Peter Huber Kältemaschinenbau Unistat 635w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat 635w Dynamisches

Peter Huber Kältemaschinenbau Unistat 635w Dynamisches — Temperature control units.

The Unistat 635w Dynamisches is a water-cooled, free-standing temperature control unit with a -60 to 200 °C range, 24 kW heating and a magnetically coupled stainless-steel pump delivering up to 196 l/min. It is the largest water-cooled modification in the Unistat 5xx/6xx series, distinguished by its 400 V 3~ 50 Hz supply, 56 A maximum current and 3x63 A fuse requirement. It serves as a high-capacity circulator for externally closed or open processes requiring precise temperature control with Pilot ONE regulation.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat 635w Dynamisches Current —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE (potential-free contact)
Heating capacity 24 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -60...200 °C
Dimensions, Weight & Mechanical Design
Weight 721 kg (net)
Electrical, Power & Drive Specifications
Current consumption 56 A (max)
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 100 mm²/s (max permissible)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C (permissible)
Cooling water connection G3/4 male
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal and hydraulic behavior

Control, installation and operation

Integration & Operating Notes

Process Limits
Heat transfer fluid viscosity must not exceed 100 mm²/s

The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s, which limits fluid selection at low temperatures and must be respected to avoid pump overload and poor circulation. Minimum filling capacity is 8.9 l and the expansion tank holds 19.9 l, setting the minimum fluid charge and expansion allowance.

Unistat 635w Technical Data, page 2
Maintenance
Shut off cooling water even for short downtimes such as overnight

Shut off the cooling water supply to the temperature control unit even for shorter downtimes (e.g. overnight) to prevent water damage and unnecessary load on the cooling water circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed.

Unistat Operation Manual, page 29

Selection & Integration Q&A

When should I choose the Unistat 635w Dynamisches instead of a smaller sibling like the 510w or 525 Dynamisches?

Choose the 635w when your process requires the higher heating capacity (24 kW) and pump flow (196 l/min max) of the largest water-cooled frame, or when cooling water is available and air-cooled siblings cannot meet the heat rejection. The 510w, 515w, 520w and 525 are smaller air-cooled or lower-output models; select them only if their capacity and cooling mode match your load.

What are the electrical supply and protection requirements for integrating this unit into my facility?

The unit requires a 400 V 3~ 50 Hz supply with a 3x63 A fuse and draws up to 56 A. It is delivered with cable but without a plug, so a suitable plug must be provided on site. Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and leakage current exceeds 3.5 mA, so RCD selection and earthing must be engineered accordingly.

What are the installation environment and duty-cycle limits from the manual?

Install in a dry indoor location with ambient temperature 5–40 °C and IP20 protection, outside ATEX zones. The unit must be transported upright and emptied; casters are not for transport. Allow about 2 hours acclimation after moving between cold and warm environments. A drip tray is required below the unit, and cooling water must be shut off even for overnight downtimes.

What maintenance schedule and accessories are required, and what is not included in base delivery?

Periodically check hoses and connections for leaks and quality and replace as needed; clean liquefier fins and the hat-type strainer/dirt trap; inspect heat transfer fluid and replace or rinse the circuit as required. The base delivery does not include a power plug, a drip tray, or cooling water hoses; these must be provided on site. Refrigerant-circuit work requires approved contractors.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat 635w Technical Data
Datasheet ·31,3 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF