The Unistat P615 Dynamisches is a free-standing temperature control unit with a magnetically coupled stainless-steel pump, covering -60 to 200 °C with 12 kW heating capacity. Its 'P' designation targets applications with high pressure drop, distinguishing it from standard 5xx siblings. It serves as a closed-loop circulator for jacketed reactors and process equipment requiring dynamic temperature control.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat P615 Dynamisches Current | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Control output | POKO ONE |
| Heating capacity | 12 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -60...200 °C |
| Weight | 576 kg |
| Current consumption | 29 A |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Free-standing 5xx-class Unistat with Pilot ONE controller and POKO ONE control output; overall dimensions 1290x955x1537 mm, net weight 576 kg.
- The 'P' suffix denotes configuration for applications with high pressure drop, per the manual's model-name convention; standard 5xx siblings (510w, 515w, 520w, 525) are the air- or water-cooled baseline variants.
- Pump is magnetically coupled, stainless steel, with adjustable speed; pump connection is M30x1.5 male, bore shape Y (60°) per DIN 3863, mating ball socket/sleeve nut per DIN 3870.
- Safety class III / FL; external sensor connection Pt100; display resolution 0.01 K.
Working range and media
- Temperature range -60 to 200 °C; heating capacity 12 kW. Cooling capacity is rated at 20 °C room temperature and may drop as ambient rises.
- Maximum permissible kinematic viscosity of heat transfer fluid is 50 mm²/s; minimum filling capacity 5.65 l, expansion tank 26.5 l.
- Refrigerant R-452A (GWP 2141, CO2 equivalent 11.1 t); pressure equipment category Art. 4.3 PED.
Electrical and control
- Power supply 400 V 3~ 50 Hz, current consumption 29 A, factory-configured 3x32 A fuse; three-phase devices under 63 A are delivered with cable but without plug.
- Leakage current exceeds 3.5 mA; EMC class A group 1 (industrial environments only); IP20.
- Interfaces: Ethernet, USB (host and device), RS232. Controller Pilot ONE with touchscreen, LED flow diagram, clock/event function.
Installation and operation specifics
- Ambient temperature 5 to 40 °C; unit must not be installed in an ATEX zone; transport upright only, emptied; casters not for transport (25% load rating); leveling feet and parking brakes before commissioning.
- Allow ~2 h acclimation when moving between cold and warm environments before power-on.
- Expansion vessel must remain unlocked for externally closed applications and locked for externally open (bath) applications.
- Overtemperature protection (OT limit heating, OT expansion vessel, Process Safety) must be set and function-tested; Process Safety mode releases maximum cooling at the OT trigger and requires compressor set to Permanently ON.
- Power-interruption behavior configurable: Off/Standby (default) or ON/Temperature control active; Power Failure Auto resumes control after outage if active.
Integration & Operating Notes
The unit draws 29 A on a 400 V 3~ 50 Hz supply and is factory-configured for a 3x32 A fuse; the upstream breaker, cable and disconnect must be rated accordingly. Leakage current exceeds 3.5 mA, so RCD selection and protective earthing must be engineered for this value.
Three-phase devices with current consumption below 63 A are delivered with a cable but without a plug; the customer must provide and fit a mating plug. Hard-wired connection must be performed by an electrician and the supply voltage/frequency verified against the rating plate.
Horizontal transport damages the compressor and a filled unit risks thermal fluid overflow, so the unit must be moved upright and drained. Casters carry only 25% of total mass symmetrically and are not suitable for transport; use an industrial truck. Leveling feet must be activated and parking brakes engaged before commissioning.
The unit must not be installed or started within an ATEX zone; explosion protection requires a separate Ex px cabinet. When moving between cold and warm environments, allow about 2 hours acclimation before switching on to prevent condensation-induced electrical faults.
For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open (bath) applications the expansion vessel must be locked to prevent fluid loss. Verify the isolating sleeve/kit position before filling.
Heat transfer fluid must not exceed 50 mm²/s kinematic viscosity or the pump may be overloaded. Minimum filling capacity is 5.65 l with a 26.5 l expansion tank; system volume and expansion space must be planned within these limits. For FL-rated fluids, fire point must be ≥65 °C.
In Process Safety mode the controller releases maximum cooling capacity at the OT trigger instead of switching off all poles; this requires the compressor to be set to Permanently ON. Configure and function-test OT limit heating and OT expansion vessel before running exothermic processes.
Pilot ONE can be set to Off/Standby (default, manual restart) or ON/Temperature control active (automatic restart after power-on, with a few-second INFO window to abort). Power Failure Auto resumes control after an outage if it was active. Select according to whether unattended restart is acceptable for the process.
Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors; general maintenance personnel may not service the refrigerant circuit. The circuit is factory-tested to a leak sensitivity of 10-4 Pa m3/s or better.
Selection & Integration Q&A
Choose the P615 when the application has high pressure drop (the 'P' designation) and requires the -60 to 200 °C range with 12 kW heating. The 5xx siblings are the baseline air- or water-cooled variants; select them when pressure drop is low and the P-configuration is unnecessary.
Pump connection is M30x1.5 male, bore shape Y (60°) per DIN 3863, with mating ball socket/sleeve nut per DIN 3870. Heat transfer fluid must not exceed 50 mm²/s kinematic viscosity; minimum filling capacity 5.65 l with 26.5 l expansion tank.
Ambient must be 5 to 40 °C, dry indoor, IP20, non-ATEX. Transport upright and emptied; allow ~2 h acclimation after cold/warm moves. Activate leveling feet and parking brakes; cooling capacity is rated at 20 °C room temperature and derates in warmer rooms.
Three-phase units under 63 A ship with cable but no plug; the customer supplies the mating plug. A condensation/fluid drip tray must be installed by the responsible body. Periodic hose inspection/replacement, cooling-water shutoff for overnight downtime, and refrigerant work by approved contractors are owner responsibilities.