Installation
The unit requires a three-phase 400 V supply with a 3x32 A fuse, so the electrical installation must be sized accordingly.
Determines the site power supply, cable and protection device rating before installation.
Source
“Power supply factory configured (3 Phase) 400V 3~ 50Hz”
Installation
Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so the customer must provide and fit the plug.
Affects electrical installation scope and procurement of the mating plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety
Leakage current exceeds 3.5 mA, which may require a suitable residual current device or protective earthing arrangement.
Important for electrical safety design and RCD selection during installation.
Source
“Attention: leakage current > 3,5mA”
Installation
The unit must be operated within an ambient temperature of 5 to 40 °C, so the installation location must be kept within this range.
Site conditions outside this range may prevent proper operation or reduce performance.
Source
“min. ambient temperature 5 °C”
Performance
Cooling capacity is rated at 20 °C room temperature and may drop if ambient temperature rises, so capacity must be derated for warmer rooms.
Prevents undersizing the unit when installed in a warm environment.
Source
“If the ambient temperature rises, the cooling capacity may drop.”
Installation
The pump connection is M30x1.5 male with bore shape Y (60°) per DIN 3863, so mating hoses or fittings must match this standard.
Ensures correct selection of hoses and adapters for the external circuit.
Source
“Pump connection M30x1,5 male”
Installation
The pump connection bore and mating ball socket/sleeve nut must follow DIN 3863 and DIN 3870, so non-compliant fittings should not be used.
Guarantees leak-free and mechanically compatible connections.
Source
“Bore shape Y (60°) according to DIN 3863”
Operation
The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, so fluid selection must respect this limit.
Using a too-viscous fluid can overload the pump and impair circulation.
Source
“max. permissible kin. viscosity 50 mm²/s”
Operation
The minimum filling capacity is 5.65 l and the expansion tank holds 26.5 l, so the system volume must be planned within these limits.
Ensures sufficient fluid volume and expansion space for safe operation.
Source
“min. filling capacity 5,65 l”
Selection
The unit uses refrigerant R-452A with GWP 2141 and a CO2 equivalent of 11.1 t, which matters for environmental compliance and leak reporting.
Affects refrigerant regulations, leak checks and environmental permitting.
Source
“R-452A (A1, H280)”
Selection
The unit is classified as pressure equipment category Art. 4.3 PED, so it falls under the Pressure Equipment Directive requirements.
Relevant for compliance documentation and acceptance at site.
Source
“Pressure equipment category Art. 4.3 PED”
Installation
The unit has IP20 protection, so it must be installed in a dry indoor location protected from water and dust ingress.
Prevents installation in wet or dusty environments that could damage the equipment.
Source
“Degree of Protection IP20”
Installation
The overall dimensions are 1290x955x1537 mm and net weight 576 kg, so access routes and floor loading must be checked before installation.
Needed for rigging, doorway clearance and structural floor assessment.
Source
“Overall dimensions WxDxH ** 1290x955x1537 mm”
Compliance
The equipment is EMC class A group 1, so it may cause radio interference in residential areas and is intended for industrial environments.
Determines suitability for industrial versus residential installation.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”