Unichiller P015-H OLÉ Datasheet

Contents

  1. Overview p. 1
    Product introduction and key features of the chiller
  2. Spec pp. 1-2
    Technical data and performance specifications per DIN 12876
  3. p. 2 table, 4 rows
  4. p. 1 table, 38 rows

Technical data according to DIN 12876

ParameterValue
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S508491

Technical data according to DIN 12876

ParameterValue
Operating temperature range-20...100 °C
temperature set point / displaydigital
Internal temperature sensorPt100
Resolution of display0,1 K
Interface digitalUSB (Device), RS232
Interface
Temperature stability at -10°C0,2 K
Alarm messageoptic, acoustic
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at 15°C1,5 kW
Cooling power at 0°C1 kW
Cooling power at -10°C0,7 kW
Cooling power at -20°C0,3 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,35 kg
CO2 equivalent0,5 t
Circulation pumpB
at 0,5 bar21 l/min
at 1,0 bar17 l/min
at 1,5 bar11 l/min
at 2,0 bar6 l/min
max. delivery25 l/min
max. delivery pressure2,5 bar
Pump connectionG3/4 male
min. filling capacity3,8 l
expansion tank1,7 l
Overall dimensions WxDxH **420x487x579 mm
Net weight65 kg
sound pressure level +/- 4 dB(A)63 dB(A)
Power supply requirement220-240V 1~ 50Hz
max. current15 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is rated for an operating temperature range of -20 to 100 °C, so the process must stay within these limits.

Defines the application window and excludes processes outside this range.

Source
“Operating temperature range -20...100 °C”
Performance

Cooling capacity drops from 1.5 kW at 15 °C to 0.3 kW at -20 °C, so low-temperature duty needs careful heat-load sizing.

Prevents undersizing the chiller for cold set points.

Source
“at 15°C 1,5 kW at 0°C 1 kW at -10°C 0,7 kW at -20°C 0,3 kW”
Installation

The chiller is air-cooled and must be installed with sufficient space for air flow and service access.

Air-cooled units need clearance; ignoring this causes overheating and reduced capacity.

Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Installation

Ambient temperature must remain between 5 °C and 40 °C, and capacity may fall if ambient rises.

Site conditions directly affect cooling performance and reliability.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

The power supply must be 220–240 V single-phase 50 Hz with a 16 A fuse and 15 A maximum current.

Determines electrical supply, breaker sizing and cabling for installation.

Source
“Power supply requirement 220-240V 1~ 50Hz max. current 15 A Fuse 16 A”
Installation

Voltage and frequency tolerances are ±5% and ±2% but must not be combined in opposite directions.

Prevents installation on unstable mains that could damage the compressor or pump.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Single-phase devices are supplied with a power cable and country-specific plug, which must be specified at order.

Avoids missing or wrong plug on delivery and delays at commissioning.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety

The refrigerant is R-449A with GHS hazard H280, so leak and ventilation precautions apply.

Affects site safety assessment and refrigerant handling requirements.

Source
“R-449A (A1, H280)”
Compliance

The unit is EMC Class A Group 1, which is intended for industrial environments and may cause radio interference in residential areas.

Determines whether the installation location is electrically compatible.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Operation

The pump delivers 21 l/min at 0.5 bar but only 6 l/min at 2.0 bar, so system pressure drop must be checked.

Ensures the pump can overcome the external circuit pressure drop.

Source
“at 0,5 bar 21 l/min at 1,0 bar 17 l/min at 1,5 bar 11 l/min at 2,0 bar 6 l/min”
Installation

The pump connection is G3/4 male, so mating hoses and fittings must match this thread.

Prevents mismatch of process connections during installation.

Source
“Pump connection G3/4 male”
Operation

Minimum filling capacity is 3.8 l and the expansion tank is 1.7 l, which sets the heat-transfer fluid charge.

Needed to prepare the correct fluid volume and expansion allowance.

Source
“min. filling capacity 3,8 l expansion tank 1,7 l”
Operation

The unit has an autostart function for power failure, so it may restart automatically after mains recovery.

Affects risk assessment and process safety after power interruptions.

Source
“Autostart function for power failure”
Maintenance

A drain valve is optional, so draining for service or fluid change requires ordering it separately.

Impacts maintenance planning and fluid replacement procedure.

Source
“drain valve #6839”