Peter Huber Kältemaschinenbau Unichiller P015-H OLÉ
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller P015-H OLÉ

Peter Huber Kältemaschinenbau Unichiller P015-H OLÉ — Recirculating chillers with Noise level 63 dB.

The Unichiller P015-H OLÉ is an air-cooled, single-phase recirculating chiller with a 'P' designation for high-pressure-drop process loops, rated -20 to 100 °C with cooling capacity falling from 1.5 kW at 15 °C to 0.3 kW at -20 °C. It is the OLÉ-controller variant of the 015-H platform, distinguished from the Pilot ONE siblings by its OLED/OLÉ control interface and USB/RS232 connectivity.

Noise level
63 dB

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBControllerSafety classTemperature range, °CInterfacePower supply
Unichiller P015-H OLÉ Current 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 015 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller 015-H OLÉ 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015w-H OLÉ 54OLÉI / NFL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 017T 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 017T-H 64Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 017Tw-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 020T —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020T-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 020Tw 60Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020Tw-H 60Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 022-H 65Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022w-H —Pilot ONEIII / FL-10–100——
Unichiller 022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025T 74Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025T-H 74Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025Tw-H —Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P007 58Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Unichiller P007-H —Pilot ONEIII / FL-20–100——
Unichiller P007-H OLÉ 58OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller P007 OLÉ 58OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller P010 57Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P010-H 57Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010-H OLÉ 57OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P010 OLÉ 57OLÉI / NFL-20–40USB (Device), RS232 Inte110-120V 1~ 50/60Hz
Unichiller P010w —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012 59Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P012-H 59Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012-H OLÉ 59OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012 OLÉ 59OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012w-H OLÉ —OLÉIII / FL-20–100USB (Device), RS232 Inte208V 2~ 60Hz
Unichiller P012w OLÉ —OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P015 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller P015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P015 OLÉ 63OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P015w 54Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller P015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P015w-H OLÉ 54OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50Hz
Unichiller P022 65Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022-H 65Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P022w-H 60Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P025 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz

Full specifications

Core Performance & Functional Specifications
Noise level 63 dB(A) (sound pressure level)
Pump connection G3/4 AG male
Pump speed adjustable no
Pump type B
Temperature range -20...100 °C
Dimensions, Weight & Mechanical Design
Weight 65 kg (net)
Electrical, Power & Drive Specifications
Current consumption 15 A (max)
Fuse rating 16 A
Power supply 220-240V 1~ 50Hz
Interfaces, Connections & Control
Controller OLÉ
Display resolution 0, 1 K
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C (permissible)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behavior under load

Installation, control and operation

Integration & Operating Notes

Process Limits
Low-temperature duty needs explicit heat-load sizing

Cooling capacity falls to 0.3 kW at -20 °C versus 1.5 kW at 15 °C, so a process sized at mid-range temperatures will be undersized at cold set points; verify the actual heat load at the intended set point before selection.

Unichiller P015-H OLÉ Datasheet, page 1
Process Limits
External circuit must be closed and pressure drop checked

The unit is designed only for externally closed applications, and the pump delivers 21 l/min at 0.5 bar but only 6 l/min at 2.0 bar; the process loop's pressure drop must be matched to this curve to avoid insufficient flow.

Unichiller Operation Manual, page 34; Datasheet page 1
Electrical
Country-specific plug must be specified at order

Single-phase devices are supplied with a power cable and a country-specific plug; the plug type must be stated when ordering, otherwise commissioning may be delayed by a missing or wrong connector.

Unichiller P015-H OLÉ Datasheet, page 2
Mounting
Air-cooled unit needs clearance and indoor siting

The chiller is air-cooled and requires sufficient space for airflow and service access; the manual specifies a wall clearance of 50 to 200 cm per side for maintenance, indoor use only, illuminance at least 300 lx and site altitude up to 2000 m.

Unichiller Operation Manual, page 25; Datasheet page 2
Mounting
Leveling feet and parking brakes must be set before start-up

Parking brakes at the casters must be activated and/or leveling feet unscrewed/activated before the unit is put into operation, to prevent the unit from moving and causing crushing injuries during operation.

Unichiller Operation Manual, page 31
Process Limits
Not permitted in ATEX zones

The unit is not explosion-protected and must not be installed or started up inside an ATEX zone; site hazardous-area classification must be checked before installation.

Unichiller Operation Manual, page 13
Control & I/O
Configure power-failure response to match process risk

The power-failure response can be set to OFF/Standby, ON or Power Failure Auto via the controller; for exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles.

Unichiller Operation Manual, pages 18–19
Maintenance
Refrigerant circuit work restricted to approved contractors

Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors; the refrigerant is R-449A (A1, H280), so leak and ventilation precautions apply and refrigerant charge versus room size must be checked at installation.

Unichiller Operation Manual, pages 15, 17; Datasheet page 1
Maintenance
Drain valve is an option, not base delivery

A drain valve (item #6839) is optional, so draining for service or fluid change requires ordering it separately; plan fluid replacement procedures accordingly.

Unichiller P015-H OLÉ Datasheet, page 2

Selection & Integration Q&A

When should I choose the P015-H OLÉ rather than the Unichiller 015-H or 015 OLÉ siblings?

Choose the P015-H OLÉ when the process loop has high pressure drop (the 'P' designation) and the OLÉ controller with USB/RS232 is required. The 015-H uses the Pilot ONE controller; the 015 OLÉ lacks the high-pressure-drop pump configuration.

What must I verify about the process loop and power supply before integrating this chiller?

Confirm the loop is externally closed and its pressure drop matches the pump curve (21 l/min at 0.5 bar down to 6 l/min at 2.0 bar), and provide 220–240 V 1~ 50 Hz with a 16 A fuse, respecting ±5% voltage / ±2% frequency tolerances not combined in opposite directions.

What site and duty-cycle limits apply to installing this air-cooled chiller?

Install indoors only, 5–40 °C ambient, illuminance ≥300 lx, altitude ≤2000 m, with 50–200 cm wall clearance per side for airflow and service. Activate leveling feet/parking brakes before start-up; the unit is not ATEX-rated.

What is not included in base delivery and what maintenance/accessory items should I plan for?

The drain valve (#6839) is optional and must be ordered separately; the country-specific plug must be specified at order. Refrigerant-circuit work requires approved contractors, and the R-449A charge versus room size must be checked at installation.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller P015-H OLÉ Datasheet
Datasheet ·87,7 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF