Unichiller GL 1300w Datasheet

Contents

  1. Overview p. 1
    Introduces the Unichiller GL 1300w chiller unit
  2. Spec pp. 1-2
    Performance, temperature, and dimensional data per DIN 12876
  3. p. 2 table, 3 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/26

Unichiller GL 1300w — Technical data according to DIN 12876

ParameterValue
Order-No.5514.0001.01
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at 20°C130 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-744 (A1, H280)
Global Warming Potential (GWP)1
UN-numberUN 2857
max. delivery320 l/min
max. delivery pressure4 bar
Delivery at 0,5 bar305 l/min
Delivery at 1,0 bar285 l/min
Delivery at 1,5 bar260 l/min
Delivery at 2,0 bar240 l/min
Delivery at 2,5 bar215 l/min
Delivery at 3,0 bar185 l/min
Delivery at 3,5 bar150 l/min
Delivery at 4,0 bar110 l/min
Delivery at 4,5 bar50 l/min
Pump connectionG1 1/4 male
Cooling water connectionG1 1/4 male
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity17,5 l
Volume of expansion50 l
Overall dimensions WxDxH **918x963x1771 mm
Power supply (3 Phase)400V 3~ 50Hz
Pressure equipment categoryII
Degree of ProtectionIP20

** Overall dimensions WxDxH

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller requires cooling water with a differential pressure of at least 1 bar and must not exceed 6 bar, so the site cooling water supply must be verified before installation.

Insufficient differential pressure will prevent proper cooling; excessive pressure may damage the cooling circuit.

Source
“min. cooling water differential pressure 1 bar”
Installation

The unit is water-cooled and requires a cooling water intake temperature between 5°C and 20°C, so the available cooling water temperature must be checked at the installation site.

Cooling water outside this range will reduce cooling capacity or cause condensation issues.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

If cooling water temperature falls below the condensation point, the cooling water pipes in the unit must be insulated to prevent condensation.

Condensation can cause water damage and electrical hazards; insulation is a mandatory installation measure.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling power drops as cooling water temperature increases, and the required cooling water flow rate may increase, so performance must be derated accordingly.

Engineers must account for reduced cooling capacity and higher water flow when specifying the cooling water supply.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Incompatible water quality can corrode or damage these materials, leading to leaks and failures.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Compliance

Voltage tolerance is ±5% with a simultaneous frequency tolerance of ±2%, but combinations such as -5% voltage with +2% frequency are not allowed.

The power supply must be within these tolerances to avoid malfunction or damage; certain combinations are explicitly prohibited.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

The equipment has a leakage current greater than 3.5 mA, which must be considered for protective earthing and residual current device selection.

High leakage current requires proper earthing and may affect the choice of RCD type to ensure electrical safety.

Source
“Attention: leakage current > 3,5mA”
Installation

Three-phase devices with current consumption less than 63A are delivered with a cable but without a plug, so a suitable plug must be provided on site.

The installer must supply and connect the appropriate plug, affecting installation planning and cost.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The unit requires minimum ambient temperature of 5°C and maximum of 40°C, so the installation location must be within this range.

Operating outside the ambient temperature range can affect performance and reliability.

Source
“min. ambient temperature 5 °C”
Selection

The chiller uses R-744 (CO2) refrigerant with GWP 1, which is a natural refrigerant with low global warming potential.

Refrigerant type affects environmental compliance, safety classification, and handling requirements.

Source
“Refrigerant (ASHRAE, GHS) R-744 (A1, H280)”
Installation

The pump connection and cooling water connection are both G1 1/4 male, so mating fittings must be selected accordingly.

Correct connection thread ensures proper hookup and prevents leaks.

Source
“Pump connection G1 1/4 male”
Operation

The unit has a digital level indicator and an atmospheric open expansion tank, so the filling level must be monitored and maintained.

Low fluid level can cause pump cavitation and cooling failure; the indicator helps maintain proper operation.

Source
“With digital level indicator.”