Installation
The chiller requires a three-phase 400 V supply with a 3x10 A fuse, so the electrical installation must be sized accordingly.
Determines the site power supply, cable and protection device rating before installation.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
The unit is water-cooled and needs a cooling water supply with at least 0.4 bar differential pressure and no more than 6 bar.
Cooling water pressure must be verified on site to guarantee heat rejection and avoid overpressure damage.
Source
“min. cooling water differential pressure 0,4 bar”
Installation
Cooling water intake temperature must stay between 5 °C and 20 °C, and pipes must be insulated if it can drop below the condensation point.
Out-of-range cooling water temperature degrades cooling power and can cause condensation and corrosion.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops significantly as the setpoint falls, from 5 kW at +20 °C to only 0.8 kW at -20 °C.
The required heat load at the lowest operating temperature must be checked against the reduced cooling capacity.
Source
“Cooling power at ambient temperature +20°C”
Selection
The circulation pump delivers 130 l/min maximum but only 20 l/min at 5.5 bar, so system pressure drop must be matched to the pump curve.
High-pressure-drop applications need the pump curve checked to confirm adequate flow at the actual system resistance.
Source
“max. delivery 130 l/min”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280, and the charge is 0.85 kg giving 1.2 t CO2 equivalent.
Refrigerant safety class, charge and CO2 equivalent affect room ventilation, leak detection and F-gas compliance obligations.
Source
“R-449A (A1, H280)”
Installation
The unit is IP20 and must be installed where ambient temperature stays between 5 °C and 40 °C.
IP20 offers no water protection and ambient limits define the permissible installation location.
Source
“min. ambient temperature 5 °C”
Installation
Three-phase devices below 63 A are supplied with a cable but without a plug, so a suitable plug must be provided on site.
The electrical contractor must supply and fit the mating plug for commissioning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Compliance
Voltage tolerance is ±5 % only with a simultaneous frequency tolerance of ±2 %, and certain combinations are not allowed.
Supply quality at the site must be checked to avoid invalid voltage/frequency combinations that could damage the unit.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
The unit needs at least 18 l filling capacity and has a 48 l expansion volume, so the external circuit volume must be planned.
Heat transfer fluid volume and expansion tank sizing depend on these values during system design.
Source
“min. filling capacity 18 l”
Operation
The unit is intended for externally closed applications only.
Open-loop or atmospheric external circuits are not permitted and would invalidate the intended use.
Source
“For externally closed applications.”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Water quality must be controlled to prevent corrosion and fouling of the mixed-material cooling water circuit.
Source
“Please use suitable cooling water.”
Installation
The unit measures 740x1160x1050 mm and weighs 300 kg, requiring adequate floor space and load capacity.
Footprint, access routes and floor loading must be confirmed before delivery and positioning.
Source
“Overall dimensions WxDxH ** 740x1160x1050 mm”
Compliance
The equipment emits Class A, Group 1 disturbances to EN55011, which restricts use in residential environments.
Class A equipment may cause interference in domestic settings and must be installed in industrial environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”