Introduces the Unistat P825w unit and its main features
Specp. 2
Lists technical data according to DIN 12876 standard
p. 2table, 26 rows
p. 1table, 30 rows
Technical data according to DIN 12876
Parameter
Value
Circulation pump
MK pump
max. delivery
67 l/min
max. delivery pressure
3 bar
Delivery at 0,5 bar
61 l/min
Delivery at 1,0 bar
55 l/min
Delivery at 1,5 bar
48 l/min
Delivery at 2,0 bar
39 l/min
Delivery at 2,5 bar
28 l/min
Pump connection
M30x1,5 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 20°C
180 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
2,4 l
Filling capacity expansion tank
3,3 l
Overall dimensions WxDxH **
460x604x1465 mm
Net weight
237 kg
Power supply requirement
400V 3~ 50Hz
max. current
15 A
Fuse
3x16A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S513166
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-85...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
3 kW
Cooling power with Thermooil at 250°C
2,3 kW
Cooling power with Thermooil at 200°C
2,3 kW
Cooling power with Thermooil at 100°C
2,3 kW
Cooling power with Thermooil at 20°C
2,2 kW
Cooling power with Ethanol at 0°C
2,2 kW
Cooling power with Ethanol at -20°C
2,1 kW
Cooling power with Ethanol at -40°C
2 kW
Cooling power with Ethanol at -60°C
1,3 kW
Cooling power with Ethanol at -80°C
0,3 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
0,95 kg
CO2 equivalent
2,0 t
Refrigerant 2nd stage (ASHRAE, GHS)
R-23 (A1, H280)
Global Warming Potential (GWP)
14800
Refrigerant quantity 2nd stage
0,39 kg
CO2 equivalent 2nd stage
5,8 t
Order-No.: 1079.0012.01; Unistat P825w
Engineering Facts
12 facts · every claim linked to a page
Installation
The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.
Determines the site power supply and breaker/fuse rating needed before installation.
Source
“Power supply requirement 400V 3~ 50Hz”
Installation
Cooling water must be supplied at 5–20 °C inlet with at least 1 bar differential pressure and no more than 6 bar, and pipes must be insulated if below dew point.
Defines the cooling water utility requirements and condensation risk at the installation site.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops significantly at low temperatures, from 2.2 kW at 0 °C to only 0.3 kW at -80 °C, which must be considered for load sizing.
Affects selection because available cooling capacity at the target low temperature may be insufficient for the process load.
Source
“at -80°C 0,3 kW”
Performance
The pump delivers 67 l/min max and 3 bar max pressure, with 28 l/min still available at 2.5 bar, so high-pressure-drop loops can be served.
Needed to verify the circulation pump can overcome the pressure drop of the external application.
Source
“max. delivery pressure 3 bar”
Selection
The unit uses R-452A (GWP 2141, 0.95 kg) and R-23 (GWP 14800, 0.39 kg) refrigerants, giving a total CO2 equivalent of about 7.8 t.
Relevant for refrigerant regulations, leak monitoring obligations and environmental compliance when selecting the unit.