Introduces the Unichiller P180w-H chiller and its main features
Specpp. 1-2
Performance, temperature, cooling capacity and dimensional data per DIN 12876
p. 2table, 8 rows
p. 1table, 31 rows
Technical data according to DIN 12876
Parameter
Value
Volume of expansion
48 l
Overall dimensions WxDxH **
940x1290x1130 mm
Power supply (3 Phase)
460V 3~ 60Hz
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/21
** Please respect space requirements. See operating conditions at www.huber-online.com
Unichiller P180w-H — Technical data according to DIN 12876
Parameter
Value
Order-No.
3041.0008.01
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power
2 kW
Cooling power at ambient temperature +20°C at 100°C
18 kW
Cooling power at ambient temperature +20°C at 20°C
18 kW
Cooling power at ambient temperature +20°C at 0°C
10 kW
Cooling power at ambient temperature +20°C at -10°C
6 kW
Cooling power at ambient temperature +20°C at -20°C
3,5 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
1,49 kg
CO2 equivalent
2,1 t
Circulation pump
E2
max. delivery
140 l/min
max. delivery pressure
5,3 bar
Delivery at 2,8 bar
126 l/min
Delivery at 3,0 bar
124 l/min
Delivery at 3,5 bar
115 l/min
Delivery at 4,0 bar
103 l/min
Delivery at 4,5 bar
83 l/min
Delivery at 5,0 bar
43 l/min
Pump connection
G1 1/4 male
Cooling water connection
G3/4 male
min. filling capacity
18 l
Cooling power values are given at ambient temperature +20°C for the listed temperatures (100°C, 20°C, 0°C, -10°C, -20°C).
Engineering Facts
14 facts · every claim linked to a page
Installation
The chiller is water-cooled and requires cooling water inlet temperature between 5°C and 20°C, so the site must provide cooling water within this range.
Cooling water outside this range will reduce cooling capacity or require additional measures, affecting utility design.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
If cooling water temperature falls below the condensation point, the cooling water pipes in the unit must be insulated.
Prevents condensation and corrosion/electrical risk in the cooling water circuit.
Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance
Higher cooling water temperature reduces cooling power and may require increased cooling water flow rate.
Cooling capacity must be derated based on actual site cooling water temperature and flow availability.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water may be used.
Water quality must be compatible with these materials to avoid corrosion and fouling.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Compliance
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Site power quality must meet these limits; otherwise the unit may not operate reliably or warranty may be affected.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Compliance
External branch circuit protection according to UL 489 is required.
Electrical installation must include UL 489 branch protection; affects panel design and compliance.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Three-phase devices with current consumption less than 63A are supplied with cable but without plug.
The installer must provide a suitable plug or hardwired connection; affects electrical installation planning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Three-phase devices with current consumption greater than 63A are supplied without cable and without plug.
Requires field wiring and connection by the installer; affects electrical installation scope.
Source
“Three-phase devices with current consumption greater than 63A --> without cable, without plug”
Operation
The unit is rated for ambient temperature from 5°C to 40°C.
Installation location must stay within this ambient range for proper operation.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Safety
The unit has adjustable overtemperature protection according to DIN 12876.
Safety function must be configured and tested during commissioning.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Selection
The chiller is intended for externally closed applications.
Application type must match; open systems are not supported.
Source
“For externally closed applications.”
Selection
The refrigerant is R-449A with GWP 1397 and CO2 equivalent 2.1 t.
Refrigerant type and charge affect environmental compliance and service requirements.
Source
“Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)”
Installation
The unit requires a 460V 3~ 60Hz power supply.
Site power must match this supply; otherwise a transformer or different configuration is needed.
Source
“Power supply (3 Phase)
460V 3~ 60Hz”
Compliance
The equipment is classified as Class A, Group 1 for electromagnetic disturbance to EN55011.
Class A equipment may require EMC measures in residential or light-industrial environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”