 Hz supply with a 3x25 A fuse and draws up to 24 A, so the electrical feed and protection must be sized accordingly.

Under-sized supply or fusing will trip or overheat the installation.

Source
“Power supply (3 Phase) 400V 3~ 50Hz max. current 24 A Fuse (3 phase) 3x25A”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide and fit the site plug.

Missing plug must be planned before commissioning to avoid downtime.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively but must not be combined in opposite directions, so the site supply quality must be verified.

Out-of-tolerance supply voids performance and may damage the compressor.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

The refrigerant R-449A is classified A1/H280 with GWP 1397 and 2.6 kg charge (3.6 t CO2e), so leak detection and F-gas record keeping apply.

F-gas regulations and room ventilation requirements depend on charge and GWP.

Source
“R-449A (A1, H280) Global Warming Potential (GWP) 1397 Refrigerant quantity 2,6 kg CO2 equivalent 3,6 t”
Installation

The unit is IP20 and must be installed at ambient temperatures between 5°C and 40°C, so it cannot be placed in wet or hot plant rooms.

Wrong ambient conditions reduce cooling capacity and may void warranty.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

The unit weighs 401 kg and measures 760x800x1615 mm, so floor loading and access routes must be checked before delivery.

Handling and rigging must be planned; the footprint and weight affect floor and door clearance.

Source
“Overall dimensions WxDxH ** 760x800x1615 mm Net weight 401 kg”
Operation

The pump delivers 210 l/min at zero head but only 31 l/min at 4.5 bar, so external circuit pressure drop must be matched to the pump curve.

A high-pressure-drop external loop will starve flow and reduce heat transfer.

Source
“max. delivery 210 l/min max. delivery pressure 4,7 bar Delivery at 4,5 bar 31 l/min”
Operation

The expansion tank volume is 22 l and minimum filling capacity is 15 l, so the external loop volume plus unit must be checked against these limits.

Too small a system volume causes poor temperature stability; too large exceeds expansion capacity.

Source
“min. filling capacity 15 l Volume of expansion 22 l”
Compliance

The unit is EMC Class A Group 1, so it is intended for industrial environments and may cause interference in residential areas.

Class A equipment may not be suitable for domestic or light-industrial locations without filtering.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water quality and treatment may be used.

Aggressive or untreated water will corrode or foul these materials and reduce heat transfer.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Selection

Cooling water consumption rises as process temperature falls: 666 l/h at 0°C, 553 l/h at -10°C and 427 l/h at -20°C at 15°C water inlet.

Facility cooling water capacity must cover the worst-case flow at the required process temperature.

Source
“Consumption at water 15°C, flow 0°C 666 l/h Consumption at water 15°C, flow -10°C 553 l/h Consumption at water 15°C, flow -20°C 427 l/h”