Peter Huber Kältemaschinenbau Unichiller GL 1300w
Peter Huber Kältemaschinenbau Unichiller GL 1300w — Recirculating chillers.
The Unichiller GL 1300w is a water-cooled recirculating chiller delivering 130 kW at 20 °C, with a -20…40 °C temperature range and a fixed-speed pressure pump. It is a member of the Unichiller GL series, positioned below the GL 1650w, and is intended for externally closed process loops requiring high cooling power from a 400 V three-phase supply.
Peter Huber Kältemaschinenbau Unichiller GL Series — Model Comparison & Selection
| Model | Cooling capacity, W | Ambient temperature limit, °C | Controller | Cooling water connection | Cooling water differential pressure, bar | Cooling water pressure, psi |
|---|---|---|---|---|---|---|
| Unichiller GL 1300w Current | 130000 | 5–40 | Pilot ONE | G1 1/4 male | 1 | 87.02 |
| Unichiller GL 1650w | 165000 | 5–40 | Pilot ONE | G1 1/4 male | 1 | 87.02 |
Full specifications
| Expansion volume | 50 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | Pressure pump, speed not variable |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity | 130 kW (at 20 °C) |
| Cooling water connection | G1 1/4 male |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- Water-cooled refrigeration circuit using R-744 (CO2, GWP 1) with an atmospheric open expansion tank (50 l expansion volume) and digital level indicator.
- Fixed-speed pressure pump (not speed-variable) with G1 1/4 male pump connection; cooling-water circuit also G1 1/4 male.
- Pilot ONE controller with Ethernet, USB (host and device) and RS232 interfaces; external Pt100 sensor input.
- IP20 protection, safety class I / NFL, designed for indoor installation only.
- Within the series, the GL 1300w sits below the GL 1650w; the manual notes that models with “P” in the model name are specifically suited to high-pressure-drop applications, which this model is not.
Behavior under load and media
- Nominal cooling capacity is 130 kW at 20 °C; the datasheet states that cooling power drops as cooling-water temperature rises, and the required cooling-water flow rate may increase, so capacity must be derated against the site cooling-water temperature.
- Cooling-water intake must be between 5 °C and 20 °C; if it falls below the condensation point, the cooling-water pipes inside the unit must be insulated.
- The cooling-water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM; only compatible water quality may be used.
- The unit is intended only for externally closed applications.
Installation, control and operation
- Requires 400 V 3~ 50 Hz supply; three-phase devices under 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site.
- Voltage tolerance ±5% with simultaneous frequency tolerance ±2%, but combinations such as -5% voltage with +2% frequency are not allowed.
- Leakage current exceeds 3.5 mA, which must be considered for protective earthing and RCD selection.
- The Pilot ONE allows configuration of the power-failure response (OFF/Standby, ON, or Power Failure Auto) and a Process Safety mode that releases maximum cooling capacity when the overtemperature trigger is reached.
- Leveling feet must be activated and parking brakes applied before startup; cooling water must be shut off even for short downtimes.
Integration & Operating Notes
Insufficient differential pressure prevents proper cooling, while pressure above 6 bar may damage the cooling circuit. Verify the site cooling-water supply against both limits before installation; the unit cannot compensate for a low-pressure supply with its fixed-speed pump.
Three-phase devices with current consumption below 63 A are delivered with a cable but without a plug, so the installer must supply and connect a suitable plug. The equipment has a leakage current greater than 3.5 mA, which must be considered for protective earthing and residual current device selection.
Voltage tolerance is ±5% with a simultaneous frequency tolerance of ±2%, but combinations such as -5% voltage with +2% frequency are not allowed. The supply must be 400 V 3~ 50 Hz and remain within the permitted combined envelope to avoid malfunction or damage.
Because the unit is water-cooled, electrical lines directly below it must be protected against liquid ingress from condensation. This affects cable routing and the placement of adjacent equipment in the installation plan.
The cooling water supply must be shut off even for short downtimes such as overnight to prevent water damage and corrosion during idle periods. Before switching on, leveling feet must be unscrewed/activated and parking brakes applied at the casters to prevent the unit from moving.
Selection & Integration Q&A
Pick the GL 1300w when 130 kW at 20 °C matches your load and the process loop has normal pressure drop. The GL 1650w offers higher capacity; models with 'P' in the name are specifically suited to applications requiring high pressure drops, which the GL 1300w is not.
Provide cooling water at 5–20 °C with a differential pressure of at least 1 bar and no more than 6 bar. Both the cooling-water and pump connections are G1 1/4 male, so mating fittings must match; insulate internal cooling-water pipes if the water falls below the condensation point.
Install indoors only, at 5–40 °C ambient, at least 300 lx illuminance, up to 2000 m altitude, with 50–200 cm side clearance for maintenance. It must not be placed in an ATEX zone. Shut off cooling water even for overnight downtime and secure leveling feet before startup.
Three-phase units under 63 A arrive with a cable but no plug, so a suitable plug must be supplied on site. Refrigerant circuit work must be performed by approved refrigeration/air-conditioning contractors; the unit uses R-744 (CO2) and the charge must be checked against room size at installation.