Peter Huber Kältemaschinenbau Unistat P615 Dynamisches

Peter Huber Kältemaschinenbau Unistat P615 Dynamisches — Temperature control units.

The Unistat P615 Dynamisches is a free-standing temperature control unit with a magnetically coupled stainless-steel pump, covering -60 to 200 °C with 12 kW heating capacity. Its 'P' designation targets applications with high pressure drop, distinguishing it from standard 5xx siblings. It serves as a closed-loop circulator for jacketed reactors and process equipment requiring dynamic temperature control.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat P615 Dynamisches Current —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 12 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -60...200 °C
Dimensions, Weight & Mechanical Design
Weight 576 kg
Electrical, Power & Drive Specifications
Current consumption 29 A
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Working range and media

Electrical and control

Installation and operation specifics

Integration & Operating Notes

Electrical
Site supply must be sized for 29 A and 3x32 A protection

The unit draws 29 A on a 400 V 3~ 50 Hz supply and is factory-configured for a 3x32 A fuse; the upstream breaker, cable and disconnect must be rated accordingly. Leakage current exceeds 3.5 mA, so RCD selection and protective earthing must be engineered for this value.

Unistat P615 Datasheet, Specifications
Electrical
Plug not included on three-phase delivery under 63 A

Three-phase devices with current consumption below 63 A are delivered with a cable but without a plug; the customer must provide and fit a mating plug. Hard-wired connection must be performed by an electrician and the supply voltage/frequency verified against the rating plate.

Unistat P615 Datasheet, Specifications; Unistat Operation Manual, Commissioning
Mounting
Transport upright and emptied; casters not for transport

Horizontal transport damages the compressor and a filled unit risks thermal fluid overflow, so the unit must be moved upright and drained. Casters carry only 25% of total mass symmetrically and are not suitable for transport; use an industrial truck. Leveling feet must be activated and parking brakes engaged before commissioning.

Unistat Operation Manual, Commissioning
Mounting
Exclude ATEX zones and allow 2 h acclimation before power-on

The unit must not be installed or started within an ATEX zone; explosion protection requires a separate Ex px cabinet. When moving between cold and warm environments, allow about 2 hours acclimation before switching on to prevent condensation-induced electrical faults.

Unistat Operation Manual, Section 1.3.3; Commissioning
Process Limits
Expansion vessel lock state depends on open vs closed application

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open (bath) applications the expansion vessel must be locked to prevent fluid loss. Verify the isolating sleeve/kit position before filling.

Unistat Operation Manual, Preparations for Operation
Process Limits
Fluid viscosity capped at 50 mm²/s and fill volume planned

Heat transfer fluid must not exceed 50 mm²/s kinematic viscosity or the pump may be overloaded. Minimum filling capacity is 5.65 l with a 26.5 l expansion tank; system volume and expansion space must be planned within these limits. For FL-rated fluids, fire point must be ≥65 °C.

Unistat P615 Datasheet, Specifications; Unistat Operation Manual, Section 1.7
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode the controller releases maximum cooling capacity at the OT trigger instead of switching off all poles; this requires the compressor to be set to Permanently ON. Configure and function-test OT limit heating and OT expansion vessel before running exothermic processes.

Unistat Operation Manual, Section 1.7.2; Setup Mode
Control & I/O
Choose power-failure restart behavior for unattended operation

Pilot ONE can be set to Off/Standby (default, manual restart) or ON/Temperature control active (automatic restart after power-on, with a few-second INFO window to abort). Power Failure Auto resumes control after an outage if it was active. Select according to whether unattended restart is acceptable for the process.

Unistat Operation Manual, Section 1.8.1
Maintenance
Refrigerant circuit work restricted to approved contractors

Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors; general maintenance personnel may not service the refrigerant circuit. The circuit is factory-tested to a leak sensitivity of 10-4 Pa m3/s or better.

Unistat Operation Manual, Section 1.4.1 and 1.4.2

Selection & Integration Q&A

When should I choose the Unistat P615 Dynamisches over the 510w/515w/520w/525 siblings?

Choose the P615 when the application has high pressure drop (the 'P' designation) and requires the -60 to 200 °C range with 12 kW heating. The 5xx siblings are the baseline air- or water-cooled variants; select them when pressure drop is low and the P-configuration is unnecessary.

What mating fittings and fluid properties must the external circuit satisfy?

Pump connection is M30x1.5 male, bore shape Y (60°) per DIN 3863, with mating ball socket/sleeve nut per DIN 3870. Heat transfer fluid must not exceed 50 mm²/s kinematic viscosity; minimum filling capacity 5.65 l with 26.5 l expansion tank.

What site and duty-cycle limits apply before commissioning?

Ambient must be 5 to 40 °C, dry indoor, IP20, non-ATEX. Transport upright and emptied; allow ~2 h acclimation after cold/warm moves. Activate leveling feet and parking brakes; cooling capacity is rated at 20 °C room temperature and derates in warmer rooms.

What maintenance and delivery-scope items are not included with the base unit?

Three-phase units under 63 A ship with cable but no plug; the customer supplies the mating plug. A condensation/fluid drip tray must be installed by the responsible body. Periodic hose inspection/replacement, cooling-water shutoff for overnight downtime, and refrigerant work by approved contractors are owner responsibilities.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat P615 Datasheet
Datasheet ·168,2 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF