Peter Huber Kältemaschinenbau Unichiller 180Tw-H
Peter Huber Kältemaschinenbau Unichiller 180Tw-H — Recirculating chillers with Noise level 59 dB.
The Unichiller 180Tw-H is a water-cooled recirculating chiller with a 400 V 3-phase supply, delivering 18 kW at 0 °C and 6 kW at -20 °C over a -20…100 °C range. Its D3 pump provides up to 210 l/min, distinguishing it from lower-capacity siblings. It serves externally closed process loops requiring precise temperature control.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller 180Tw-H Current | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Noise level | 59 dB(A) |
| Expansion volume | 22 l |
| Heating capacity | 2 kW |
| Minimum fill volume | 15 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | D3 |
| External sensor connection | Pt100 |
| Temperature range | -20...100 °C |
| Weight | 401 kg |
| Current consumption | 24 A (max.) |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity | 15 kW (at 15 °C) at 0 °C: 18 kW · at -10 °C: 13 kW · at -20 °C: 6 kW |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 427 л/час (at 15°C, flow -20°C) at 15°C, flow -10°C: 553 л/час · at 15°C, flow 0°C: 666 l/h |
| Cooling water differential pressure | 1 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and positioning
- Water-cooled refrigeration circuit using R-449A (2.6 kg, GWP 1397) with a 2 kW heater and 22 l expansion volume.
- Pump type D3 delivers up to 210 l/min at zero head, falling to 31 l/min at 4.5 bar; pump speed is not adjustable.
- Compared to sibling models (e.g., Unichiller 110T), the 180Tw-H offers higher cooling capacity and lower noise (59 dB(A) vs 74 dB(A)).
- Controller: Pilot ONE with Ethernet, USB (host and device), and RS232 interfaces; external Pt100 sensor input.
Behavior under load
- Cooling capacity is 18 kW at 0 °C and 15 °C, 13 kW at -10 °C, and 6 kW at -20 °C, requiring derating for low-temperature duty.
- Cooling water consumption at 15 °C inlet: 666 l/h at 0 °C process, 553 l/h at -10 °C, and 427 l/h at -20 °C.
- The unit is designed for externally closed applications only; models with 'P' in the name are for high pressure drop, but this model lacks that suffix.
Installation and operation
- Requires 400 V 3~ 50 Hz supply with 3×25 A fuse; maximum current 24 A. Delivered with cable but without plug.
- Cooling water connections: G3/4 male; pressure range 1 bar min differential to 6 bar max. Water inlet temperature 5…20 °C; insulate pipes if below condensation point.
- Ambient temperature 5…40 °C, IP20, indoor use only with illuminance ≥300 lx and altitude ≤2000 m.
- Wall clearance 50–200 cm per side for maintenance; leveling feet or parking brakes must be engaged before operation.
- Power failure response configurable via Pilot ONE: OFF/Standby, ON, or Power Failure Auto. Process Safety mode releases maximum cooling on overtemperature trigger.
Integration & Operating Notes
Supply pressure below 1 bar differential prevents safe commissioning; above 6 bar may damage the unit. Size the building water system and any pressure-reducing valve to maintain this window under all flow conditions.
Free-standing models require this clearance on either side to allow service access. Plan the installation footprint accordingly, considering the unit dimensions of 760×800×1615 mm and weight of 401 kg.
Minimum filling capacity is 15 l; expansion volume is 22 l. Too small a volume causes poor temperature stability, while exceeding 22 l may overflow the expansion tank. Calculate the total volume of the unit plus external circuit.
Leaving the cooling water supply on during overnight or short downtimes can cause water damage and corrosion. Install a shut-off valve in the cooling water supply line and close it whenever the unit is not in use.
Three-phase devices below 63 A are supplied with a cable but no plug. The customer must provide and fit a suitable plug (3×25 A fuse required) before commissioning. Plan for this in the installation schedule.
Selection & Integration Q&A
Choose the 180Tw-H for higher cooling capacity (18 kW at 0 °C vs lower in 110 series) and lower noise (59 dB(A) vs 74 dB(A) for 110T). The 110Tw offers 65 dB(A) but less capacity. Select the 180Tw-H when process loads exceed 13 kW at -10 °C or when acoustic limits are tight.
Cooling water: G3/4 male connections, supply pressure 1–6 bar differential, inlet temperature 5–20 °C, flow up to 666 l/h at 0 °C process. Electrical: 400 V 3~ 50 Hz, 24 A max, 3×25 A fuse, cable provided but no plug. Ensure voltage/frequency tolerances (±5%/±2%) are not combined oppositely.
Install indoors only, ambient 5–40 °C, IP20, illuminance ≥300 lx, altitude ≤2000 m. Maintain 50–200 cm wall clearance per side. Engage leveling feet or parking brakes before operation. Not for ATEX zones. Shut off cooling water during downtimes. External loop volume must be 15–22 l.
Refrigerant circuit work only by certified contractors; F-gas records required (R-449A, 2.6 kg, 3.6 t CO2e). Cooling water circuit materials (copper, stainless steel, etc.) require compatible water treatment. Base delivery excludes plug (customer must provide) and external hoses/fittings. Regular maintenance includes checking cooling water quality and pipe insulation.