Product introduction and general description of the Unistat 635w unit
Specpp. 1-2
Technical specifications and performance data per DIN 12876
p. 2table, 18 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
max. permissible kin. viscosity
100 mm²/s
Cooling water connection
G3/4 male
min. cooling water differential pressure
1 bar
max. cooling water differential pressure
4 bar
max. cooling water pressure
6 bar
max. System pressure
6 bar
min. filling capacity
8,9 l
Filling capacity expansion tank
19,9 l
Overall dimensions WxDxH **
918x963x1771 mm
Net weight
721 kg
Power supply requirement
400V 3~ 50Hz
max. current
56 A
Fuse
3x63A
Pressure equipment category
II
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S569436
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Characteristic
Value
Operating temperature range
-60...200 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Interface analog
optional
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
24 kW
Cooling power with Thermooil at 200°C
27 kW
Cooling power with Thermooil at 100°C
27 kW
Cooling power with Thermooil at 20°C
27 kW
Cooling power with Ethanol at 0°C
27 kW
Cooling power with Ethanol at -20°C
25 kW
Cooling power with Ethanol at -40°C
14 kW
Cooling power with Ethanol at -60°C
3,5 kW
Refrigeration machine
water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-1270 (A3, H220)
Global Warming Potential (GWP)
0
Refrigerant quantity
2,56 kg
CO2 equivalent
0,0 t
Circulation pump:
MK pump
max. delivery
196 l/min
max. delivery pressure
2,5 bar
Delivery at 0,5 bar
175 l/min
Delivery at 1,0 bar
152 l/min
Delivery at 1,5 bar
123 l/min
Delivery at 2,0 bar
86 l/min
Pump connection
M38x1,5 male
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a 400 V 3~ 50 Hz supply with a 3x63 A fuse and draws up to 56 A, so the electrical installation must be sized accordingly.
Determines feeder cable, breaker and upstream distribution capacity before installation.
Source
“Power supply requirement 400V 3~ 50Hz max. current 56 A Fuse 3x63A”
Installation
Cooling water must be supplied at 1–4 bar differential pressure and no more than 6 bar static pressure through G3/4 male connections.
Ensures the facility cooling water loop can meet flow and pressure limits without overpressurizing the unit.
Source
“min. cooling water differential pressure 1 bar max. cooling water differential pressure 4 bar max. cooling water pressure 6 bar”
Operation
Rated cooling capacity is only valid at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure, and the unit is designed for 5–20 °C cooling water intake.
Prevents overestimating cooling performance if site cooling water is warmer than the reference conditions.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -outlet. This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
If cooling water temperature can fall below the condensation point, the cooling water pipes inside the unit must be insulated.
Avoids condensation and corrosion risk when chilled water is used.
Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality and treatment to prevent corrosion or fouling of the cooling circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Compliance
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.
Defines acceptable site power quality and protects the compressor and electronics from out-of-tolerance supply.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2% Example -5% voltage and +2% frequency -> not allowed!”
Safety
Leakage current exceeds 3.5 mA, which affects protective earthing and RCD selection.
Requires proper earth connection and consideration of residual current device type and rating.
Source
“Attention: leakage current > 3,5mA”
Installation
Three-phase devices below 63 A are delivered with cable but without plug, so a suitable plug must be provided on site.
Prevents installation delay because no mating plug is included.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Selection
The unit uses R-1270 refrigerant (A3, H220) with GWP 0 and a charge of 2.56 kg, which has safety implications for the installation room.
A3 flammable refrigerant affects room classification, ventilation and safety planning.