Peter Huber Kältemaschinenbau Unichiller P050w
Peter Huber Kältemaschinenbau Unichiller P050w — Recirculating chillers with Noise level 62 dB.
The Unichiller P050w is a water-cooled recirculating chiller with a Pilot ONE controller, covering -20 to 40 °C and requiring a 400 V 3~ 50 Hz supply. Its 'P' designation marks it for high-pressure-drop external circuits, distinguishing it from T-series siblings. It serves closed-loop process cooling where heat loads fall as setpoint drops.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Pump connection | Pump type | Safety class | Temperature range, °C | Power supply |
|---|---|---|---|---|---|---|
| Unichiller P050w Current | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 030T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 460V 3~ 60Hz |
| Unichiller 030Tw | 61 | G3/4 AG male | B | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 040T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040T-H | 69 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 040Tw | 70 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040Tw-H | 70 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 050-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 050w-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 055T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 055T-H | 74 | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 055Tw | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 060Tw-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 070T-H | 66 | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 070Tw-H | — | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 075w | — | G1 1/4 AG male | E1 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 075w-H | — | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100-H | 58 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 100T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 100T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100Tw | 65 | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller GL 035w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 050w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075 | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 100w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050 | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w-H | 62 | G1 1/4 AG male | E2 | III / FL | -20–100 | — |
| Unichiller P075 | 61 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100 | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P100-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100w | 58 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P100w-H | 58 | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
Full specifications
| Noise level | 62 dB(A) |
| Expansion volume | 48 l |
| Minimum fill volume | 18 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | E2 |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Weight | 300 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 137 л/час (at 15°C, flow -20°C) at 15°C, flow -10°C: 167 л/час · at 15°C, flow 0°C: 216 l/h |
| Cooling Water Pressure | 6 bar |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
The P050w is a water-cooled, free-standing chiller in the Unichiller series. Unlike the T-series siblings (030T, 030Tw, 030Tw-H, 040T), the 'P' in its model name designates it for applications requiring high pressure drops, per the operation manual. The unit is IP20, weighs 300 kg, and measures 740x1160x1050 mm. It uses refrigerant R-449A (A1, H280) with a 0.85 kg charge (1.2 t CO2 equivalent). The pump is a fixed-speed E2 type with G1 1/4 AG male connections; cooling water connections are G1/2 male.
Thermal and hydraulic behavior
Cooling capacity falls steeply with setpoint: 5 kW at +20 °C, dropping to 0.8 kW at -20 °C (at +20 °C ambient). The pump delivers up to 130 l/min, but only 20 l/min at 5.5 bar, so system pressure drop must be matched to the pump curve. Cooling water consumption depends on flow temperature: 216 l/h at 0 °C, 167 l/h at -10 °C, and 137 l/h at -20 °C, all at 15 °C intake. The unit requires a minimum fill volume of 18 l and provides 48 l expansion volume; it is intended for externally closed applications only.
Installation, control and operation
The Pilot ONE controller offers Ethernet, USB (host and device), and RS232 interfaces, plus a Pt100 external sensor connection. Power-failure response is configurable (OFF/Standby, ON, or Power Failure Auto). Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached, rather than switching off. The unit must be installed indoors at 5–40 °C ambient, with 50–200 cm wall clearance per side for maintenance, and leveling feet or parking brakes engaged before operation. Electrical lines below the unit must be protected from condensate.
Integration & Operating Notes
The chiller is delivered with a cable but without a plug for three-phase devices below 63 A. The electrical contractor must provide a suitable mating plug and protection rated 3x10 A. Voltage tolerance is ±5 % only with simultaneous frequency tolerance of ±2 %; certain voltage/frequency combinations are not permitted and could damage the unit.
Intake temperature outside 5–20 °C degrades cooling power and risks condensation; pipes must be insulated if the water can drop below the dew point. Differential pressure must be at least 0.4 bar and must not exceed 6 bar. The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used.
The unit is intended for externally closed applications only; open-loop or atmospheric circuits are not permitted. The external circuit volume must be planned with a minimum filling capacity of 18 l and the 48 l expansion volume in mind. The 'P' designation means the unit suits high-pressure-drop loops, but the pump curve (130 l/min max, 20 l/min at 5.5 bar) must still be checked against actual system resistance.
Selection & Integration Q&A
Choose the P050w when your external circuit has high pressure drop, since the 'P' designation marks it for such applications. The T-series siblings are not specified for high pressure drop. Also compare cooling capacity at your setpoint: the P050w delivers 5 kW at +20 °C but only 0.8 kW at -20 °C.
Cooling water: G1/2 male connections, 5–20 °C intake, 0.4–6 bar differential pressure, and consumption of 216 l/h at 0 °C flow (137 l/h at -20 °C). Process loop: G1 1/4 AG male pump connections, externally closed circuit, minimum fill 18 l, 48 l expansion volume. Only cooling water compatible with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.
Install indoors only, at 5–40 °C ambient, up to 2000 m altitude, with illuminance at least 300 lx. Not permitted in ATEX zones. Allow 50–200 cm wall clearance per side for maintenance. Engage leveling feet or parking brakes before operation. Protect electrical lines below the unit from condensate.
Refrigerant circuit work must be done by approved refrigeration contractors. Shut off cooling water even for overnight downtimes. The base delivery includes a cable but no plug for the 400 V 3~ supply; a suitable plug must be provided on site. No other accessories or consumables are listed in the provided documents.