Unichiller P022w Datasheet

Contents

  1. Overview p. 1
    Product introduction and key features of the Unichiller P022w
  2. Spec pp. 1-2
    Technical data and performance specifications according to DIN 12876
  3. p. 2 table, 11 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Volume of expansion1,7 l
Overall dimensions WxDxH **420x487x579 mm
Net weight67 kg
sound pressure level +/- 4 dB(A)60 dB(A)
Power supply requirement208-230V 1~/2~ 60Hz
max. current12,5 A
Fuse2x16 A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.1/24

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterUnichiller P022w
Order-No.3010.0137.01
Operating temperature range-10...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at ambient temperature +20°C at 15°C2,2 kW
Cooling power at ambient temperature +20°C at 0°C1,6 kW
Cooling power at ambient temperature +20°C at -10°C1 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,45 kg
CO2 equivalent0,6 t
Circulation pumpB
max. delivery25 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar21 l/min
Delivery at 1,0 bar17 l/min
Delivery at 1,5 bar11 l/min
Delivery at 2,0 bar6 l/min
Pump connectionG3/4 male
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 15°C95 l/h
Consumption at water 15°C, flow 0°C96 l/h
Consumption at water 15°C, flow -10°C77 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity3,8 l

Engineering Facts

14 facts · every claim linked to a page
Installation

The chiller requires cooling water with a minimum differential pressure of 1 bar and must not exceed 6 bar, so the site water supply must be verified before installation.

Insufficient or excessive cooling water pressure will prevent correct operation or damage the unit.

Source
“min. cooling water differential pressure 1 bar”
Installation

The unit is designed for cooling water intake between 5°C and 20°C, and pipes must be insulated if the water temperature can fall below the condensation point.

Operating outside this range reduces cooling power and risks condensation, so the site cooling water temperature must be checked.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power is rated at 2.2 kW at 15°C, 1.6 kW at 0°C and 1 kW at -10°C, so capacity must be selected against the actual process temperature.

Cooling capacity drops significantly at lower temperatures, affecting process sizing.

Source
“Cooling power at ambient temperature +20°C”
Installation

The power supply must be 208-230V single or two-phase at 60 Hz with a maximum current of 12.5 A and 2x16 A fusing.

Electrical supply and protection must be provisioned accordingly before commissioning.

Source
“208-230V 1~/2~ 60Hz”
Compliance

Voltage and frequency tolerances are ±5% and ±2% respectively, but they may not be combined in opposite directions.

Supply quality must be verified to avoid invalid operating conditions and equipment stress.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant pump ordered.

Using incompatible media will damage the pump and void reliable operation.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Prevents freezing and damage in the cooling circuit at low temperatures.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

The refrigerant is R-449A with GWP 1397 and a charge of 0.45 kg, giving 0.6 t CO2 equivalent, which matters for leak reporting and environmental compliance.

Refrigerant type and charge determine environmental obligations and service requirements.

Source
“Refrigerant quantity 0,45 kg”
Installation

The unit has IP20 protection and must be installed in an ambient temperature between 5°C and 40°C.

IP20 offers no water protection, and ambient limits constrain the installation location.

Source
“min. ambient temperature 5 °C”
Installation

The pump connections are G3/4 male and cooling water connections are G1/2 male, so matching hose couplings must be prepared.

Correct fittings are needed for leak-free hydraulic connection.

Source
“Pump connection G3/4 male”
Performance

The circulation pump delivers 25 l/min maximum and 2.5 bar maximum pressure, with flow falling to 6 l/min at 2.0 bar.

Pump curve must be checked against system pressure drop to ensure adequate flow.

Source
“max. delivery pressure 2,5 bar”
Installation

The minimum filling capacity is 3.8 l and expansion volume is 1.7 l, which sets the heat transfer fluid charge and expansion allowance.

Determines fluid volume needed and expansion vessel sizing.

Source
“min. filling capacity 3,8 l”
Compliance

The equipment is classified Class A, Group 1 for electromagnetic disturbance to EN55011.

Class A equipment may require EMC measures in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Three-phase devices below 63 A are supplied with cable but without a plug, so a suitable plug must be provided on site.

Affects electrical installation planning and commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”