Introduces the Unistat 615w unit and its main features
Specpp. 1-2
Performance, temperature range, and technical data per DIN 12876
p. 2table, 18 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 20°C
580 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
6,5 l
Filling capacity expansion tank
26 l
Overall dimensions WxDxH **
630x704x1520 mm
Net weight
367 kg
sound pressure level +/- 4 dB(A)
62 dB(A)
Power supply requirement
400V 3~ 50Hz
max. current
25 A
Fuse
3x32A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S471936
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Parameter
Unistat 615w
Order-No.
1074.0001.01
Operating temperature range
-60...200 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
12 kW
Cooling power with Thermooil at 200°C
9,5 kW
Cooling power with Thermooil at 100°C
9,5 kW
Cooling power with Thermooil at 20°C
9,5 kW
Cooling power with Ethanol at 0°C
9,5 kW
Cooling power with Ethanol at -20°C
8 kW
Cooling power with Ethanol at -40°C
4,6 kW
Cooling power with Ethanol at -60°C
1,2 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
2,55 kg
CO2 equivalent
5,5 t
Circulation pump: max. delivery
82 l/min
Circulation pump: max. delivery pressure
1,5 bar
Delivery at 0,25 bar
75 l/min
Delivery at 0,5 bar
67 l/min
Delivery at 0,75 bar
59 l/min
Delivery at 1,0 bar
47 l/min
Delivery at 1,25 bar
29 l/min
Pump connection
M30x1,5 male
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a 400 V three-phase supply with a 25 A maximum current and 3x32 A fusing, so the electrical installation must be sized accordingly.
Determines the site power distribution, cable cross-section and breaker rating needed before installation.
Source
“Power supply requirement 400V 3~ 50Hz max. current 25 A Fuse 3x32A”
Installation
Cooling water must be supplied at 5–20 °C with at least 1 bar differential pressure and no more than 6 bar, and the circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.
Defines the facility cooling-water supply conditions and water quality compatibility required for reliable operation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
The cooling water connection is G1/2 male and the unit consumes about 580 l/h at 15 °C water and 20 °C flow, so the water supply and return lines must handle this flow.
Needed to size cooling-water piping, valves and facility supply capacity.
Source
“Cooling water connection G1/2 male Consumption at water 15°C, flow 20°C 580 l/h”
Installation
The unit is water-cooled and uses R-452A refrigerant with a GWP of 2141 and 2.55 kg charge (5.5 t CO2 equivalent), which affects refrigerant handling and environmental compliance.
Important for refrigerant regulations, leak detection and environmental permitting at the installation site.
Source
“Refrigerant (ASHRAE, GHS) R-452A (A1, H280) Global Warming Potential (GWP) 2141 Refrigerant quantity 2,55 kg CO2 equivalent 5,5 t”
Installation
The unit weighs 367 kg net and measures 630x704x1520 mm, so floor loading and access routes must be checked before moving it into place.
Determines rigging, floor loading and doorway/access clearance requirements.
Source
“Overall dimensions WxDxH ** 630x704x1520 mm Net weight 367 kg”
Operation
The operating temperature range is -60 to 200 °C with 0.01 K display resolution and Pt100 sensors, suitable for precise external closed and open applications.
Confirms the unit meets the required process temperature span and stability for the application.
Source
“Operating temperature range -60...200 °C Temperature stability at -10°C 0,01 K”
Performance
Cooling power is 9.5 kW with Thermooil from 20 to 200 °C but drops to 1.2 kW with ethanol at -60 °C, so low-temperature duty must be checked against the load.
Prevents oversizing expectations at the cold end and ensures the process heat load is removable at -60 °C.
Source
“Cooling power with Ethanol at -60°C 1,2 kW”
Performance
The circulation pump delivers up to 82 l/min and 1.5 bar, with 47 l/min at 1.0 bar, so external circuit pressure drop must be matched to the pump curve.
Needed to verify the pump can overcome the pressure drop of the external application loop.
Source
“max. delivery 82 l/min max. delivery pressure 1,5 bar Delivery at 1,0 bar 47 l/min”
Operation
The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, which limits fluid selection at low temperatures.
Ensures the chosen heat transfer fluid remains pumpable within the unit's limits across the operating range.
Source
“max. permissible kin. viscosity 50 mm²/s”
Installation
The minimum filling capacity is 6.5 l and the expansion tank holds 26 l, so the external loop volume plus unit volume must be considered when charging fluid.
Determines the minimum fluid charge and expansion volume available for the external circuit.
Source
“min. filling capacity 6,5 l Filling capacity expansion tank 26 l”
Installation
The unit is rated IP20 and must be operated between 5 and 40 °C ambient, so it must be installed in a dry, temperature-controlled indoor location.
Sets the environmental envelope and ingress protection limits for the installation room.
Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Compliance
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and a -5% voltage with +2% frequency is explicitly not allowed.
Ensures the site power quality is within the unit's permitted supply envelope before commissioning.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Three-phase devices below 63 A are supplied with cable but without a plug, so a suitable plug and connection must be provided on site.
Affects the electrical installation scope and the need to procure a matching plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety
The unit has safety classification III/FL and adjustable overtemperature protection according to DIN 12876, relevant for flammable or sensitive fluids.
Confirms the safety category for use with flammable heat transfer fluids and the required protective settings.