The chiller is rated for an operating temperature range of -20 to 100 °C, so the process must stay within these limits.
Defines the application window and excludes processes outside this range.
Source
“Operating temperature range -20...100 °C”
| Parameter | Value |
|---|---|
| Overall dimensions WxDxH ** | 420x487x579 mm |
| Net weight | 62 kg |
| sound pressure level +/- 4 dB(A) | 63 dB(A) |
| Power supply requirement | 220-240V 1~ 50Hz |
| max. current | 15,5 A |
| Fuse | 16 A |
| Pressure equipment category | Art. 4.3 PED |
| Degree of Protection | IP20 |
| min. ambient temperature | 5 °C |
| max. ambient temperature | 40 °C |
| from Serial-No.: | S612715 |
** Please respect space requirements. See operating conditions at www.huber-online.com
| Parameter | Value |
|---|---|
| Operating temperature range | -20...100 °C |
| temperature set point / display | 5,7" colour Touchscreen |
| Internal temperature sensor | Pt100 |
| Sensor external connection | Pt100 |
| Temperature stability at -10°C | 0,2 K |
| Interface digital | Ethernet, USB (Host u. Device), RS232 |
| Safety classification | III / FL |
| Heating power at 240V | 2,1 kW |
| Heating power at 230V | 2 kW |
| Heating power at 220V | 1,8 kW |
| Cooling power at ambient temperature +20°C | |
| at 15°C | 1,5 kW |
| at 0°C | 1 kW |
| at -10°C | 0,7 kW |
| at -20°C | 0,3 kW |
| Refrigeration machine | air-cooled, natural refrigerant |
| Refrigerant (ASHRAE, GHS) | R-290 (A3, H220) |
| Global Warming Potential (GWP) | 0,02 |
| Refrigerant quantity | 0,17 kg |
| CO2 equivalent | 0,0 t |
| UN-number | UN 3358 |
| Circulation pump: | B |
| max. delivery | 25 l/min |
| max. delivery pressure | 2,5 bar |
| Delivery at 0,5 bar | 21 l/min |
| Delivery at 1,0 bar | 17 l/min |
| Delivery at 1,5 bar | 11 l/min |
| Delivery at 2,0 bar | 6 l/min |
| Pump connection | G3/4 male |
| min. filling capacity | 3,8 l |
| Volume of expansion | 1,7 l |
Строка «Cooling power at ambient temperature +20°C» напечатана с точкой-заполнителем без значения; в таблице оставлена пустой.
The chiller is rated for an operating temperature range of -20 to 100 °C, so the process must stay within these limits.
Defines the application window and excludes processes outside this range.
“Operating temperature range -20...100 °C”
Cooling capacity drops sharply at low temperatures, from 1.5 kW at 15 °C to 0.3 kW at -20 °C, so low-temperature duty needs careful sizing.
Prevents undersizing the unit for cold setpoints.
“at 15°C 1,5 kW at 0°C 1 kW at -10°C 0,7 kW at -20°C 0,3 kW”
The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects room ventilation and ignition-source control.
Flammable refrigerant classification drives installation safety requirements.
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
The unit requires a 220-240 V single-phase 50 Hz supply with a maximum current of 15.5 A and a 16 A fuse.
Determines electrical supply, cabling and protection sizing.
“Power supply requirement 220-240V 1~ 50Hz max. current 15,5 A Fuse 16 A”
Voltage and frequency tolerances are coupled: ±5% voltage is allowed only with simultaneous ±2% frequency tolerance, and some combinations are not allowed.
Avoids invalid supply conditions that could damage the unit.
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
The unit is IP20 and must be installed in an environment between 5 °C and 40 °C ambient temperature.
Defines the permissible installation environment and protection against ingress.
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Cooling capacity is valid at 20 °C room temperature and may drop if ambient temperature rises.
Requires derating when installed in warm rooms.
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
The pump connection is G3/4 male, so hoses and fittings must match this thread.
Ensures correct plumbing connection at installation.
“Pump connection G3/4 male”
The circulation pump delivers 25 l/min maximum and 2.5 bar maximum pressure, with flow falling as pressure rises.
Needed to check whether the pump can overcome the process pressure drop.
“max. delivery 25 l/min max. delivery pressure 2,5 bar”
The minimum filling capacity is 3.8 l and expansion volume is 1.7 l, which sets the heat transfer fluid charge.
Determines fluid volume and expansion vessel sizing.
“min. filling capacity 3,8 l Volume of expansion 1,7 l”
The equipment is classified as EN55011 Class A, Group 1, which is industrial EMC classification.
Class A equipment may not be suitable for residential environments.
“Classification (disturbance) to EN55011: Class A, Group 1”
Single/two-phase devices are supplied with a power cable and country-specific plug that must be specified at ordering.
Prevents delivery with the wrong plug for the installation country.
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
The unit has adjustable overtemperature protection according to DIN 12876, which must be set for the application.
Overtemperature protection is a key safety function for the process fluid and equipment.
“With adjustable overtemperature protection according to DIN 12876.”