Peter Huber Kältemaschinenbau Unistat 615w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat 615w Dynamisches

Peter Huber Kältemaschinenbau Unistat 615w Dynamisches — Temperature control units with Noise level 62 dB.

The Unistat 615w Dynamisches is a water-cooled, free-standing temperature control unit with a -60 to 200 °C range, 12 kW heating and a variable-speed magnetically coupled pressure pump. It is the 6xx-series water-cooled modification, distinguished from air-cooled siblings by its cooling-water circuit and higher capacity. It serves precise external closed and open temperature control loops in process engineering.

Noise level
62 dB

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat 615w Dynamisches Current 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Noise level 62 dB(A) (sound pressure level)
Control output POKO ONE (potential-free contact)
Expansion volume 26 l
Heating capacity 12 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type Pressure pump, variable speed, magnetically coupled
External sensor connection Pt100
Temperature range -60...200 °C
Dimensions, Weight & Mechanical Design
Weight 367 kg (net)
Electrical, Power & Drive Specifications
Current consumption 25 A (max)
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s (max permissible)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C (permissible)
Cooling water connection G1/2 male
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, connection and control

Integration & Operating Notes

Electrical
Site power must be sized for 25 A and 3x32 A fusing

The 400 V three-phase supply must be dedicated and protected at 3x32 A; the unit is delivered with cable but no plug, so the customer must provide a matching plug and breaker. Voltage tolerance is only ±5% with ±2% frequency, and a -5% voltage with +2% frequency is not permitted.

Unistat 615w Technical Data, page 2
Cooling & Media
Cooling water must be 5–20 °C at 1–6 bar differential

The water-cooled condenser requires a facility supply of 5–20 °C with at least 1 bar differential pressure and no more than 6 bar; flow is about 580 l/h at 15 °C water and 20 °C flow. Circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so water quality must be compatible. Shut off cooling water even for overnight downtime.

Unistat 615w Technical Data, page 2; Unistat Operation Manual, page 29
Process Limits
Low-temperature cooling capacity collapses to 1.2 kW at -60 °C

Cooling power is 9.5 kW with Thermooil from 20 to 200 °C but only 1.2 kW with ethanol at -60 °C. Process heat loads at the cold end must be checked against this figure; the maximum permissible kinematic viscosity of 50 mm²/s further restricts fluid choice at low temperature.

Unistat 615w Technical Data, page 1
Mounting
Transport upright and empty; casters are not for transport

The unit must be transported and stored upright only; horizontal transport damages the compressor. Only an emptied unit may be moved. Casters, if present, carry only 25% of total mass symmetrically and are not suitable for transport; use an industrial truck. Leveling feet must be activated and parking brakes engaged before commissioning.

Unistat Operation Manual, page 22
Process Limits
Expansion vessel lock state depends on open vs closed application

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open (bath) applications it must be locked to prevent fluid loss. The wrong setting is a direct damage risk, not a preference.

Unistat Operation Manual, page 33
Control & I/O
Process Safety mode requires compressor set to Permanently ON

When Process Safety is enabled, the controller releases maximum cooling capacity at the overtemperature trigger instead of switching off all poles; this mode requires the compressor to be set to Permanently ON. Power Failure Auto can resume temperature control automatically after an outage if it was active, which must be assessed for unattended operation.

Unistat Operation Manual, page 19–20
Maintenance
Coolant-circuit work only by approved refrigeration contractors

Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors; general maintenance personnel may not service the refrigerant circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed. A condensation water / heat transfer fluid drip tray must be installed below the unit.

Unistat Operation Manual, Section 1.4.2, page 19; Section 1.4, page 17

Selection & Integration Q&A

When should I choose the Unistat 615w Dynamisches instead of the 510w, 515w or 520w water-cooled siblings?

Choose the 615w when the process needs the higher heating capacity (12 kW) and the 6xx platform's larger expansion volume (26 l) and pump delivery (82 l/min, 1.5 bar). The 510w/515w/520w are lower-capacity water-cooled units; the 525 Dynamisches is air-cooled and quieter (56 dB(A)) but lacks a cooling-water circuit.

What must be provided on site for the electrical and cooling-water connections?

A dedicated 400 V 3~ 50 Hz supply with 3x32 A fusing and a matching plug (cable supplied, plug not). Cooling water at 5–20 °C, 1–6 bar differential, G1/2 male connection, about 580 l/h at 15 °C water and 20 °C flow. Voltage tolerance ±5% with ±2% frequency; -5% voltage with +2% frequency is not allowed.

What are the installation environment and duty-cycle limits from the manual?

Indoors only, IP20, ambient 5–40 °C, not within an ATEX zone. Transport and store upright and empty; allow about 2 h acclimation after a cold-to-warm move before power-on. Casters are not for transport. Shut off cooling water even overnight. Medical-device and direct-food use are not permissible.

What maintenance and accessories are not included in the base delivery?

The plug for the three-phase cable, the condensation water / heat transfer fluid drip tray, and any gas detection sensor or processing unit are not part of the base unit and must be engineered by the operating company. Refrigerant-circuit service requires approved refrigeration contractors; hoses and heat transfer fluid are wear items to be inspected and replaced periodically.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat 615w Technical Data
Datasheet ·74,6 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF